The effect of pupil dilation on biometric parameters of the Lenstar 900.

Berker Bakbak1, Bengu E Koktekir, Sansal Gedik, Huseyin Guzel.   

Abstract

PURPOSE: To evaluate the effect of pupil dilation on the biometric parameters of the Lenstar LS 900 (Haag Streit AG, Koeniz, Switzerland).
METHODS: In this cross-sectional study of 33 eyes with cataracts, axial length (AL), anterior chamber depth, intraocular lens (IOL) power, keratometry, and pupil diameter measurements were performed using Lenstar biometry before and after pupil dilatation. Intraobserver repeatability was assessed by taking 2 consecutive recordings of biometric parameters using Lenstar biometry in the undilated pupils of 30 eyes with cataracts.
RESULTS: The mean difference in AL measurements from intraobserver readings was -0.001 ± 0.01 mm. The study group showed a statistically significant change in the first keratometry reading and anterior chamber depth values (P < 0.05); however, the mean difference in AL and IOL power was statistically insignificant between the undilated and dilated pupil Lenstar biometry readings. Three cases in the study group (9.1%) and 1 case in the control group (3.3%) demonstrated changes in IOL power >0.50 diopter.
CONCLUSIONS: Dilated and undilated pupil size did not affect the measurement of IOL using the Lenstar biometry.

Entities:  

Mesh:

Year:  2013        PMID: 23023403     DOI: 10.1097/ICO.0b013e31825fee82

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  7 in total

1.  The effect of pupil dilation on AL-Scan biometric parameters.

Authors:  Ertuğrul Can; Mustafa Duran; Tuğba Çetinkaya; Nurşen Arıtürk
Journal:  Int Ophthalmol       Date:  2015-06-27       Impact factor: 2.031

2.  Effect of pupillary dilation on Haigis formula-calculated intraocular lens power measurement by using optical biometry.

Authors:  Bharkbhum Khambhiphant; Suganlaya Sasiwilasagorn; Nattida Chatbunchachai; Krit Pongpirul
Journal:  Clin Ophthalmol       Date:  2016-07-27

3.  Effect of Cycloplegia on Keratometric and Biometric Parameters in Keratoconus.

Authors:  Nihat Polat; Abuzer Gunduz
Journal:  J Ophthalmol       Date:  2016-12-12       Impact factor: 1.909

4.  Influence of pupil dilation on the Barrett universal II (new generation), Haigis (4th generation), and SRK/T (3rd generation) intraocular lens calculation formulas: a retrospective study.

Authors:  Takeshi Teshigawara; Akira Meguro; Nobuhisa Mizuki
Journal:  BMC Ophthalmol       Date:  2020-07-20       Impact factor: 2.209

5.  Influence of pupil dilation on predicted postoperative refraction and recommended IOL to obtain target postoperative refraction calculated by using third- and fourth-generation calculation formulas.

Authors:  Takeshi Teshigawara; Akira Meguro; Nobuhisa Mizuki
Journal:  Clin Ophthalmol       Date:  2018-10-02

6.  Effect of pupil dilation on biometric measurements and intraocular lens power calculations in schoolchildren.

Authors:  Xiaogang Wang; Jing Dong; Maolong Tang; Xiaoliang Wang; Haitao Wang; Suhua Zhang
Journal:  PLoS One       Date:  2018-09-13       Impact factor: 3.240

7.  Relationship between Myopia Progression and School Entrance Age: A 2.5-Year Longitudinal Study.

Authors:  Linjie Liu; Dandan Jiang; Chunchun Li; Yaoyao Lin; Wenzhe Zhou; Haishao Xiao; Yanyan Chen
Journal:  J Ophthalmol       Date:  2021-03-30       Impact factor: 1.909

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.