Major stent deformation/pseudofracture of 7 Crown Endeavor/Micro Driver stent platform: incidence and causative factors.

Mark Pitney1, Kate Pitney, Nigel Jepson, Daniel Friedman, Tuan Nguyen-Dang, Joseph Matthews, Robert Giles, David Taylor.   

Abstract

AIMS: This is the initial report of stent deformation/pseudofracture of the 7 crown Endeavor/Micro Driver stent platform (2.25-2.75 mm), whereby the post-dilation balloon catches and causes major stent deformation angiographically appearing as a large stent fracture. We sought to determine frequency and cause. METHODS AND
RESULTS: Analysis of 1,000 consecutive Endeavor/Micro Driver stents (7 crown) deployed at our institution. Bench testing was also performed by deploying 10 stents in 2.5 mm tubing and then attempting re-crossing with non-compliant balloons. CLINICAL
RESULTS: There were 14 cases of major stent deformation/pseudofracture representing an incidence of 1.4% or 1.8% of 775 stents that were post-dilated. Of the 14 deformed stents, re-stenting was required in nine and MACE occurred in five (36%.). Benchtop results: Balloon "catch" was reproducible at an initial frequency of 9% and affected three of the 10 stents. With the use of provocative measures stent deformation occurred in 50% of test stents.
CONCLUSIONS: Stent deformation/pseudofracture of the 7 crown Endeavor/Micro Driver platform occurred in 1.8% of cases that were post-dilated. Of these, 36% experience MACE. Deformation can be reproduced on the benchtop. Operators should be aware of the complication, and develop strategies to deal with it.

Entities:  

Mesh:

Year:  2011        PMID: 21646069     DOI: 10.4244/EIJV7I2A41

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  10 in total

1.  Impact of stent mis-sizing and mis-positioning on coronary fluid wall shear and intramural stress.

Authors:  Henry Y Chen; Bon-Kwon Koo; Deepak L Bhatt; Ghassan S Kassab
Journal:  J Appl Physiol (1985)       Date:  2013-05-30

2.  Dislodged and mechanically distorted stent stuck within a previously implanted drug-eluting stent.

Authors:  Hiroaki Watabe; Akira Sato; Tomoya Hoshi; Kenichi Obara; Ryo Kawamura; Kazutaka Aonuma
Journal:  Heart Vessels       Date:  2012-09-20       Impact factor: 2.037

3.  Longitudinal deformation - price we pay for better deliverability of coronary stent platforms.

Authors:  Ezhilan Janakiraman; Vijayakumar Subban; Suma M Victor; Ajit S Mullasari
Journal:  Indian Heart J       Date:  2012-07-27

4.  Distal longitudinal deformation of a Synergy stent by jailed Rotawire guidewire.

Authors:  A M Leong; P J L Ong; H H Ho; T Watson
Journal:  Herz       Date:  2016-06-29       Impact factor: 1.443

5.  Longitudinal Deformation of Distal Edge in a New-Generation Stent Caused by Guidewire Entrapment.

Authors:  Adam Taleb; Gaurav Parikh; Nikolaos Kakouros
Journal:  Tex Heart Inst J       Date:  2018-02-01

6.  Clinical Impact of Stent Design.

Authors:  Rebecca L Noad; Colm G Hanratty; Simon J Walsh
Journal:  Interv Cardiol       Date:  2014-04

7.  Longitudinal Stent Deformation in Undeployed Stents: A Case Series.

Authors:  Parminder Singh Otaal; Dinakar Bootla
Journal:  J Tehran Heart Cent       Date:  2021-07

8.  Inelastic Deformation of Coronary Stents: Two-Level Model.

Authors:  Pavel S Volegov; Nikita A Knyazev; Roman M Gerasimov; Vadim V Silberschmidt
Journal:  Materials (Basel)       Date:  2022-10-07       Impact factor: 3.748

9.  Changes of plasma B-type natriuretic peptide levels after high-pressure post-dilation following coronary stent deployment.

Authors:  Gang-Yong Wu; Gang-Jun Zong; Jing-Kai Chen; Yang Xia; Man-Qing Chen; Xiao Wang; Xiao-Ying Wang; Lu-Lu Wang; Tian-Xiao Wang
Journal:  PLoS One       Date:  2013-12-06       Impact factor: 3.240

Review 10.  Long and short of optimal stent design.

Authors:  Timothy Watson; Mark W I Webster; John A Ormiston; Peter N Ruygrok; James T Stewart
Journal:  Open Heart       Date:  2017-10-30
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.