Jayashree Kharge1, Praveen Sreekumar, Kumara Swamy, Ashalatha Bharatha, Raghu Thagachagere Ramegowda, Manjunath Cholenahally Nanjappa.
Abstract
The breakage of a stent-delivery catheter at the shaft is a rare and dangerous complication during coronary intervention. We report a simple balloon technique for the successful retrieval, from within a guiding catheter, of both an unexpanded stent and its delivery system.Keywords: Angioplasty, balloon, coronary/instrumentation; device removal/methods; equipment failure; stents/adverse outcomes
Mesh:
Year: 2012 PMID: 23109758 PMCID: PMC3461686