Yusuke Tahara1, Kenichi Nakashi, Ke Ji, Akihiro Ikeda, Kiyoshi Toko.
Abstract
We have developed a new portable taste sensor with a lipid/polymer membrane and conducted experiments to evaluate the sensor's performance. The fabricated sensor consists of a taste sensor chip (40 mm × 26 mm × 2.2 mm) with working and reference electrodes and a portable sensor device (80 mm × 25 mm × 20 mm). The working electrode consists of a taste-sensing site comprising a poly(hydroxyethyl)methacrylate (pHEMA) hydrogel layer with KCl as the electrolyte layer and a lipid/polymer membrane as the taste sensing element. The reference electrode comprises a polyvinyl chloride (PVC) membrane layer with a small hole and a pHEMA layer with KCl. The whole device is the size of a USB memory stick, making it suitable for portable use. The sensor's response to tannic acid as the standard astringency substance showed good accuracy and reproducibility, and was comparable with the performance of a commercially available taste sensing system. Thus, it is possible for this sensor to be used for in-field evaluations and it can make a significant contribution to the food industry, as well as in various fields of research.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23325168 PMCID: PMC3574722 DOI: 10.3390/s130101076
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1.(a) Fabricated taste sensor chip. (b) Structure of the working electrode. (c) Structure of the reference electrode.
Figure 2.Block diagram of portable taste sensor device.
Basic samples used in the taste experiments.
| Reference solution | 30 mM KCl, 0.3 mM tartaric acid |
| Saltiness | 300 mM KCl, 0.3 mM tartaric acid |
| Sourness | 30 mM KCl, 3mM tartaric acid |
| Umami | 10 mM monosodium glutamate |
| Bitterness (+) | 0.1 mM Quinine-HCl |
| Bitterness (–) | 0.01 vol% iso-alpha acid |
| Astringency | 0.05 wt% tannic acid |
| Sweetness | 1 M Sucrose |
Figure 3.Potential variation of the fabricated reference electrode. The experiment was carried out in the reference solution.
Figure 4.Response of the portable taste sensor to tannic acid. (a) Relative values. (b) CPA values.