James J Lewis1, Violet N Chihota, Minty van der Meulen, P Bernard Fourie, Katherine L Fielding, Alison D Grant, Susan E Dorman, Gavin J Churchyard.
Abstract
BACKGROUND: "TBDx" is an innovative smear microscopy system that automatically loads slides onto a microscope, focuses and digitally captures images and then classifies smears as positive or negative using computerised algorithms.Entities:
Mesh:
Year: 2012 PMID: 23209666 PMCID: PMC3510232 DOI: 10.1371/journal.pone.0050173
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1The TBDx system, including microscope, slide rack and computer system.
Label 1: Prior 200 Slide Loader, with 4 slide cassettes containing 50 slides each. Label 2: Olympus Microscope, Olympus Camera and Prior Automated Slide Stage. Label 3: Joystick for manual stage movement. Label 4: Computer running TBDx integration, detection, and reporting software.
Frequency and percentage distributions of microscopist's smear status by culture results in the 981 specimens.
| Microscopist reading | Culture positive for | Culture negative (n = 712) |
| Smear negative | 127 (47.2%) | 702 (98.6%) |
| Scanty positive | 8 (3.0%) | 9 (1.3%) |
| 1+ positive | 33 (12.3%) | 1 (0.1%) |
| 2+ positive | 32 (11.9%) | 0 |
| 3+ positive | 69 (25.7%) | 0 |
The sensitivity, specificity, PPV and NPV of various microscopy methods, using culture as the gold standard.
| Sensitivity | Specificity | PPV | NPV | FoV reviewed by microscopist | |
|
| |||||
| Microscopist | 52.8% | 98.6% | 93.4% | 84.7% | 98,100 |
| TBDx | 75.8% | 43.5% | 33.7% | 82.7% | 0 |
| TBDx, with microscopist review of digitised images from TBDx scanty smears | 61.7% | 72.6% | 46.0% | 83.4% | 1,092 |
|
| |||||
| A: TBDx, with microscopist review of digitised images from TBDx scanty smears (3–9 AFB = TBDx scanty) | 48.0% | 91.4% | 67.9% | 82.3% | 597 |
| B: TBDx, with microscopist's original reading of smears classified as TBDx scanty | |||||
| (1–9 AFB = TBDx scanty) | 49.4% | 98.9% | 94.3% | 83.8% | 52,000 |
| (2–9 AFB = TBDx scanty) | 45.0% | 99.2% | 95.3% | 82.7% | 27,500 |
| (3–9 AFB = TBDx scanty) | 42.0% | 99.2% | 95.0% | 81.9% | 14,900 |
PPV = positive predictive value; NPV = negative predictive value; FoV = fields of view.
Assumed that all smears read by the microscopist had 100 FoV reviewed per slide, as data on exact numbers read for positive smears were not recorded. Exact numbers of digital images of FoV reviewed were recorded and so these are exact.
Figure 2Receiver operating characteristic curve showing the influence of different cut-offs to define positivity with TBDx (circles) and contrasted to the research microscopist (diamond).
AFB = acid-fast bacilli.