Literature DB >> 23359608

Comparison of ceramic-on-metal and metal-on-metal hip prostheses under adverse conditions.

Sophie Williams1, Mazen Al-Hajjar, Graham H Isaac, John Fisher.   

Abstract

Ceramic-on-metal (COM) hip replacements, where the head is BIOLOX® Delta ceramic and the liner is CoCrMo alloy, have demonstrated reduced wear under standard simulator conditions compared to metal-on-metal (MOM) bearings. COM hips are now being used clinically around the world. MOM hip resurfacings have raised concerns regarding poor clinical performance and increased in vivo wear was associated with steeply inclined acetabular components and translationally malpositioned components. The aim of this study was to compare the wear rates of MOM and COM total hip prostheses under adverse edge-loading conditions in a hip simulator test. COM and MOM 36 mm hip prostheses were tested in a hip simulator, with liners mounted to provide a clinical inclination angle of 55°. A simplified gait cycle and microseparation conditions were applied for two million cycles in 25% new born calf serum. The overall mean volumetric wear rate of COM bearings under adverse conditions was 0.36 ± 0.55 mm³/million cycles; this was significantly less than MOM wear (1.32 ± 0.91 mm³/million cycles). Under these adverse conditions; the contact zone on the head intersects the rim of the cup causing substantially elevated contact stresses, disrupting the protective boundary and mixed lubrication regime causing changes in types and severity of wear mechanisms. In COM bearings, the harder head does not become damaged when there is lubricant starvation and wear does not accelerate. In conclusion, COM bearings showed reduced wear compared to MOM bearings under standard and adverse clinically relevant simulator conditions and COM bearings may provide an advantage over MOM bearings under edge-loading conditions clinically.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23359608     DOI: 10.1002/jbm.b.32880

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  8 in total

Review 1.  [Total hip arthroplasty in young patients : Bearings and custom-made prostheses].

Authors:  C Benignus; M Morlock; J Beckmann
Journal:  Orthopade       Date:  2019-04       Impact factor: 1.087

2.  No Difference in Reoperations at 2 Years Between Ceramic-on-metal and Metal-on-metal THA: A Randomized Trial.

Authors:  C Anderson Engh; Supatra Sritulanondha; Abigail Korczak; Terrence David Whalen; Douglas D R Naudie; Richard W McCalden; Steven J MacDonald
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

3.  Wear performance of ceramic-on-metal hip bearings.

Authors:  Jörn Reinders; Robert Sonntag; Christian Heisel; Tobias Reiner; Leo Vot; Jan Philippe Kretzer
Journal:  PLoS One       Date:  2013-08-29       Impact factor: 3.240

4.  Future bearing surfaces in total hip arthroplasty.

Authors:  Jun-Dong Chang
Journal:  Clin Orthop Surg       Date:  2014-02-14

5.  Clinical Results and Serum Metal Ion Concentrations following Ceramic-on-Metal Total Hip Arthroplasty at a Mean Follow-Up of 60 Months.

Authors:  W Maurer-Ertl; D Pranckh-Matzke; J Friesenbichler; G Bratschitsch; L A Holzer; M Maier; A Leithner
Journal:  Biomed Res Int       Date:  2017-03-08       Impact factor: 3.411

6.  An in vitro simulation model to assess the severity of edge loading and wear, due to variations in component positioning in hip joint replacements.

Authors:  O O'Dwyer Lancaster-Jones; S Williams; L M Jennings; J Thompson; G H Isaac; J Fisher; M Al-Hajjar
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2017-09-23       Impact factor: 3.368

7.  Wear of composite ceramics in mixed-material combinations in total hip replacement under adverse edge loading conditions.

Authors:  Mazen Al-Hajjar; Silvia Carbone; Louise M Jennings; Sabine Begand; Thomas Oberbach; Daniel Delfosse; John Fisher
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-04-08       Impact factor: 3.368

8.  Protrusion of a ceramic femoral head through the acetabular metallic cup in total-hip arthroplasty: A case report.

Authors:  Ding Zhao; De-Bao Zhang; Dong-Feng Han; Gui-Shan Gu
Journal:  Medicine (Baltimore)       Date:  2020-06-12       Impact factor: 1.817

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.