Literature DB >> 23366780

On the cause and control of residual voltage generated by electrical stimulation of neural tissue.

Ashwati Krishnan1, Shawn K Kelly.   

Abstract

Functional electrical stimulation of neural tissue is traditionally performed with symmetric cathodic-first biphasic pulses of current through an electrode/electrolyte interface. When the interface is modeled by a series R-C circuit, as is sometimes done for stimulator circuit design, the appearance of a net residual voltage across the electrode cannot be explained. Residual voltage can cause polarization of the electrode and pose a problem for safe electrical stimulation. This paper aims to (1) theoretically explain one reason for the residual voltage, which is the inclusion of the Faradaic impedance (2) suggest a simple dynamic feedback mechanism to eliminate residual voltage.

Mesh:

Year:  2012        PMID: 23366780      PMCID: PMC4876046          DOI: 10.1109/EMBC.2012.6346819

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  10 in total

1.  Brief, noninjurious electric waveform for stimulation of the brain.

Authors:  J C LILLY; J R HUGHES; E C ALVORD; T W GALKIN
Journal:  Science       Date:  1955-04-01       Impact factor: 47.728

2.  Charge density and charge per phase as cofactors in neural injury induced by electrical stimulation.

Authors:  D B McCreery; W F Agnew; T G Yuen; L Bullara
Journal:  IEEE Trans Biomed Eng       Date:  1990-10       Impact factor: 4.538

3.  System design of a cochlear implant.

Authors:  R L White
Journal:  IEEE Eng Med Biol Mag       Date:  1987

4.  A Low-Power Blocking-Capacitor-Free Charge-Balanced Electrode-Stimulator Chip With Less Than 6 nA DC Error for 1-mA Full-Scale Stimulation.

Authors:  R Sarpeshkar
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2007-09       Impact factor: 3.833

5.  An active approach for charge balancing in functional electrical stimulation.

Authors:  Kriangkrai Sooksood; Thomas Stieglitz; Maurits Ortmanns
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2010-06       Impact factor: 3.833

Review 6.  Neural stimulation and recording electrodes.

Authors:  Stuart F Cogan
Journal:  Annu Rev Biomed Eng       Date:  2008       Impact factor: 9.590

Review 7.  Electrical stimulation of neural tissue to evoke behavioral responses.

Authors:  E J Tehovnik
Journal:  J Neurosci Methods       Date:  1996-03       Impact factor: 2.390

Review 8.  Historical evolution of circuit models for the electrode-electrolyte interface.

Authors:  L A Geddes
Journal:  Ann Biomed Eng       Date:  1997 Jan-Feb       Impact factor: 3.934

9.  Development and implantation of a minimally invasive wireless subretinal neurostimulator.

Authors:  Douglas B Shire; Shawn K Kelly; Jinghua Chen; Patrick Doyle; Marcus D Gingerich; Stuart F Cogan; William A Drohan; Oscar Mendoza; Luke Theogarajan; John L Wyatt; Joseph F Rizzo
Journal:  IEEE Trans Biomed Eng       Date:  2009-04-28       Impact factor: 4.538

10.  Potential-biased, asymmetric waveforms for charge-injection with activated iridium oxide (AIROF) neural stimulation electrodes.

Authors:  Stuart F Cogan; Philip R Troyk; Julia Ehrlich; Timothy D Plante; David E Detlefsen
Journal:  IEEE Trans Biomed Eng       Date:  2006-02       Impact factor: 4.538

  10 in total
  4 in total

1.  On Using Residual Voltage to Estimate Electrode Model Parameters for Damage Detection.

Authors:  Ashwati Krishnan; Shawn K Kelly
Journal:  IEEE Biomed Circuits Syst Conf       Date:  2015-10

2.  Redundant safety features in a high-channel-count retinal neurostimulator.

Authors:  Shawn K Kelly; William F Ellersick; Ashwati Krishnan; Patrick Doyle; Douglas B Shire; John L Wyatt; Joseph F Rizzo
Journal:  IEEE Biomed Circuits Syst Conf       Date:  2014-10

3.  A CMOS Current Steering Neurostimulation Array With Integrated DAC Calibration and Charge Balancing.

Authors:  Elliot Greenwald; Christoph Maier; Qihong Wang; Robert Beaulieu; Ralph Etienne-Cummings; Gert Cauwenberghs; Nitish Thakor
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2017-01-16       Impact factor: 3.833

4.  Residual voltage as an ad-hoc indicator of electrode damage in biphasic electrical stimulation.

Authors:  Ashwati Krishnan; Mats Forssell; Zhanhong Du; X Tracy Cui; Gary K Fedder; Shawn K Kelly
Journal:  J Neural Eng       Date:  2021-08-12       Impact factor: 5.043

  4 in total