Literature DB >> 23083199

A multiplanar radiography method for assessing cup orientation in total hip arthroplasty.

Shahram Amiri1, Bassam A Masri, Donald Garbuz, Carolyn Anglin, David R Wilson.   

Abstract

Correct orientation of the acetabular cup considering patient-specific functional pelvic angles is an important factor for improving outcomes and avoiding complications after total hip arthroplasty. This study introduces a new, noninvasive radiographic tool for accurately determining a patient's specific pelvic tilt angle preoperatively, as well as accurately assessing acetabular cup orientation with respect to bony landmarks intraoperatively and postoperatively. The method was validated by imaging a bone replica model of the pelvis with implanted hip components, in comparison to digitized references, and verified with a cadaveric specimen. Pelvic tilt was measured with an accuracy of 0.1 deg and SD of 0.4 deg. Operative cup inclination and anteversion showed accuracies of 0.6 deg and 2.5 deg, with SD of 0.4 deg and 0.6 deg, respectively; these could be improved further by subtracting systematic bias. The method shows accuracy advantages over existing radiographic and fluoroscopic methods and exposes the subjects to a lower radiation dose compared to the similar computed tomography methods. These results suggest that the proposed method is feasible for assessing cup placement with reference to the functional and anatomical references. Furthermore, the ability to reference the same bony landmarks preoperatively, intraoperatively, and postoperatively has important research and clinical advantages.

Entities:  

Mesh:

Year:  2012        PMID: 23083199     DOI: 10.1115/1.4007664

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  4 in total

1.  A low-cost tracked C-arm (TC-arm) upgrade system for versatile quantitative intraoperative imaging.

Authors:  Shahram Amiri; David R Wilson; Bassam A Masri; Carolyn Anglin
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-12-10       Impact factor: 2.924

Review 2.  Radiographic assessment of the cup orientation after total hip arthroplasty: a literature review.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Ann Transl Med       Date:  2020-02

3.  Three-dimensional orientation and location-dependent varying rules of radiographic angles of the acetabular cup.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Int Orthop       Date:  2018-02-17       Impact factor: 3.075

4.  The synergetic effect of pelvic rotation and X-ray offset on radiographic angles of the acetabular cup.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Guo-Fei Sun; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Med Biol Eng Comput       Date:  2019-09-09       Impact factor: 2.602