Measurement of exhaled nitric oxide: comparison of different analysers.

Stephanie Korn1, Ina Telke, Oliver Kornmann, Roland Buhl.   

Abstract

BACKGROUND AND
OBJECTIVE: Exhaled nitric oxide (NO) is used as a surrogate marker to monitor eosinophilic airway inflammation, assist in diagnosis, and support treatment decisions for asthma patients. The aim of this study was to compare five NO analysers: Medisoft (M), Aerocrine Niox (N), Aerocrine Niox flex (NF), Aerocrino Niox mino (NM) and EcoMedics (E).
METHODS: In 85 subjects (25 asthma patients, 25 COPD patients, 35 healthy volunteers; median age 36.5 years, range 23-79, 47% female), three NO measurements per individual were performed, using each analyser according to the American Thoracic Society/European Respiratory Society guidelines. Subjects evaluated the devices on the basis of comfort and ease of performing the measurement.
RESULTS: Median NO concentrations were 24.8 parts per billion (ppb) (range 6.3-262.7) for M, 14.5 ppb (0.0-196.6) for N, 15.2 ppb (5.6-67.7) for NF, 16.3 ppb (5.0-208.3) for NM, and 13.1 ppb (4.0-103.0) for E. There was significant correlation among the absolute NO values measured with all the devices (0.743 for M vs NF < r < 0.979 for N vs NF). Bland-Altman plots suggested an acceptable degree of agreement among the results obtained with the three Niox analysers. Measurements obtained with the Medisoft and Ecomedics analysers were not directly comparable with those obtained with the other devices. The greatest differences in absolute NO levels for individual patients were between measurements with the M and E analysers (8.3 ppb, range 27.4-159.7, P < 0.001). Acceptance of the measurements by patients was high, independent of the device used.
CONCLUSIONS: Exhaled NO values obtained with different devices were not directly comparable and may differ to a clinically relevant extent, depending on which device is used.
© 2010 The Authors; Respirology © 2010 Asian Pacific Society of Respirology.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20920124     DOI: 10.1111/j.1440-1843.2010.01847.x

Source DB:  PubMed          Journal:  Respirology        ISSN: 1323-7799            Impact factor:   6.424


  8 in total

1.  Measurement of Fractional Exhaled Nitric Oxide: Comparison of Three Different Analysers.

Authors:  Stephanie Korn; Maike Wilk; Stefanie Voigt; Stephan Weber; Thomas Keller; Roland Buhl
Journal:  Respiration       Date:  2019-07-09       Impact factor: 3.580

Review 2.  Noninvasive effects measurements for air pollution human studies: methods, analysis, and implications.

Authors:  Jaime Mirowsky; Terry Gordon
Journal:  J Expo Sci Environ Epidemiol       Date:  2015-01-21       Impact factor: 5.563

Review 3.  An Evidence-Based Review of Application Devices for Nitric Oxide Concentration Determination from Exhaled Air in the Diagnosis of Inflammation and Treatment Monitoring.

Authors:  Magdalena Wyszyńska; Monika Nitsze-Wierzba; Aleksandra Czelakowska; Jacek Kasperski; Joanna Żywiec; Małgorzata Skucha-Nowak
Journal:  Molecules       Date:  2022-07-03       Impact factor: 4.927

Review 4.  Clinical application of exhaled nitric oxide measurement in pediatric lung diseases.

Authors:  Angelo Manna; Carlo Caffarelli; Margherita Varini; Carlotta Povesi Dascola; Silvia Montella; Marco Maglione; Francesco Sperlì; Francesca Santamaria
Journal:  Ital J Pediatr       Date:  2012-12-31       Impact factor: 2.638

Review 5.  Asthma in children and adolescents: a comprehensive approach to diagnosis and management.

Authors:  Christopher Chang
Journal:  Clin Rev Allergy Immunol       Date:  2012-08       Impact factor: 8.667

Review 6.  Recent Advances in Phthalocyanine and Porphyrin-Based Materials as Active Layers for Nitric Oxide Chemical Sensors.

Authors:  Darya Klyamer; Roman Shutilov; Tamara Basova
Journal:  Sensors (Basel)       Date:  2022-01-24       Impact factor: 3.576

7.  Fraction of exhaled nitric oxide (FeNO ) norms in healthy Tunisian adults.

Authors:  Sonia Rouatbi; Mohamed Ali Chouchene; Ines Sfaxi; Mohamed Ben Rejeb; Zouhair Tabka; Helmi Ben Saad
Journal:  Biomed Res Int       Date:  2014-06-03       Impact factor: 3.411

Review 8.  Fractional exhaled nitric oxide-measuring devices: technology update.

Authors:  Mauro Maniscalco; Carolina Vitale; Alessandro Vatrella; Antonio Molino; Andrea Bianco; Gennaro Mazzarella
Journal:  Med Devices (Auckl)       Date:  2016-06-23
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.