Standards for medical devices in MRI: present and future.

Terry O Woods1.   

Abstract

The purpose of this review is to define the current standards addressing safety of medical devices in MRI and to describe ongoing standards development efforts. The American Society for Testing and Materials (ASTM International) began developing standard test methods for determining the MR safety of medical devices in MRI in 1997. To date, five ASTM standards addressing testing and marking medical devices and other items for use in the MR environment have been published. International Standards Organization (ISO) 14630, the general requirements standard for nonactive surgical implants, is currently being revised to include information about MR safety of passive implants and to reference the ASTM standards. To address the unique safety issues of active implants, and in particular, active implants with leads, like pacemakers and neurostimulators, a joint working group between ISO TC150/SC6 on active implants and International Electrotechnical Commission (IEC) SC 62B MT40 on magnetic resonance equipment for medical diagnosis is working to develop a technical specification for active implantable medical devices (AIMDs) in MRI. While much progress has been made, work still needs to continue to develop a complete body of test methods to allow the evaluation of the safety of medical devices in the MR environment.

Mesh:

Year:  2007        PMID: 17969160     DOI: 10.1002/jmri.21140

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  20 in total

1.  Deployment of RFID in healthcare facilities-experimental design in MRI department.

Authors:  Chen-Yang Cheng; Jyh-Wen Chai
Journal:  J Med Syst       Date:  2011-11-10       Impact factor: 4.460

2.  Evaluation of 39 medical implants at 7.0 T.

Authors:  David X Feng; Joseph P McCauley; Fea K Morgan-Curtis; Redoan A Salam; David R Pennell; Mary E Loveless; Adrienne N Dula
Journal:  Br J Radiol       Date:  2015-10-20       Impact factor: 3.039

Review 3.  Implementation of a comprehensive MR safety course for medical students.

Authors:  Steffen Sammet; Christina L Sammet
Journal:  J Magn Reson Imaging       Date:  2015-07-14       Impact factor: 4.813

4.  Evaluation of a Workflow to Define Low Specific Absorption Rate MRI Protocols for Patients With Active Implantable Medical Devices.

Authors:  Jessica A Martinez; Kévin Moulin; Bryan Yoo; Yu Shi; Hyun J Kim; Pablo J Villablanca; Daniel B Ennis
Journal:  J Magn Reson Imaging       Date:  2020-01-10       Impact factor: 4.813

Review 5.  Occupational exposure to electromagnetic fields in magnetic resonance environment: basic aspects and review of exposure assessment approaches.

Authors:  Valentina Hartwig; Stefania Romeo; Olga Zeni
Journal:  Med Biol Eng Comput       Date:  2018-01-18       Impact factor: 2.602

6.  Design and self-assembly of siRNA-functionalized RNA nanoparticles for use in automated nanomedicine.

Authors:  Kirill A Afonin; Wade W Grabow; Faye M Walker; Eckart Bindewald; Marina A Dobrovolskaia; Bruce A Shapiro; Luc Jaeger
Journal:  Nat Protoc       Date:  2011-12-01       Impact factor: 13.491

7.  Modeling Endovascular MRI Coil Coupling With Transmit RF Excitation.

Authors:  Madhav Venkateswaran; Orhan Unal; Samuel Hurley; Alexey Samsonov; Peng Wang; Sean B Fain; Krishna N Kurpad
Journal:  IEEE Trans Biomed Eng       Date:  2016-03-04       Impact factor: 4.538

8.  A 1.5T MRI-conditional 12-lead electrocardiogram for MRI and intra-MR intervention.

Authors:  Zion Tsz Ho Tse; Charles L Dumoulin; Gari D Clifford; Jeff Schweitzer; Lei Qin; Julien Oster; Michael Jerosch-Herold; Raymond Y Kwong; Gregory Michaud; William G Stevenson; Ehud J Schmidt
Journal:  Magn Reson Med       Date:  2014-03       Impact factor: 4.668

Review 9.  Magnetic resonance safety.

Authors:  Steffen Sammet
Journal:  Abdom Radiol (NY)       Date:  2016-03

10.  Numerical investigations of MRI RF field induced heating for external fixation devices.

Authors:  Yan Liu; Jianxiang Shen; Wolfgang Kainz; Songsong Qian; Wen Wu; Ji Chen
Journal:  Biomed Eng Online       Date:  2013-02-09       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.