| Literature DB >> 23138348 |
Stephen P Perfetto1, David Ambrozak, Richard Nguyen, Pratip K Chattopadhyay, Mario Roederer.
Abstract
The quality assurance program presented here provides a means to maximize and maintain the performance of individual flow cytometers in a facility. To optimize performance, we recommend performing all three steps (optimization, calibration and standardization) in this program when a new flow cytometer is installed or whenever the flow cytometer's optical path is altered (e.g., lasers, filters or detectors are replaced). The complete process requires 3-4 h. On a more frequent basis, only a subset of these procedures need to be performed as a part of daily maintenance routines. The data generated can be tracked to monitor the instrument and determine whether service is needed. In addition, the data can provide a metric for whether repairs and upgrades have improved or harmed performance, and for future instrument-to-instrument comparisons. In sum, the procedures presented here represent an updated framework for optimizing, calibrating and standardizing a flow cytometer for daily use.Mesh:
Substances:
Year: 2012 PMID: 23138348 DOI: 10.1038/nprot.2012.126
Source DB: PubMed Journal: Nat Protoc ISSN: 1750-2799 Impact factor: 13.491