| Literature DB >> 23074593 |
Alimohammad Hajizeinali1, Hakimeh Sadeghian, Mehrnaz Rezvanfard, Mohammad Alidoosti, Seyed Ebrahim Kassaian, Ebrahim Nematipour.
Abstract
BACKGROUND: Transcatheter occlusion of the patent ductus arteriosus (PDA) is a minimally invasive treatment. The appropriate device size is chosen based on the angiographic measurement of the PDA. The current study aimed to assess the relationship between the transthoracic echocardiographic (TTE) measurements of the PDA prior to the occlusion procedure and the actual size of the deployed device.Entities:
Keywords: Dactus arteriosus; Echocardiography; Septal occluder device; patent
Year: 2010 PMID: 23074593 PMCID: PMC3466856
Source DB: PubMed Journal: J Tehran Heart Cent ISSN: 1735-5370
Figure 1Visualization of patent ductus arteriosus (PDA) in parasternal short-axis view (A) and echocardiographic measurements of PDA in the same view (B)
PA, Pulmonary artery; DA, Descending aorta; AS, Diameter of PDA at aortic site; PS, Diameter of PDA at pulmonic site; L, Length of PDA
Demographic data and echocardiographic measurements of all patients according to the type of occluder device
| Subject | Device type | Demographic Data | Echocardiographic measurements | PDA measurement (mm) | Device diameter (mm) | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
| |||||||||||||
| Age (y) | Sex | LVEF (%) | LA size (mm) | LVDs (mm) | LVDd (mm) | PAP (mmHg) | PG(s) (mmHg) | PG(d) (mmHg) | PDA type | Length | Aortic width (mm) | Pulmonic width (mm) | |||
|
| |||||||||||||||
| 1 | Am | 22 | F | 60 | 39 | 31 | 47 | ND | ND | ND | B | 11 | 7 | 11 | 8 |
| 2 | Am | 7 | F | 60 | 26 | 27 | 38 | ND | 55 | 109 | B | 8 | 4 | 6 | 4 |
| 3 | Am | 46 | F | 55 | 33 | 44 | 61 | 41 | 59 | 74 | A | 13 | 8 | 6 | 8 |
| 4 | Am | 13 | F | 55 | 31 | 28 | 42 | 32 | 71 | 42 | A | 7 | 5 | 4 | 6 |
| 5 | Am | 18 | M | 60 | 23 | 35 | 45 | 25 | ND | ND | B | 14 | 7 | 8 | 6 |
| 6 | COF | 14 | M | 60 | 27 | 30 | 45 | 35 | 37 | 83 | A | 6 | 4 | 2 | 4 |
| 7 | COF | 27 | F | 60 | 39 | 36 | 52 | 27 | 63 | 89 | ND | ND | 6 | ND | 6 |
| Mean (SD) | 21.0 (12.8) | 58.6 (2.4) | 31.1 (6.2) | 33.0 (6.4) | 47.1 (7.5) | 32.0 (6.4) | 57.0 (12.6) | 79.4 (24.5) | 9.8 (3.3) | 6.0 (1.6) | 6.2 (3.1) | 6.0 (1.6) | |||
Echocardiographic measurement was assessed based on American society of echocardiography (ASE) guidelines and standards
PDA, Patent ducts arteriosus; LVEF, Left ventricular ejection fraction; LA, Left atrium; LVDs, Left ventricular dimension (systolic); LVDd, Left ventricular dimension (diastolic); PAP, Pulmonary artery pressure; PG(s), Peak gradient (systolic) across PDA; PG(d), Peak gradient (diastolic) across PDA; Am, Amplatzer; F, Female; ND, No data were available; M, Male; COF, Cardi-O-Fix