Abstract
PURPOSE: The objectives of this study were to survey the radiographic exposure parameters, to measure the patient doses for intraoral dental radiography nationwide, and thus to establish the diagnostic reference levels (DRLs) in intraoral dental X-ray examination in Korea.Entities:
Keywords: Radiation Dosage; Radiation Protection; Radiography, Dental
Year: 2012 PMID: 23301210 PMCID: PMC3534178 DOI: 10.5624/isd.2012.42.4.237
Source DB: PubMed Journal: Imaging Sci Dent ISSN: 2233-7822
Regional distribution of dental radiographic equipment selected for this study
The number without parenthesis indicates that of dental radiographic equipment used for the survey of radiographic exposure parameters. The number within parentheses indicates that of dental radiographic equipment used for the measurement of patient dose.
Fig. 1The ionization chamber of DAP meter is positioned at the end of the exit cone of the intraoral X-ray machine for PED and DAP measurement.
Exposure parameters for adult mandibular molar intraoral dental radiography
FSD: focal spot - skin distance measured as the distance from focal spot to the end of the exit cone of the intraoral x-ray machine
Comparison of radiographic exposure parameters according to the size of hospital
*,†,‡: statistically significant at p<0.05
Comparison of radiographic exposure parameters according to the type of image receptor system
*,†,‡: statistically significant at p<0.05. DR: Digital Radiography, CR: Computed Radiography
Patient doses for adult mandibular molar intraoral dental radiography
Fig. 2Measured PED values for intraoral dental radiography.
Fig. 3Measured DAP values for intraoral dental radiography.
Comparison of mean patient does according to the size of hospital
*,†: statistically significant at p<0.05
Comparison of mean patient dose according to the type of image receptor system
*,†: statistically significant at p<0.05. DR: Digital Radiography, CR: Computed Radiography
Comparison of mean patient does according to the installation duration
Comparison of mean patient dose according to type of dental X-ray machine