Numerical investigations of MRI RF field induced heating for external fixation devices.

Yan Liu1, Jianxiang Shen, Wolfgang Kainz, Songsong Qian, Wen Wu, Ji Chen.   

Abstract

BACKGROUND: The magnetic resonance imaging (MRI) radio frequency (RF) field induced heating on external fixation devices can be very high in the vicinity of device screws. Such induced RF heating is related to device constructs, device placements, as well as the device insertion depth into human subjects. In this study, computational modeling is performed to determine factors associated with such induced heating.
METHODS: Numerical modeling, based on the finite-difference time-domain (FDTD) method, is used to evaluate the temperature rises near external device screw tips inside the ASTM phantom for both 1.5-T and 3-T MRI systems. The modeling approach consists of 1) the development of RF coils for 1.5-T and 3-T, 2) the electromagnetic simulations of energy deposition near the screw tips of external fixation devices, and 3) the thermal simulations of temperature rises near the tips of these devices.
RESULTS: It is found that changing insertion depth and screw spacing could largely affect the heating of these devices. In 1.5-T MRI system, smaller insertion depth and larger pin spacing will lead to higher temperature rise. However, for 3-T MRI system, the relation is not very clear when insertion depth is larger than 5 cm or when pin spacing became larger than 20 cm. The effect of connection bar material on device heating is also studied and the heating mechanism of the device is analysed.
CONCLUSIONS: Numerical simulation is used to study RF heating for external fixation devices in both 1.5-T and 3-T MRI coils. Typically, shallower insertion depth and larger pin spacing with conductive bar lead to higher RF heating. The heating mechanism is explained using induced current along the device and power decay inside ASTM phantom.

Entities:  

Mesh:

Year:  2013        PMID: 23394173      PMCID: PMC3610272          DOI: 10.1186/1475-925X-12-12

Source DB:  PubMed          Journal:  Biomed Eng Online        ISSN: 1475-925X            Impact factor:   2.819


  10 in total

1.  Safety of orthopedic implants in magnetic resonance imaging: an experimental verification.

Authors:  Ritabh Kumar; Richard A Lerski; Stephen Gandy; Benedict A Clift; Rami J Abboud
Journal:  J Orthop Res       Date:  2006-09       Impact factor: 3.494

2.  Safety evaluation of large external fixation clamps and frames in a magnetic resonance environment.

Authors:  Roger Luechinger; Peter Boesiger; John A Disegi
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2007-07       Impact factor: 3.368

Review 3.  Standards for medical devices in MRI: present and future.

Authors:  Terry O Woods
Journal:  J Magn Reson Imaging       Date:  2007-11       Impact factor: 4.813

4.  Numerical simulation of SAR and B1-field inhomogeneity of shielded RF coils loaded with the human head.

Authors:  J Chen; Z Feng; J M Jin
Journal:  IEEE Trans Biomed Eng       Date:  1998-05       Impact factor: 4.538

5.  Heating of metallic rods induced by time-varying gradient fields in MRI.

Authors:  Khaled El Bannan; William Handler; Blaine Chronik; Shaun P Salisbury
Journal:  J Magn Reson Imaging       Date:  2013-01-04       Impact factor: 4.813

6.  Computational and experimental studies of an orthopedic implant: MRI-related heating at 1.5-T/64-MHz and 3-T/128-MHz.

Authors:  Yan Liu; Ji Chen; Frank G Shellock; Wolfgang Kainz
Journal:  J Magn Reson Imaging       Date:  2012-07-31       Impact factor: 4.813

7.  Evaluation of RF heating on hip joint implant in phantom during MRI examinations.

Authors:  Hiroyuki Muranaka; Takayoshi Horiguchi; Yoshitake Ueda; Shuji Usui; Nobuyoshi Tanki; Osamu Nakamura
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2010-07-20

8.  Heating of metallic implants and instruments induced by gradient switching in a 1.5-Tesla whole-body unit.

Authors:  Hansjörg Graf; Günter Steidle; Fritz Schick
Journal:  J Magn Reson Imaging       Date:  2007-11       Impact factor: 4.813

9.  Temperature measurement on neurological pulse generators during MR scans.

Authors:  Wolfgang Kainz; Georg Neubauer; Richard Uberbacher; François Alesch; Dulciana Dias Chan
Journal:  Biomed Eng Online       Date:  2002-09-12       Impact factor: 2.819

10.  Complexity of MRI induced heating on metallic leads: experimental measurements of 374 configurations.

Authors:  Eugenio Mattei; Michele Triventi; Giovanni Calcagnini; Federica Censi; Wolfgang Kainz; Gonzalo Mendoza; Howard I Bassen; Pietro Bartolini
Journal:  Biomed Eng Online       Date:  2008-03-03       Impact factor: 2.819

  10 in total
  2 in total

1.  Safety and image quality of MR-conditional external fixators for 1.5 Tesla extremity MR.

Authors:  David H Ballard; John D Garrett; Alberto A Simoncini; Silvia Barbeito; Massimo Max Morandi
Journal:  Emerg Radiol       Date:  2021-01-15

2.  Use of a radio frequency shield during 1.5 and 3.0 Tesla magnetic resonance imaging: experimental evaluation.

Authors:  Christopher P Favazza; Deirdre M King; Heidi A Edmonson; Joel P Felmlee; Phillip J Rossman; Nicholas J Hangiandreou; Robert E Watson; Krzysztof R Gorny
Journal:  Med Devices (Auckl)       Date:  2014-10-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.