| Literature DB >> 16626491 |
K Alving1, C Janson, L Nordvall.
Abstract
BACKGROUND: Exhaled nitric oxide (NO) measurement has been shown to be a valuable tool in the management of patients with asthma. Up to now, most measurements have been done with stationary, chemiluminescence-based NO analysers, which are not suitable for the primary health care setting. A hand-held NO analyser which simplifies the measurement would be of value both in specialized and primary health care. In this study, the performance of a new electrochemical hand-held device for exhaled NO measurements (NIOX MINO) was compared with a standard stationary chemiluminescence unit (NIOX).Entities:
Mesh:
Substances:
Year: 2006 PMID: 16626491 PMCID: PMC1462993 DOI: 10.1186/1465-9921-7-67
Source DB: PubMed Journal: Respir Res ISSN: 1465-9921
Device comparison. Some characteristics that are not identical in the two devices are given. Most other characteristics are similar, for example NO scrubbing of inhaled air and the exhalation flow control.
| NIOX | NIOX MINO | |
| Measurement range (ppb) | 1.5–200 | 5–300 |
| Accuracy | < 50 ppb: ± 2.5 ppb | < 50 ppb: ± 5 ppb |
| Visual feedback | Computer screen | Display via mirror |
| Calibration on site | Yes | No |
| Dimensions (cm) | 50 × 30 × 40 | 24 × 13 × 10 |
| Weight (kg) | 40 | 0.8 |
Figure 1Appearance of devices. Illustrations of (A) the NIOX and (B) the NIOX MINO. Note that the relative size is not proportional (see Table 1 for device dimensions).
Success rate. Fraction of patients (%) that were able to achieve three accepted measurements out of at most 6 attempts under guidance (clinical setting), or one accepted measurement out of at most 3 attempts without guidance (simulated home use).
| Mode | Clinical setting | Home use | |
| Instrument | NIOX | NIOX MINO | NIOX MINO |
| All subjects (n = 71) | 94 | 92 | 92 |
| Children (n = 37) | 95 | 84 | 84 |
| Adults (n = 34) | 94 | 100 | 100 |
Figure 2Device agreement. Scatter plots of the mean of three valid FENO measurements using the NIOX vs (A) the mean of three valid FENO measurements, or vs (B) the first valid FENO measurement using the NIOX MINO (n = 61). Repeatability coefficients (ICCs) were (A) 0.97 and (B) 0.98, respectively (p < 0.001 for both).
Figure 3Device agreement. Bland-Altman plots of the mean of three valid FENO measurements using the NIOX vs (A) the mean of three valid FENO measurements, or vs (B) the first valid FENO measurements using the NIOX MINO (n = 61).
Figure 4Repeatability. Histograms of the intra-subject SD of three valid FENO measurements using (A) the NIOX (n = 63) and (B) the NIOX MINO (n = 62). The estimated distribution of SD is indicated with the line.