Literature DB >> 23298114

Collimator design for a dedicated molecular breast imaging-guided biopsy system: proof-of-concept.

Amanda L Weinmann1, Carrie B Hruska, Amy L Conners, Michael K O'Connor.   

Abstract

PURPOSE: Molecular breast imaging (MBI) is a dedicated nuclear medicine breast imaging modality that employs dual-head cadmium zinc telluride (CZT) gamma cameras to functionally detect breast cancer. MBI has been shown to detect breast cancers otherwise occult on mammography and ultrasound. Currently, a MBI-guided biopsy system does not exist to biopsy such lesions. Our objective was to consider the utility of a novel conical slant-hole (CSH) collimator for rapid (<1 min) and accurate monitoring of lesion position to serve as part of a MBI-guided biopsy system.
METHODS: An initial CSH collimator design was derived from the dimensions of a parallel-hole collimator optimized for MBI performed with dual-head CZT gamma cameras. The parameters of the CSH collimator included the collimator height, cone slant angle, thickness of septa and cones of the collimator, and the annular areas exposed at the base of the cones. These parameters were varied within the geometric constraints of the MBI system to create several potential CSH collimator designs. The CSH collimator designs were evaluated using Monte Carlo simulations. The model included a breast compressed to a thickness of 6 cm with a 1-cm diameter lesion located 3 cm from the collimator face. The number of particles simulated was chosen to represent the count density of a low-dose, screening MBI study acquired with the parallel-hole collimator for 10 min after a ∼150 MBq (4 mCi) injection of Tc-99m sestamibi. The same number of particles was used for the CSH collimator simulations. In the resulting simulated images, the count sensitivity, spatial resolution, and accuracy of the lesion depth determined from the lesion profile width were evaluated.
RESULTS: The CSH collimator design with default parameters derived from the optimal parallel-hole collimator provided 1-min images with error in the lesion depth estimation of 1.1 ± 0.7 mm and over 21 times the lesion count sensitivity relative to 1-min images acquired with the current parallel-hole collimator. Sensitivity was increased via more vertical cone slant angles, larger annular areas, thinner cone walls, shorter cone heights, and thinner radiating septa. Full width at half maximum trended in the opposite direction as sensitivity for all parameters. There was less error in the depth estimates for less vertical slant angles, smaller annular areas, thinner cone walls, cone heights near 1 cm, and generally thinner radiating septa.
CONCLUSIONS: A Monte Carlo model was used to demonstrate the feasibility of a CSH collimator design for rapid biopsy application in molecular breast imaging. Specifically, lesion depth of a 1-cm diameter lesion positioned in the center of a typical breast can be estimated with error of less than 2 mm using circumferential count profiles of images acquired in 1 min.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23298114      PMCID: PMC3537760          DOI: 10.1118/1.4770274

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  Radionuclide-guided stereotactic prebiopsy localization of nonpalpable breast lesions with normal mammograms.

Authors:  I Khalkhali; F S Mishkin; L E Diggles; S R Klein
Journal:  J Nucl Med       Date:  1997-07       Impact factor: 10.057

2.  Evaluation of molecular breast imaging in women undergoing myocardial perfusion imaging with Tc-99m sestamibi.

Authors:  Carrie B Hruska; Deborah J Rhodes; Douglas A Collins; Cindy L Tortorelli; J Wells Askew; Michael K O'Connor
Journal:  J Womens Health (Larchmt)       Date:  2012-03-09       Impact factor: 2.681

3.  High-resolution scintimammography: a pilot study.

Authors:  Rachel F Brem; Joelle M Schoonjans; Douglas A Kieper; Stan Majewski; Steven Goodman; Cahid Civelek
Journal:  J Nucl Med       Date:  2002-07       Impact factor: 10.057

4.  Lexicon for standardized interpretation of gamma camera molecular breast imaging: observer agreement and diagnostic accuracy.

Authors:  Amy Lynn Conners; Carrie B Hruska; Cindy L Tortorelli; Robert W Maxwell; Deborah J Rhodes; Judy C Boughey; Wendie A Berg
Journal:  Eur J Nucl Med Mol Imaging       Date:  2012-06       Impact factor: 9.236

5.  Occult breast cancer: scintimammography with high-resolution breast-specific gamma camera in women at high risk for breast cancer.

Authors:  Rachel F Brem; Jocelyn A Rapelyea; Gilat Zisman; Kevin Mohtashemi; Joyce Raub; Christine B Teal; Stan Majewski; Benjamin L Welch
Journal:  Radiology       Date:  2005-08-26       Impact factor: 11.105

6.  Breast tumor localization using sestamibi.

Authors:  Leonard R Coover; Paul J Malaspina
Journal:  Clin Nucl Med       Date:  2002-03       Impact factor: 7.794

7.  Proof of concept for low-dose molecular breast imaging with a dual-head CZT gamma camera. Part II. Evaluation in patients.

Authors:  Carrie B Hruska; Amanda L Weinmann; Christina M Tello Skjerseth; Eric M Wagenaar; Amy L Conners; Cindy L Tortorelli; Robert W Maxwell; Deborah J Rhodes; Michael K O'Connor
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

8.  Proof of concept for low-dose molecular breast imaging with a dual-head CZT gamma camera. Part I. Evaluation in phantoms.

Authors:  Carrie B Hruska; Amanda L Weinmann; Michael K O'Connor
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

9.  High-resolution fluorodeoxyglucose positron emission tomography with compression ("positron emission mammography") is highly accurate in depicting primary breast cancer.

Authors:  Wendie A Berg; Irving N Weinberg; Deepa Narayanan; Mary E Lobrano; Eric Ross; Laura Amodei; Lorraine Tafra; Lee P Adler; Joseph Uddo; William Stein; Edward A Levine
Journal:  Breast J       Date:  2006 Jul-Aug       Impact factor: 2.431

10.  Scintimammography with dedicated breast camera detects and localizes occult carcinoma.

Authors:  Leonard R Coover; Gina Caravaglia; Phyllis Kuhn
Journal:  J Nucl Med       Date:  2004-04       Impact factor: 10.057

View more
  3 in total

Review 1.  Nuclear imaging of the breast: translating achievements in instrumentation into clinical use.

Authors:  Carrie B Hruska; Michael K O'Connor
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

Review 2.  Methodological aspects of 99mTc-sestamibi guided biopsy in breast cancer.

Authors:  A Collarino; R A Valdés Olmos; A F van der Hoeven; L M Pereira Arias-Bouda
Journal:  Clin Transl Imaging       Date:  2016-07-16

3.  Direct-Conversion Molecular Breast Imaging of Invasive Breast Cancer: Imaging Features, Extent of Invasive Disease, and Comparison Between Invasive Ductal and Lobular Histology.

Authors:  Amy Lynn Conners; Katie N Jones; Carrie B Hruska; Jennifer R Geske; Judy C Boughey; Deborah J Rhodes
Journal:  AJR Am J Roentgenol       Date:  2015-09       Impact factor: 3.959

  3 in total