| Literature DB >> 23273317 |
Angelo Manna1, Carlo Caffarelli, Margherita Varini, Carlotta Povesi Dascola, Silvia Montella, Marco Maglione, Francesco Sperlì, Francesca Santamaria.
Abstract
Fractional exhaled nitric oxide (FeNO) is a non invasive method for assessing the inflammatory status of children with airway disease. Different ways to measure FeNO levels are currently available. The possibility of measuring FeNO levels in an office setting even in young children, and the commercial availability of portable devices, support the routine use of FeNO determination in the daily pediatric practice. Although many confounding factors may affect its measurement, FeNO is now widely used in the management of children with asthma, and seems to provide significantly higher diagnostic accuracy than lung function or bronchial challenge tests. The role of FeNO in airway infection (e.g. viral bronchiolitis and common acquired pneumonia), in bronchiectasis, or in cases with diffuse lung disease is less clear. This review focuses on the most recent advances and the current clinical applications of FeNO measurement in pediatric lung disease.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23273317 PMCID: PMC3545741 DOI: 10.1186/1824-7288-38-74
Source DB: PubMed Journal: Ital J Pediatr ISSN: 1720-8424 Impact factor: 2.638
FeNO reference values in healthy children and adolescents
| 37 | Caucasian, 4-18 years | 3.1 ppb | CLD 700; Eco Physics, Dürnten, Switzerland | |
| 159 | Caucasian, 6–15 years | 8.7 ppb | CLD 700, Al-Med, Ecophysics, Durnten, Switzerland | |
| 22 | Caucasian, 11-12 years | 4.0 ppb | LR 2000, Logan System, Rochester, Kent, UK | |
| 157 | Caucasian, 7-13 years | 7.4 ppb | NOA 280, Seivers Instruments Inc., Boulder, CO | |
| 107 | Caucasian, 4-18 years | 9.2 ppb | LR 2000, Logan Research, Rochester, UK | |
| 96 | Caucasian, 9-10 years | 13.3 ppb | Sievers Instruments, Boulder, CO, USA, kit bag collection | |
| 405 | Caucasian, Hispanic, Asian, Black, 4–17 years | 9.7 ppb | NIOX, Aerocrine AB, Stockholm, Sweden | |
| 291 | Chinese, 11-18 years | 19.9 ppb | NIOX, Aerocrine AB, Stockholm, Sweden | |
| | | Caucasian, 11-18 years | 12.7 ppb | |
| 114 | Caucasian, 7.2-15.7 years | Mean values not reported; range 7-14 ppb | NIOX, Aerocrine AB, Stockholm, Sweden | |
| 657 | White, Black, Asian, 9.1-12.9 years | 12.7 ppb | CLD 88sp, Eco Medics AG; Durnten, Switzerland | |
| 693 | Asiatic, 5-18 years | 13.7 ppb | CLD 88sp, Eco Medics AG; Durnten, Switzerland |
*Expressed as mean values.
Diagnostic accuracy of FeNO measurement in patients with asthma
| Children, 3-7 years | 83 (25) | > 9.7 ppb | 86 | 92 | |
| Adults | 160 (66) | > 16 ppb | 49 | 90 | |
| Adults | 40 (47) | > 7 ppb | 82 | 89 | |
| Children, 12-18 years | 106 (70) | > 19 ppb | 86 | 89 | |
| Children, 10-12 years | 60 (45) | > 20.4 ppb | 41 | 97 | |
| Children, 6-14 years | 57 (40) | > 19 ppb | 91 | 87 | |
| Children, 5-18 years | 70 (4.5) | > 28 ppb | 64 | 70 | |
| Adults, Children | 42 (30) | > 27 ppb | 78 | 92 | |
| Children, 8-16 years | 167 (68) | > 22 ppb | 57 | 87 |