Implementation and experimental results of 4D tumor tracking using robotic couch.

I Buzurovic1, Y Yu, M Werner-Wasik, T Biswas, P R Anne, A P Dicker, T K Podder.   

Abstract

PURPOSE: This study presents the implementation and experimental results of a novel technique for 4D tumor tracking using a commercially available and commonly used treatment couch and evaluates the tumor tracking accuracy in clinical settings.
METHODS: Commercially available couch is capable of positioning the patient accurately; however, currently there is no provision for compensating physiological movement using the treatment couch in real-time. In this paper, a real-time couch tracking control technique is presented together with experimental results in tumor motion compensation in four dimensions (superior-inferior, lateral, anterior-posterior, and time). To implement real-time couch motion for tracking, a novel control system for the treatment couch was developed. The primary functional requirements for this novel technique were: (a) the treatment couch should maintain all previous∕normal features for patient setup and positioning, (b) the new control system should be used as a parallel system when tumor tracking would be deployed, and (c) tracking could be performed in a single direction and∕or concurrently in all three directions of the couch motion (longitudinal, lateral, and vertical). To the authors' best knowledge, the implementation of such technique to a regular treatment couch for tumor tracking has not been reported so far. To evaluate the performance of the tracking couch, we investigated the mechanical characteristics of the system such as system positioning resolution, repeatability, accuracy, and tracking performance. Performance of the tracking system was evaluated using dosimetric test as an endpoint. To investigate the accuracy of real-time tracking in the clinical setting, the existing clinical treatment couch was replaced with our experimental couch and the linear accelerator was used to deliver 3D conformal radiation therapy (3D-CRT) and intensity modulated radiation therapy (IMRT) treatment plans with and without tracking. The results of radiation dose distribution from these two sets of experiments were compared and presented here.
RESULTS: The mechanical accuracies were 0.12, 0.14, and 0.18 mm in X, Y, and Z directions. The repeatability of the desired motion was within ±0.2 mm. The differences of central axis dose between the 3D-CRT stationary plan and two tracking plans with different motion trajectories were 0.21% and 1.19%. The absolute dose differences of both 3D tracking plans comparing to the stationary plan were 1.09% and 1.20%. Comparing the stationary IMRT plan with the tracking IMRT plan, it was observed that the central axis dose difference was -0.87% and the absolute difference of both IMRT plans was 0.55%.
CONCLUSIONS: The experimental results revealed that the treatment couch could be successfully used for real-time tumor tracking with a high level of accuracy. It was demonstrated that 4D tumor tracking was feasible using existing couch with implementation of appropriate tracking methodology and with modifications in the control system.

Entities:  

Mesh:

Year:  2012        PMID: 23127089      PMCID: PMC3494731          DOI: 10.1118/1.4758064

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  22 in total

1.  Issues in respiratory motion compensation during external-beam radiotherapy.

Authors:  Cihat Ozhasoglu; Martin J Murphy
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-04-01       Impact factor: 7.038

2.  DMLC motion tracking of moving targets for intensity modulated arc therapy treatment: a feasibility study.

Authors:  Jens Zimmerman; Stine Korreman; Gitte Persson; Herb Cattell; Michelle Svatos; Amit Sawant; Raghu Venkat; David Carlson; Paul Keall
Journal:  Acta Oncol       Date:  2009       Impact factor: 4.089

3.  An analysis of the treatment couch and control system dynamics for respiration-induced motion compensation.

Authors:  Warren D D'Souza; Thomas J McAvoy
Journal:  Med Phys       Date:  2006-12       Impact factor: 4.071

4.  Predicting respiratory tumor motion with multi-dimensional adaptive filters and support vector regression.

Authors:  Nadeem Riaz; Piyush Shanker; Rodney Wiersma; Olafur Gudmundsson; Weihua Mao; Bernard Widrow; Lei Xing
Journal:  Phys Med Biol       Date:  2009-09-04       Impact factor: 3.609

5.  Speed and amplitude of lung tumor motion precisely detected in four-dimensional setup and in real-time tumor-tracking radiotherapy.

Authors:  Hiroki Shirato; Keishiro Suzuki; Gregory C Sharp; Katsuhisa Fujita; Rikiya Onimaru; Masaharu Fujino; Norio Kato; Yasuhiro Osaka; Rumiko Kinoshita; Hiroshi Taguchi; Shunsuke Onodera; Kazuo Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-03-15       Impact factor: 7.038

6.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

7.  Active Tracking and Dynamic Dose Delivery for robotic couch in radiation therapy.

Authors:  Ivan Buzurovic; Yan Yu; Tarun K Podder
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2011

8.  Fluoroscopic tumor tracking for image-guided lung cancer radiotherapy.

Authors:  Tong Lin; Laura I Cerviño; Xiaoli Tang; Nuno Vasconcelos; Steve B Jiang
Journal:  Phys Med Biol       Date:  2009-01-16       Impact factor: 3.609

9.  Dosimetric advantages of four-dimensional adaptive image-guided radiotherapy for lung tumors using online cone-beam computed tomography.

Authors:  Asif Harsolia; Geoffrey D Hugo; Larry L Kestin; Inga S Grills; Di Yan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-02-01       Impact factor: 7.038

10.  Compensating for tumor motion by a 6-degree-of-freedom treatment couch: is patient tolerance an issue?

Authors:  Reinhart A Sweeney; Winfried Arnold; Eva Steixner; Meinhard Nevinny-Stickel; Peter Lukas
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-01-23       Impact factor: 7.038

View more
  5 in total

1.  Development of a four-axis moving phantom for patient-specific QA of surrogate signal-based tracking IMRT.

Authors:  Nobutaka Mukumoto; Mitsuhiro Nakamura; Masahiro Yamada; Kunio Takahashi; Mami Akimoto; Yuki Miyabe; Kenji Yokota; Shuji Kaneko; Akira Nakamura; Satoshi Itasaka; Yukinori Matsuo; Takashi Mizowaki; Masaki Kokubo; Masahiro Hiraoka
Journal:  Med Phys       Date:  2016-12       Impact factor: 4.071

2.  Feasibility study for marker-based VMAT plan optimization toward tumor tracking.

Authors:  Azeez A Omotayo; Sankar Venkataraman; Niranjan Venugopal; Boyd McCurdy
Journal:  J Appl Clin Med Phys       Date:  2020-06-11       Impact factor: 2.102

3.  The ideal couch tracking system-Requirements and evaluation of current systems.

Authors:  Alexander Jöhl; Stefanie Ehrbar; Matthias Guckenberger; Stephan Klöck; Andreas Mack; Mirko Meboldt; Melanie Zeilinger; Stephanie Tanadini-Lang; Marianne Schmid Daners
Journal:  J Appl Clin Med Phys       Date:  2019-09-19       Impact factor: 2.102

4.  SlicerArduino: A Bridge between Medical Imaging Platform and Microcontroller.

Authors:  Paolo Zaffino; Alessio Merola; Domenico Leuzzi; Virgilio Sabatino; Carlo Cosentino; Maria Francesca Spadea
Journal:  Bioengineering (Basel)       Date:  2020-09-11

5.  A respiratory compensating system: design and performance evaluation.

Authors:  Ho-Chiao Chuang; Ding-Yang Huang; Der-Chi Tien; Ren-Hong Wu; Chung-Hsien Hsu
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.