Biometric measurements in highly myopic eyes.

Peiyang Shen1, Yingfeng Zheng, Xiaohu Ding, Bin Liu, Nathan Congdon, Ian Morgan, Mingguang He.   

Abstract

PURPOSE: To assess the repeatability and accuracy of optical biometry (Lenstar LS900 optical low-coherence reflectometry [OLCR] and IOLMaster partial coherence interferometry [PCI]) and applanation ultrasound biometry in highly myopic eyes.
SETTING: Division of Preventive Ophthalmology, Zhongshan Ophthalmic Center, Guangzhou, China.
DESIGN: Comparative evaluation of diagnostic technology.
METHODS: Biometric measurements were taken in highly myopic subjects with a spherical equivalent (SE) of -6.00 diopters (D) or higher and an axial length (AL) longer than 25.0 mm. Measurements of AL and anterior chamber depth (ACD) obtained by OLCR were compared with those obtained by PCI and applanation A-scan ultrasound. Right eyes were analyzed. Repeatability was evaluated using the coefficient of variation (CoV) and agreement, using Bland-Altman analyses.
RESULTS: The mean SE was -11.20 D ± 4.65 (SD). The CoVs for repeated AL measurements using OLCR, PCI, and applanation ultrasound were 0.06%, 0.07%, and 0.20%, respectively. The limits of agreement (LoA) for AL were 0.11 mm between OLCR and PCI, 1.01 mm between OLCR and applanation ultrasound, and 1.03 mm between PCI and ultrasound. The ACD values were 0.29 mm, 0.53 mm, and 0.51 mm, respectively. These repeatability and agreement results were comparable in eyes with extreme myopia (AL ≥ 27.0 mm) or posterior staphyloma. The mean radius of corneal curvature was similar between OLCR and PCI (7.66 ± 0.24 mm versus 7.64 ± 0.25 mm), with an LoA of 0.12 mm.
CONCLUSION: Optical biometry provided more repeatable and precise measurements of biometric parameters, including AL and ACD, than applanation ultrasound biometry in highly myopic eyes. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2012 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23228592     DOI: 10.1016/j.jcrs.2012.08.064

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  7 in total

1.  Comparison of optical biometry and applanation ultrasound measurements of the axial length of the eye.

Authors:  Fouad R Nakhli
Journal:  Saudi J Ophthalmol       Date:  2014-05-06

2.  Repeatability of ocular biometry with IOLMaster 700 in subjects with clear lens.

Authors:  Manuel Garza-Leon; Hugo A Fuentes-de la Fuente; Ana V García-Treviño
Journal:  Int Ophthalmol       Date:  2016-10-21       Impact factor: 2.031

3.  Comparison axial length measurements from three biometric instruments in high myopia.

Authors:  Xiao-Gang Wang; Jing Dong; Yu-Lan Pu; Hui-Jun Liu; Qiang Wu
Journal:  Int J Ophthalmol       Date:  2016-06-18       Impact factor: 1.779

4.  Outcomes of cataract operations in extreme high axial myopia.

Authors:  Jasmine K M Lam; Tommy C Y Chan; Alex L K Ng; Vanissa W S Chow; Victoria W Y Wong; Vishal Jhanji
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-06-16       Impact factor: 3.117

5.  Comparison of anterior segment measurements obtained by three different devices in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-06-02       Impact factor: 3.411

6.  Effectiveness of combined macular buckle under direct vision and vitrectomy with ILM peeling in refractory macular hole retinal detachment with extreme high axial myopia: a 24-month comparative study.

Authors:  Jin Ma; Honghui Li; Xiaohu Ding; Silvia Tanumiharjo; Lin Lu
Journal:  Br J Ophthalmol       Date:  2017-03-14       Impact factor: 4.638

7.  Distribution of axial length, anterior chamber depth, and corneal curvature in an aged population in South China.

Authors:  Hui Chen; Haotian Lin; Zhuoling Lin; Jingjing Chen; Weirong Chen
Journal:  BMC Ophthalmol       Date:  2016-05-01       Impact factor: 2.209

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.