Literature DB >> 23294781

Evaluation of a continuous blood glucose monitoring system using central venous microdialysis.

Fanny Schierenbeck1, Anders Franco-Cereceda, Jan Liska.   

Abstract

BACKGROUND: Glycemic control in critically ill patients has been shown to be beneficial. In this prospective study, we evaluated the accuracy and technical feasibility of a continuous glucose monitoring system using intravascular microdialysis.
METHOD: Fifty patients undergoing cardiac surgery were monitored using a 4 Fr intravenous microdialysis catheter (Eirus SLC™, Dipylon Medical AB, Solna, Sweden) percutaneously placed with the tip of the catheter positioned in the superior vena cava. The catheter was connected to the Eirus™ monitoring system, and the patients were monitored for up to 48 h postoperatively in the intensive care unit (ICU). As reference, arterial blood samples were taken every hour and analyzed in a blood gas analyzer.
RESULTS: Data were available from 48 patients. A total of 994 paired (arterial blood gas microdialysis) samples were obtained. Glucose correlation coefficient (R2) was 0.85. Using Clarke error grid analysis, 100% of the paired samples were in region AB, and 99% were in region A. Mean glucose level was 8.3 mmol/liter (149 mg/dl), mean relative difference was 0.2%, and mean absolute relative difference was 5%. A total of 99.2% of the paired samples were correct according to International Organization for Standardization (ISO) criteria. Bland-Altman analysis showed that bias ± limits of agreement were 0.02 ± 1.1 mmol/liter (0.36 ± 20 mg/dl).
CONCLUSIONS: Central venous microdialysis using the Eirus monitoring system is a highly accurate and reliable method for continuous blood glucose monitoring up to 48 h in ICU patients undergoing cardiac surgery. The system may thus be useful in critically ill ICU patients.
© 2012 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23294781      PMCID: PMC3570876          DOI: 10.1177/193229681200600615

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  34 in total

1.  Intensive insulin therapy in critically ill patients.

Authors:  G van den Berghe; P Wouters; F Weekers; C Verwaest; F Bruyninckx; M Schetz; D Vlasselaers; P Ferdinande; P Lauwers; R Bouillon
Journal:  N Engl J Med       Date:  2001-11-08       Impact factor: 91.245

Review 2.  The original Clarke Error Grid Analysis (EGA).

Authors:  William L Clarke
Journal:  Diabetes Technol Ther       Date:  2005-10       Impact factor: 6.118

3.  Intensive insulin therapy in the medical ICU.

Authors:  Greet Van den Berghe; Alexander Wilmer; Greet Hermans; Wouter Meersseman; Pieter J Wouters; Ilse Milants; Eric Van Wijngaerden; Herman Bobbaers; Roger Bouillon
Journal:  N Engl J Med       Date:  2006-02-02       Impact factor: 91.245

4.  Intraoperative hyperglycemia and perioperative outcomes in cardiac surgery patients.

Authors:  Gunjan Y Gandhi; Gregory A Nuttall; Martin D Abel; Charles J Mullany; Hartzell V Schaff; Brent A Williams; Lisa M Schrader; Robert A Rizza; M Molly McMahon
Journal:  Mayo Clin Proc       Date:  2005-07       Impact factor: 7.616

5.  Continuous intravenous insulin infusion reduces the incidence of deep sternal wound infection in diabetic patients after cardiac surgical procedures.

Authors:  A P Furnary; K J Zerr; G L Grunkemeier; A Starr
Journal:  Ann Thorac Surg       Date:  1999-02       Impact factor: 4.330

Review 6.  How does blood glucose control with insulin save lives in intensive care?

Authors:  Greet Van den Berghe
Journal:  J Clin Invest       Date:  2004-11       Impact factor: 14.808

7.  Intravascular microdialysis as a method for measuring glucose and lactate during and after cardiac surgery.

Authors:  Fanny Möller; Jan Liska; Fredrik Eidhagen; Anders Franco-Cereceda
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

8.  Continuous insulin infusion reduces mortality in patients with diabetes undergoing coronary artery bypass grafting.

Authors:  Anthony P Furnary; Guangqiang Gao; Gary L Grunkemeier; YingXing Wu; Kathryn J Zerr; Stephen O Bookin; H Storm Floten; Albert Starr
Journal:  J Thorac Cardiovasc Surg       Date:  2003-05       Impact factor: 5.209

Review 9.  Hyperglycemia in the critically ill patient.

Authors:  Michelle M Gearhart; Shilpa K Parbhoo
Journal:  AACN Clin Issues       Date:  2006 Jan-Mar

10.  Perioperative hyperglycemia is a strong correlate of postoperative infection in type II diabetic patients after coronary artery bypass grafting.

Authors:  Murat Guvener; Ilhan Pasaoglu; Metin Demircin; Mehmet Oc
Journal:  Endocr J       Date:  2002-10       Impact factor: 2.349

View more
  19 in total

1.  Assessment of changes in blood glucose concentration with intravascular microdialysis.

Authors:  Marc-Olivier Fischer; Corentin Gouëzel; Sabine Fradin; Vladimir Saplacan; Jean-Louis Gérard; Jean-Luc Fellahi; Jean-Luc Hanouz
Journal:  J Clin Monit Comput       Date:  2018-02-05       Impact factor: 2.502

2.  Dynamic properties of glucose complexity during the course of critical illness: a pilot study.

Authors:  Emmanuel Godat; Jean-Charles Preiser; Jean-Christophe Aude; Pierre Kalfon
Journal:  J Clin Monit Comput       Date:  2019-03-19       Impact factor: 2.502

3.  Analysis: continuous glucose monitoring in the intensive care unit.

Authors:  W Kenneth Ward
Journal:  J Diabetes Sci Technol       Date:  2012-11-01

4.  In Silico Testing of an Artificial-Intelligence-Based Artificial Pancreas Designed for Use in the Intensive Care Unit Setting.

Authors:  Leon DeJournett; Jeremy DeJournett
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

5.  Multiplexed and fully automated detection of metabolic biomarkers using microdialysis probe.

Authors:  Champak Das; Guochun Wang; Qian Sun; Bradley Ledden
Journal:  Sens Actuators B Chem       Date:  2016-07-25       Impact factor: 7.460

6.  Clinical Impact of Accurate Point-of-Care Glucose Monitoring for Tight Glycemic Control in Severely Burned Children.

Authors:  Nam K Tran; Zachary R Godwin; Amanda N Steele; Steven E Wolf; Tina L Palmieri
Journal:  Pediatr Crit Care Med       Date:  2016-09       Impact factor: 3.624

7.  Glucose Management Technologies for the Critically Ill.

Authors:  Pedro D Salinas; Carlos E Mendez
Journal:  J Diabetes Sci Technol       Date:  2019-01-13

Review 8.  Vascular Glucose Sensor Symposium: Continuous Glucose Monitoring Systems (CGMS) for Hospitalized and Ambulatory Patients at Risk for Hyperglycemia, Hypoglycemia, and Glycemic Variability.

Authors:  Jeffrey I Joseph; Marc C Torjman; Paul J Strasma
Journal:  J Diabetes Sci Technol       Date:  2015-06-15

Review 9.  Point-of-care blood glucose testing for diabetes care in hospitalized patients: an evidence-based review.

Authors:  Rajesh Rajendran; Gerry Rayman
Journal:  J Diabetes Sci Technol       Date:  2014-06-05

10.  Accuracy of 2 Different Continuous Glucose Monitoring Systems in Patients Undergoing Cardiac Surgery.

Authors:  Fanny Schierenbeck; Anders Franco-Cereceda; Jan Liska
Journal:  J Diabetes Sci Technol       Date:  2016-07-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.