Abstract
We have developed a simple and inexpensive method that improves sensitivity of protein and antigen detection in standard PAGE procedures. Our technique uses a sample microloader device with a funnel-like structure, filled with a 4% stacking gel. When attach to the top of a polyacrylamide slab gel, the proteins in a sample are concentrated by electrophoresis into a small volume as they emerge from the device's narrow outlet. Our microloader has several advantages over previous devices, including simple assembly, high versatility, and absence of cross-contamination between lanes. Addition of this device to a slab gel results in a fivefold increase in the sensitivity of antigen detection in a Western blot. As a result, less protein is required for loading and signal detection. Our protocol is a straightforward modification of a standard experimental technique, and is especially useful when only limited sample quantities are available.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23400834 PMCID: PMC3768259 DOI: 10.1002/elps.201200534
Source DB: PubMed Journal: Electrophoresis ISSN: 0173-0835 Impact factor: 3.535