Literature DB >> 22247274

Differences in encapsulating lead tissue in patients who underwent transvenous lead removal.

Agnieszka Kołodzińska1, Andrzej Kutarski, Łukasz Koperski, Marcin Grabowski, Barbara Małecka, Grzegorz Opolski.   

Abstract

AIM: The aim of our study was to characterize specific tissue reaction of encapsulating lead tissue in patients who underwent transvenous lead removal and evaluate condition of the outer leads' insulation. METHODS AND
RESULTS: Fifty-six leads (27 atrial, 24 ventricular, 5 implantable cardioverter-defibrillators) were removed from 31 patients (mean age 70 years). Indications for removal were chronic pocket infection (CPI) (9 patients ), infective endocarditis (IE) (6), and non-infective indications (NI) (16). Leads with their surrounding tissue were fixed in paraformaldehyde. Tissues were embedded in paraffin wax, stained with haematoxylin-eosin, and examined histologically. The outer leads' insulations were examined in stereomicroscope. The mean lead age encapsulated by connective tissue sheath was 89, whereas encapsulated by granulation tissue was 47.34 months (P= 0.03). Calcification was present in 13 patients. Haemosiderine was observed only in patients with severe abrasion with perforation in the pocket (P= 0.04). Vasculogenesis was present in one (6%) with NI, five (56%) with CPI, and three (50%) with IE (P= 0.02) and was associated with elevated white blood cells (WBC) (P= 0.04). Eosinophilia was associated with elevated WBC (P= 0.04). The most frequently observed are third level of degradation (severe with perforation) in the intracardiac part of the silicone leads. Insulation damage due to environmental stress cracking concerned all leads with polyurethane overlay.
CONCLUSIONS: Granulation in encapsulating tissue was present in patients with younger leads. Vasculogenesis was observed more often in IE and CPI patients, which might indirectly indicate thickness of the sheath. Eosinophila may indicate allergic component of inflammation. Insulation damage frequently concerned the intracardiac part.

Entities:  

Mesh:

Year:  2012        PMID: 22247274     DOI: 10.1093/europace/eur384

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  7 in total

1.  Extraction of defibrillator leads recalled for cable externalization and failure.

Authors:  Divyang Patel; Evan Adelstein; Jan Nemec; G Stuart Mendenhall; Raveen Bazaz; Sandeep Jain; Samir Saba
Journal:  J Interv Card Electrophysiol       Date:  2012-11-21       Impact factor: 1.900

2.  Strategies to increase the INGEVITY lead strength during lead extraction procedures based on laboratory bench testing.

Authors:  Pierce Vatterott; Andrew De Kock; Eric F Hammill; Robert Lewis
Journal:  Pacing Clin Electrophysiol       Date:  2021-07-21       Impact factor: 1.912

3.  High shocking and pacing impedances due to defibrillation lead calcification.

Authors:  Robert G Hauser; Jay Sengupta; Susan Casey; Chuen Tang; Larissa I Stanberry; Raed Abdelhadi
Journal:  J Interv Card Electrophysiol       Date:  2019-12-18       Impact factor: 1.900

4.  Quantitative and Qualitative Assessment of Adhesive Thrombo-Fibrotic Lead Encapsulations (TFLE) of Pacemaker and ICD Leads in Arrhythmia Patients-A Post Mortem Study.

Authors:  Jonas Keiler; Marko Schulze; Ronja Dreger; Armin Springer; Alper Öner; Andreas Wree
Journal:  Front Cardiovasc Med       Date:  2020-11-30

5.  Early, Delayed and Late Cardiac Implantable Electronic Device Infections: Do the Timing of Onset and Pathogens Matter?

Authors:  Anna Polewczyk; Wojciech Jacheć; Maciej Polewczyk; Dorota Szczęśniak-Stańczyk; Andrzej Kutarski
Journal:  J Clin Med       Date:  2022-07-06       Impact factor: 4.964

6.  Prevalence and Predisposing Factors of Non-infectious Cardiac Implantable Electronic Device Lead Masses as Incidental Finding During Transoesophageal Echocardiography: A Retrospective Cohort Study.

Authors:  Tanja Kuecken; Ruta Jasaityte; Cara Bülow; Jessica Gross; Anja Haase-Fielitz; Michael Neuss; Christian Butter
Journal:  Front Cardiovasc Med       Date:  2022-06-14

7.  "Ghost", a Well-Known but Not Fully Explained Echocardiographic Finding during Transvenous Lead Extraction: Clinical Significance.

Authors:  Dorota Nowosielecka; Wojciech Jacheć; Anna Polewczyk; Łukasz Tułecki; Paweł Stefańczyk; Andrzej Kutarski
Journal:  Int J Environ Res Public Health       Date:  2022-10-01       Impact factor: 4.614