Comparing the Tonojet disposable tonometer with the traditional Goldmann tonometer in glaucomatous and non-glaucomatous eyes.

S M J Farrell1, I Dooley, E O'Connell, S Bashir, A Foley-Nolan, F Kearns, P Logan, T Fulcher.   

Abstract

Applanation tonometry is widely performed as the most accurate method of measuring intraocular pressure in the clinical setting. Disposable tonometer heads reduce the risk of transmitting infection during applanation and eliminate tonometer head cleaning, which is time-consuming. We performed a study to investigate whether the Tonojet disposable tonometer (Luneau, France) gave comparable intraocular pressure measurements to the standard Goldmann tonometer produced by Haag Streit (Switzerland). The study was performed in a busy general secondary referral ophthalmology clinic in Beaumont Hospital over a 15-day period. We measured intraocular pressure with the Goldmann and Tonojet tonometers in 293 eyes. Patients were randomised as to which tonometer was used first. A diagnosis of glaucoma or ocular hypertension was noted. Central corneal thickness (CCT) was measured. The measurements taken with the Tonojet were on average 2.2 mmHg lower than corresponding measurements taken with the Goldmann applanation tonometer head. The bias was present over the entire range of pressures. The 95 % limits of agreement between the two tonometers ranged from -5.9 to +1.4 mmHg. In the cohort of eyes with ocular hypertension/glaucoma the difference was significantly higher (2.8 vs. 2.1 mmHg). The bias was independent of CCT. We found that the Tonojet disposable tonometer tends to underestimate intraocular pressure compared with the Goldmann. The bias is non-systematic, present over the entire range of pressures and independent of corneal thickness. We conclude that the Tonojet is not sufficiently accurate to constitute a suitable alternative to the Goldmann.

Entities:  

Mesh:

Year:  2013        PMID: 23341184     DOI: 10.1007/s10792-012-9711-5

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  23 in total

1.  [Further contribution to applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1961-06       Impact factor: 3.250

2.  [Applanation tonometry].

Authors:  H GOLDMANN; T SCHMIDT
Journal:  Ophthalmologica       Date:  1957-10       Impact factor: 3.250

Review 3.  Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.

Authors:  M J Doughty; M L Zaman
Journal:  Surv Ophthalmol       Date:  2000 Mar-Apr       Impact factor: 6.048

4.  Corneal curvature in applanation tonometry.

Authors:  H H Mark
Journal:  Am J Ophthalmol       Date:  1973-08       Impact factor: 5.258

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  A comparison of clinical performance between disposable and Goldmann tonometers.

Authors:  A P Maino; H J Uddin; A B Tullo
Journal:  Eye (Lond)       Date:  2006-05       Impact factor: 3.775

7.  The advanced glaucoma intervention study, 6: effect of cataract on visual field and visual acuity. The AGIS Investigators.

Authors: 
Journal:  Arch Ophthalmol       Date:  2000-12

8.  Axial length in applanation tonometry.

Authors:  Harry H Mark; Kim P Robbins; Tami L Mark
Journal:  J Cataract Refract Surg       Date:  2002-03       Impact factor: 3.351

9.  Comparison of glaucomatous progression between untreated patients with normal-tension glaucoma and patients with therapeutically reduced intraocular pressures. Collaborative Normal-Tension Glaucoma Study Group.

Authors: 
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

10.  The effectiveness of intraocular pressure reduction in the treatment of normal-tension glaucoma. Collaborative Normal-Tension Glaucoma Study Group.

Authors: 
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

View more
  1 in total

Review 1.  A Systematic Review Regarding Tonometry and the Transmission of Infectious Diseases.

Authors:  Nicole Atkins; William Hodge; Bruce Li
Journal:  J Clin Med Res       Date:  2018-01-26
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.