A comparative study on reliability of point of care sodium and potassium estimation in a pediatric intensive care unit.

Viswas Chhapola1, Sandeep Kumar Kanwal, Rajni Sharma, Virendra Kumar.   

Abstract

OBJECTIVE: To compare sodium and potassium levels in children as done with Blood Gas Analyzer (BGA) at point of care testing in pediatric ICU vs. that done in laboratory electrolyte analyzer.
METHODS: This prospective method comparison study was done from February to April 2012 in Pediatric ICU of tertiary care hospital at Delhi. Sixty consecutive patients were tested during the period. Paired blood samples for venous blood gas to be tested on BGA and serum electrolytes to be tested on auto-analyzers (AA) were taken as per standard technique. Data was collected and 59 paired samples were analyzed for sodium and potassium levels. They were analyzed according to CLSI document EP15-A2 using ACB method comparison software.
RESULTS: Mean sodium measured on the BGA was 132.8 ± 12.2 mmol/L where as measured by AA was 141.5 ± 11.1 mmol/L. The mean difference between the two was -8.76 mmol/L (p < 0.001). The difference was statistically significant in all three subgroups of hypernatremia, isonatremia and hyponatremia (p < 0.001). Potassium level in BGA was 3.53 ± 0.81 mmol/L and AA was 4.28 ± 1.05 mmol/L. The mean difference between the BGA and AA was -0.75 mmol/L (p < 0.0001). The difference was statistically significant in patients with normokalemia and hyperkalemia (p < 0.0001). The difference was non significant in patients with hypokalemia (p = 0.051).
CONCLUSIONS: Blood gas analyzers underestimates Na + and K + values if sampling is done using liquid sodium heparin and if all other potential pre-analytical errors of testing are taken care of. The Bland Altman's analysis in the present study showed a significant systematic bias and very wide limits of agreement for both sodium and potassium, which is not clinically acceptable.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23392748     DOI: 10.1007/s12098-013-0977-z

Source DB:  PubMed          Journal:  Indian J Pediatr        ISSN: 0019-5456            Impact factor:   1.967


  10 in total

Review 1.  Errors in laboratory medicine.

Authors:  Pierangelo Bonini; Mario Plebani; Ferruccio Ceriotti; Francesca Rubboli
Journal:  Clin Chem       Date:  2002-05       Impact factor: 8.327

2.  Comparison of point-of-care versus central laboratory measurement of electrolyte concentrations on calculations of the anion gap and the strong ion difference.

Authors:  Hiroshi Morimatsu; Jens Rocktäschel; Rinaldo Bellomo; Shigehiko Uchino; Donna Goldsmith; Geoffrey Gutteridge
Journal:  Anesthesiology       Date:  2003-05       Impact factor: 7.892

3.  Charts of operational process specifications ("OPSpecs charts") for assessing the precision, accuracy, and quality control needed to satisfy proficiency testing performance criteria.

Authors:  J O Westgard
Journal:  Clin Chem       Date:  1992-07       Impact factor: 8.327

4.  Electrolytes in sick neonates - which sodium is the right answer?

Authors:  Richard I King; Richard J Mackay; Christopher M Florkowski; Adrienne M Lynn
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2012-01-03       Impact factor: 5.747

5.  Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes.

Authors:  Miranda van Berkel; Volkher Scharnhorst
Journal:  Clin Chem Lab Med       Date:  2010-12-14       Impact factor: 3.694

6.  Heparin interference in whole blood sodium measurements in a pediatric setting.

Authors:  Paul M Yip; Man Khun Chan; Nuala Zielinski; Khosrow Adeli
Journal:  Clin Biochem       Date:  2006-01-26       Impact factor: 3.281

Review 7.  Statistics review 4: sample size calculations.

Authors:  Elise Whitley; Jonathan Ball
Journal:  Crit Care       Date:  2002-05-10       Impact factor: 9.097

8.  The use of heparin in preparing samples for blood-gas analysis.

Authors:  Chris Higgins
Journal:  MLO Med Lab Obs       Date:  2007-10

9.  Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes.

Authors:  Anunaya Jain; Imron Subhan; Mahesh Joshi
Journal:  Int J Emerg Med       Date:  2009-02-24

10.  Electrolytes assessed by point-of-care testing - Are the values comparable with results obtained from the central laboratory?

Authors:  Binila Chacko; John V Peter; Shalom Patole; Jude J Fleming; Ratnasamy Selvakumar
Journal:  Indian J Crit Care Med       Date:  2011-01
  10 in total
  12 in total

1.  Agreement of arterial sodium and arterial potassium levels with venous sodium and venous potassium in patients admitted to intensive care unit.

Authors:  Sunil Kumar Nanda; Lopamudra Ray; Asha Dinakaran
Journal:  J Clin Diagn Res       Date:  2015-02-01

2.  Evaluating the utility of rapid point-of-care potassium testing for the early identification of hyperkalemia in patients with chronic kidney disease in the emergency department.

Authors:  Je Sung You; Yoo Seok Park; Hyun Soo Chung; Hye Sun Lee; Youngseon Joo; Jong Woo Park; Sung Phil Chung; Shin Ho Lee; Hahn Shick Lee
Journal:  Yonsei Med J       Date:  2014-09       Impact factor: 2.759

3.  Potassium disturbance associated with an inpatient childhood asthma pathway.

Authors:  Jonathan Hung; Timothy Kraft; Brent Seefried; David W Johnson; Michelle Bailey; Deonne Dersch-Mills
Journal:  Paediatr Child Health       Date:  2019-10-12       Impact factor: 2.253

4.  Use of liquid heparin for blood gas sampling in pediatric intensive care unit: A comparative study of effects of varying volumes of heparin on blood gas parameters.

Authors:  Viswas Chhapola; Sandeep Kumar; Pallavi Goyal; Rajni Sharma
Journal:  Indian J Crit Care Med       Date:  2013-11

5.  Point of care blood gases with electrolytes and lactates in adult emergencies.

Authors:  Dheeraj Kapoor; Meghana Srivastava; Pritam Singh
Journal:  Int J Crit Illn Inj Sci       Date:  2014-07

6.  Analysis of bias in measurements of potassium, sodium and hemoglobin by an emergency department-based blood gas analyzer relative to hospital laboratory autoanalyzer results.

Authors:  Jian Bo Zhang; Ji Lin; Xiao Dong Zhao
Journal:  PLoS One       Date:  2015-04-07       Impact factor: 3.240

7.  Bedside ABG, electrolytes, lactate and procalcitonin in emergency pediatrics.

Authors:  Prerna Batra; Ajeet Kumar Dwivedi; Neha Thakur
Journal:  Int J Crit Illn Inj Sci       Date:  2014-07

8.  Are sodium and potassium results on arterial blood gas analyzer equivalent to those on electrolyte analyzer?

Authors:  Shalini Gupta; Ashwani K Gupta; Kamaljit Singh; Minni Verma
Journal:  Indian J Crit Care Med       Date:  2016-04

9.  Is liquid heparin comparable to dry balanced heparin for blood gas sampling in intensive care unit?

Authors:  Viswas Chhapola; Sandeep Kumar; Pallavi Goyal
Journal:  Indian J Crit Care Med       Date:  2014-01

10.  To heparinize or not to heparinize: Effect on arterial blood gas measurements.

Authors:  Valentine Lobo
Journal:  Indian J Crit Care Med       Date:  2014-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.