| Literature DB >> 23349650 |
Harish K Malhotra1, Vincent Wu, Zhou Wang, Sachin Patil.
Abstract
PURPOSE: Standard treatment plan libraries are often used to ensure a quick turn-around time for vaginal cylinder treatments. Recently there is increasing interest in transitioning from conventional 2D radiograph based brachytherapy to 3D image based brachytherapy, which has resulted in a substantial increase in treatment planning time and decrease in patient through-put. We describe a novel technique that significantly reduces the treatment planning time for CT-based vaginal cylinder brachytherapy.Entities:
Keywords: high-dose-rate brachytherapy; vaginal brachytherapy
Year: 2012 PMID: 23349650 PMCID: PMC3552630 DOI: 10.5114/jcb.2012.29365
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Standard treatment plan library with first 7 active dwell positions for 2.0 cm cylinder diameters for stainless steel vaginal cylinder applicator (Nucletron). Included are coordinates of the points on the applicator surface and at a depth of 5 mm from the applicator surface; also included is the location of the dwell position coincident with the origin of the treatment plan. It is important to know that in this table the dose values shown at cylinder surface and at 5 mm from it are for representative purposes only. Their actual values will change based on the dose-prescription, normalization as well as optimization selected by the user on the fly
| Origin = 3rd dwell position | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Applicator pts – surface | ||||||||||||
| Name |
|
|
| Coord. system | Act. dose [cGy] | Act. dose (%) | Normalization | Norm. dose (%) | Optimization | Opt. rel. dose | Opt. weight | |
| A1 | 0.0 | 16.5 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A2 | 3.5 | 15.8 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A3 | 6.2 | 14.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A4 | 8.2 | 11.5 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A5 | 9.2 | 9.2 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A6 | 10.0 | 5.2 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A7 | 10.0 | 0.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A8 | 10.0 | −5.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A9 | 10.0 | −15.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A10 | 10.0 | −10.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
| A11 | 10.0 | −20.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 | |
|
| ||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |
| A12 | 0.0 | 21.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A13 | 3.0 | 20.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A14 | 5.5 | 19.3 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A15 | 8.1 | 18.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A16 | 10.0 | 16.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A17 | 12.0 | 14.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A18 | 13.5 | 11.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A19 | 14.8 | 8.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A20 | 15.0 | 5.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A21 | 15.0 | 0.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A22 | 15.0 | −5.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A23 | 15.0 | −10.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A24 | 15.0 | −15.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
| A25 | 15.0 | −20.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 | |
Standard treatment plan library with first 7 active dwell positions for 2.5 cm cylinder diameters for stainless steel vaginal cylinder applicator (Nucletron). Included are coordinates of the points on the applicator surface and at a distance of 5 mm; also included is the location of the dwell position coincident with the origin of the treatment plan. It is important to know that in this table the dose values shown at cylinder surface and at 5 mm from it are for representative purposes only. Their actual values will change based on the dose-prescription, normalization as well as optimization selected by the user on the fly
| Origin = 4th dwell position | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Applicator pts – surface | |||||||||||
| Name |
|
|
| Coord. system | Act. dose [cGy] | Act. dose (%) | Normalization | Norm. dose (%) | Optimization | Opt. rel. dose | Opt. weight |
| A1 | 0.0 | 21.5 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A2 | 5.0 | 20.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A3 | 9.3 | 16.5 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A4 | 11.8 | 11.5 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A5 | 12.5 | 5.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A6 | 12.5 | 0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A7 | 12.5 | −5.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A8 | 12.5 | −10.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A9 | 12.5 | −15.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
|
| |||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| A10 | 0.0 | 26.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A11 | 4.5 | 25.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A12 | 9.6 | 22.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A13 | 12.5 | 20.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A14 | 15.5 | 16.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A15 | 16.4 | 13.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A16 | 17.0 | 10.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A17 | 17.5 | 5.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A18 | 17.5 | 0.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A19 | 17.5 | −5.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A20 | 17.5 | −10.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A21 | 17.5 | −15.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
Standard treatment plan library with first 7 active dwell positions for 3.0 cm cylinder diameters for stainless steel vaginal cylinder applicator (Nucletron). Included are coordinates of the points on the applicator surface and at a distance of 5 mm; also included is the location of the dwell position coincident with the origin of the treatment plan. It is important to know that in this table the dose values shown at cylinder surface and at 5 mm from it are for representative purposes only. Their actual values will change based on the dose-prescription, normalization as well as optimization selected by the user on the fly
| Origin = 3rd dwell position | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Applicator pts – surface | |||||||||||
| Name |
|
|
| Coord. system | Act. dose [cGy] | Act. dose (%) | Normalization | Norm. dose (%) | Optimization | Opt. rel. dose | Opt. weight |
| A1 | 0.0 | 16.5 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A2 | 5.7 | 15.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A3 | 10.3 | 11.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A4 | 12.7 | 7.7 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A5 | 14.0 | 4.7 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A6 | 15.0 | 0.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A7 | 15.0 | −10.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A8 | 15.0 | −20.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
|
| |||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| A9 | 0.0 | 21.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A10 | 6.7 | 20.2 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A11 | 10.4 | 17.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A12 | 15.0 | 13.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A13 | 18.7 | 5.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A14 | 20.0 | 0.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A15 | 20.0 | −5.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A16 | 20.0 | −10.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A17 | 20.0 | −15.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A18 | 20.0 | −20.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
Standard treatment plan library with first 7 active dwell positions for 3.5 cm cylinder diameters for stainless steel vaginal cylinder applicator (Nucletron). Included are coordinates of the points on the applicator surface and at a distance of 5 mm; also included is the location of the dwell position coincident with the origin of the treatment plan. It is important to know that in this table the dose values shown at cylinder surface and at 5 mm from it are for representative purposes only. Their actual values will change based on the dose-prescription, normalization as well as optimization selected by the user on the fly
| Origin = 4th dwell position | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Applicator pts – surface | |||||||||||
| Name |
|
|
| Coord. system | Act. dose [cGy] | Act. dose (%) | Normalization | Norm. dose (%) | Optimization | Opt. rel. dose | Opt. weight |
| A1 | 0.0 | 21.5 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A2 | 4.5 | 21.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A3 | 7.8 | 20.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A4 | 12.5 | 16.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A5 | 16.3 | 10.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A6 | 17.5 | 5.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A7 | 17.5 | 0.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A8 | 17.5 | −5.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A9 | 17.5 | −10.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
| A10 | 17.5 | −15.0 | 0 | Applicator | 1000 | 100 | no | 100 | no | 100 | 100 |
|
| |||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| A11 | 0.0 | 26.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A12 | 6.5 | 25.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A13 | 10.5 | 23.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A14 | 16.0 | 18.5 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A15 | 19.0 | 15.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A16 | 22.0 | 10.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A17 | 22.5 | 5.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A18 | 22.5 | 0.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A19 | 22.5 | −5.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A20 | 22.5 | −10.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
| A21 | 22.5 | −15.0 | 0 | Applicator | 700 | 100 | yes | 100 | yes | 100 | 100 |
Fig. 1A screen capture showing the dose-points overlaid on a patient CT scan. Also shown are the isodose curves and other relevant information