Literature DB >> 23149504

Implant primary stability determined by resonance frequency analysis: correlation with insertion torque, histologic bone volume, and torsional stability at 6 weeks.

Christian Makary1, Alberto Rebaudi, Gilberto Sammartino, Nada Naaman.   

Abstract

OBJECTIVE: To determine if resonance frequency analysis (RFA) correlated with insertion torque (IT), bone volume (BV), and clinical osseointegration at 6 weeks.
MATERIALS AND METHODS: For 18 patients, BV was evaluated histologically by retrieving bone core biopsies before placement of 40 dental implants. Peak IT was recorded at implant placement, and RFA values (implant stability quotient [ISQ]) were noted at baseline, 3 weeks, and 6 weeks. Osseointegration was evaluated at 6 weeks when torquing abutments to 30 N. ISQ values were correlated with IT, BV and abutment torquing results. Data were statistically analyzed.
RESULTS: ISQ values significantly decreased at 3 weeks and increased at 6 weeks. There was a significant positive correlation between BV and ISQ at baseline and at 3 weeks, but not at 6 weeks, and between ISQ and jaw location, implant diameter and IT at baseline, 3 weeks, and 6 weeks. There was a significant correlation between spinning/painful implants during abutment torquing and low ISQ, low BV, and low IT values.
CONCLUSION: Correlations between BV and IT values, and ISQ suggest that RFA may indicate primary implant stability. BV, IT, and ISQ values may anticipate the degree of osseointegration at 6 weeks.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23149504     DOI: 10.1097/ID.0b013e31826918f1

Source DB:  PubMed          Journal:  Implant Dent        ISSN: 1056-6163            Impact factor:   2.454


  12 in total

1.  Effects on the torsional vibration behavior in the investigation of dental implant osseointegration using resonance frequency analysis: a numerical approach.

Authors:  Min Zhai; Bing Li; Dehua Li
Journal:  Med Biol Eng Comput       Date:  2017-02-07       Impact factor: 2.602

2.  Measurements of Implant Stability Following Sinus Lift: A Pilot Clinical Study.

Authors:  Muhamed Ajanović; Alma Kamber-Ćesir; Adis Hamzić; Selma Tosum
Journal:  Acta Stomatol Croat       Date:  2015-09

3.  The effects of elevated hemoglobin A(1c) in patients with type 2 diabetes mellitus on dental implants: Survival and stability at one year.

Authors:  Thomas W Oates; Patrick Galloway; Peggy Alexander; Adriana Vargas Green; Guy Huynh-Ba; Jocelyn Feine; C Alex McMahan
Journal:  J Am Dent Assoc       Date:  2014-12       Impact factor: 3.634

4.  Immediate and early loading of hydrothermally treated, hydroxyapatite-coated dental implants: a 7-year prospective randomized clinical study.

Authors:  Afarin Arghami; David Simmons; Jeanne St Germain; Pooja Maney
Journal:  Int J Implant Dent       Date:  2021-03-10

5.  Prospective randomized clinical trial of hydrophilic tapered implant placement at maxillary posterior area: 6 weeks and 12 weeks loading.

Authors:  Seong-Beom Kim; Pil-Young Yun; Sang-Yun Kim; Yang-Jin Yi; Ji-Yun Kim; Young-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2016-10-21       Impact factor: 1.904

6.  Resonance frequency analysis of dental implants placed at the posterior maxilla varying the surface treatment only: A randomized clinical trial.

Authors:  Marcelo M Novellino; Newton Sesma; Piero R Zanardi; Dalva C Laganá
Journal:  Clin Implant Dent Relat Res       Date:  2017-06-20       Impact factor: 3.932

7.  Repeatability of the resonance frequency analysis values in implants with a new technology.

Authors:  M-C Díaz-Castro; A Falcao; P López-Jarana; C Falcao; J-V Ríos-Santos; A Fernández-Palacín; M Herrero-Climent
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2019-09-01

8.  The evaluation of implant stability measured by resonance frequency analysis in different bone types.

Authors:  Naser Sargolzaie; Sarah Samizade; Hamidreza Arab; Habibollah Ghanbari; Leila Khodadadifard; Amin Khajavi
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2019-02-26

9.  Study to assess the relationship between insertion torque value and implant stability quotient and its influence on timing of functional implant loading.

Authors:  Hasan Sarfaraz; Shefali Johri; P Sucheta; Sripathi Rao
Journal:  J Indian Prosthodont Soc       Date:  2018 Apr-Jun

10.  Influence of Implant Thread Morphology on Primary Stability: A Prospective Clinical Study.

Authors:  Maria Menini; Francesco Bagnasco; Ivan Calimodio; Nicolò Di Tullio; Francesca Delucchi; Domenico Baldi; Francesco Pera
Journal:  Biomed Res Int       Date:  2020-08-05       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.