Mechanical evaluation of orthodontic mini-implants of different lengths.

Matheus Melo Pithon1, Daniel Santos Fonseca Figueiredo, Dauro Douglas Oliveira.   

Abstract

PURPOSE: The current study sought to evaluate the influence of the length of the mini-implant on its mechanical properties.
MATERIALS AND METHODS: A total of 405 mini-implants from the same manufacturer were divided into 3 groups (n = 135) according to whether they were 6, 8, or 10 mm in length. An insertion torque test was performed using minipig osseous tissue of different cortical thicknesses (1 mm to 6 mm). To assess the maximum fracture torque during insertion, the mini-implants were inserted into swine femoral cortical bone. A digital torque gauge was positioned perpendicularly to the bone surface to measure this parameter. To evaluate the deformation and subsequent fracture due to a bending force, the mini-implants were subjected to loading orthogonal to their length at a speed of 0.5 mm per minute.
RESULTS: The insertion torque increased with increasing screw length and increasing cortical bone thickness. There were no statistically significant differences between the different implant lengths with respect to the insertion torque force required for fracture (P > .05) or the forces required for deformation (P > .05) and fracture (P > .05) when submitted to flexion.
CONCLUSIONS: Increasing the length of the screw does not increase its mechanical strength, but can efficiently reinforce the initial stability of mini-implants.
Copyright © 2013 American Association of Oral and Maxillofacial Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23273488     DOI: 10.1016/j.joms.2012.10.002

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  11 in total

1.  Does Change in Thread Shape Influence the Pull Out Strength of Mini Implants? An In vitro Study.

Authors:  Aniruddh V Yashwant; S Dilip; R Krishnaraj; K Ravi
Journal:  J Clin Diagn Res       Date:  2017-05-01

2.  Orthodontic mini-implant stability at different insertion depths : Sensitivity of three stability measurement methods.

Authors:  Manuel Nienkemper; Natascha Santel; Ralf Hönscheid; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2016-06-07       Impact factor: 1.938

3.  Effects of monocortical and bicortical mini-implant anchorage on bone-borne palatal expansion using finite element analysis.

Authors:  Robert J Lee; Won Moon; Christine Hong
Journal:  Am J Orthod Dentofacial Orthop       Date:  2017-05       Impact factor: 2.650

4.  Disc Anchoring with an Orthodontic Mini-Screw for Chronic Meniscocondylar Dislocation of TMJ.

Authors:  Thomas Zachariah; R S Neelakandan; M I Thaufiq Ahamed
Journal:  J Maxillofac Oral Surg       Date:  2014-12-06

5.  Impact of mini-implant length on stability at the initial healing period: a controlled clinical study.

Authors:  Manuel Nienkemper; Benedict Wilmes; Alexander Pauls; Dieter Drescher
Journal:  Head Face Med       Date:  2013-10-20       Impact factor: 2.151

6.  Quantitative evaluation of palatal bone thickness for the placement of orthodontic miniscrews in adults with different facial types.

Authors:  Yunji Wang; Ye Qiu; Henglang Liu; Jinlong He; Xiaoping Fan
Journal:  Saudi Med J       Date:  2017-10       Impact factor: 1.484

7.  Quantification of pro-inflammatory cytokines and osteoclastogenesis markers in successful and failed orthodontic mini-implants.

Authors:  Marcela Cristina Damião Andrucioli; Mírian Aiko Nakane Matsumoto; Sandra Yasuyo Fukada; Maria Conceição Pereira Saraiva; Ana Zilda Nazar Bergamo; Fábio Lourenço Romano; Raquel Assed Bezerra da Silva; Lea Assed Bezerra da Silva; Paulo Nelson-Filho
Journal:  J Appl Oral Sci       Date:  2019-10-07       Impact factor: 2.698

8.  Mechanical strength of stainless steel and titanium alloy mini-implants with different diameters: an experimental laboratory study.

Authors:  Sérgio Estelita Barros; Viviane Vanz; Kelly Chiqueto; Guilherme Janson; Eduardo Ferreira
Journal:  Prog Orthod       Date:  2021-03-22       Impact factor: 2.750

9.  Clinical and MRI Evaluation of Orthodontic Mini-Screws for Disc Repositioning in Internal Derangement of TMJ: A Prospective Study.

Authors:  K Rajkumar; S K Roy Chowdhury; Ramen Sinha
Journal:  J Maxillofac Oral Surg       Date:  2017-05-15

10.  Successful and failed mini-implants: microbiological evaluation and quantification of bacterial endotoxin.

Authors:  Marcela Cristina Damião Andrucioli; Mírian Aiko Nakane Matsumoto; Maria Conceição Pereira Saraiva; Magda Feres; Luciene Cristina de Figueiredo; Carlos Artério Sorgi; Lucia Helena Faccioli; Raquel Assed Bezerra da Silva; Lea Assed Bezerra da Silva; Paulo Nelson-Filho
Journal:  J Appl Oral Sci       Date:  2018-07-10       Impact factor: 2.698

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.