H Griessenberger1, D P J Heib, A B Kunz, K Hoedlmoser, M Schabus.
Abstract
PURPOSE: Classically, professional assessment of sleep is done in the sleep laboratory using whole-night polysomnography (PSG). However, given a misbalance between accredited sleep laboratories and the large amount of patients suffering from sleep disorders, only few receive appropriate diagnostic assessment. Recently, some low-cost home sleep scoring systems have been proposed, yet such systems are rarely tested scientifically. The aim of the present study was to evaluate the staging accuracy of the home sleep scoring system Zeo (Newton, MA, USA).Entities:
Mesh:
Year: 2012 PMID: 22996794 PMCID: PMC3655221 DOI: 10.1007/s11325-012-0757-4
Source DB: PubMed Journal: Sleep Breath ISSN: 1520-9512 Impact factor: 2.816
Percent agreement table for epoch labeling between Zeo and comparison scorings (Somnolyzer 24 × 7, Expert and SleepExplorer)
| Zeo | |||||
|---|---|---|---|---|---|
| Wake (%) | REM (%) | Light (%) | Deep (%) | ||
| Somnolyzer 24 × 7 | Wake | 40.83 | 26.79 | 29.24 | 3.14 |
| REM | 11.54 | 73.69 | 13.40 | 1.37 | |
| Light | 7.21 | 9.17 | 80.04 | 3.59 | |
| Deep | 1.30 | 0.08 | 36.48 | 62.13 | |
| Expert scorer | Wake | 40.05 | 24.03 | 30.96 | 4.96 |
| REM | 13.94 | 70.65 | 14.28 | 1.13 | |
| Light | 4.88 | 9.46 | 79.13 | 6.52 | |
| Deep | 0.48 | 0.67 | 29.94 | 68.91 | |
| SleepExplorer | Wake | 31.39 | 30.93 | 34.01 | 3.67 |
| REM | 13.39 | 67.77 | 18.00 | 0.83 | |
| Light | 2.89 | 4.62 | 84.93 | 7.56 | |
| Deep | 0.86 | 0.49 | 29.45 | 69.20 | |
Classification results between Somnolyzer 24 × 7 and comparison scorings (Zeo, Expert and SleepExplorer)
| Wake | REM | Light | Deep | |||
|---|---|---|---|---|---|---|
| Zeo | ||||||
| % OA | 72.6 | Sensitivity | 40.8 | 73.7 | 80.0 | 62.1 |
| Cohen’s kappa | 0.56 | PPV | 32.0 | 67.4 | 79.2 | 82.2 |
| Expert | ||||||
| % OA | 80.9 | Sensitivity | 70.1 | 80.1 | 86.4 | 69.8 |
| Cohen’s kappa | 0.69 | PPV | 53.1 | 81.6 | 82.4 | 93.7 |
| SleepExplorer | ||||||
| % OA | 74.2 | Sensitivity | 83.4 | 74.2 | 75.3 | 67.2 |
| Cohen’s kappa | 0.61 | PPV | 37.7 | 70.2 | 83.4 | 91.3 |
% OA percent overall agreement, sensitivity percentage of actual sleep state epochs that are labelled as sleep state epochs by classifier, PPV positive predictive value (percentage of epochs labelled as sleep state that are correctly labelled)
Fig. 1Bland–Altman plots of sleep parameters SOL and WASO showing differences between Somnolyzer 24 × 7 and a Zeo, b expert c SleepExplorer scorings. The x-axes indicate the average from both the study standard and comparison scoring of wake after sleep onset and sleep onset latency. The difference is expressed as the comparison score minus the study standard score. The mean difference and the limits of agreement (±1.96 SD) are represented as dashed lines. Note that high variabilities are dominant for Zeo scorings
Fig. 2Scatter plots depicting the agreement of Somnolyzer 24 × 7 with Zeo for SOL, WASO and REM (top to bottom). Note that the SOL agreement for Zeo and study standard shows high deviations. Dots are scattered around the 45° line of identity