Aime M Gloi1, Robert Buchanan.
Abstract
The aim of this study was to compare lumpectomy cavity depth measurements obtained through ultrasound (U/S) and retrospective computed tomography (CT). Twenty-five patients with stage T1-2 invasive breast cancer formed the cohort of this study. Their U/S and CT measurements were converted into electron energy and compared. The mean U/S depth was 3.6 ± 1.3 cm, while the mean CT depth was 4.9 ± 1.9 cm; the listed error ranges are one standard deviation. Electron energies for treatment ranged from 6 MeV to 12 MeV based on the U/S determination. There was no significant correlation between cavity depths measured by U/S and CT (R(2)= 0.459, P < 0.002). Furthermore, only 20% of CT-based electron energy determinations matched the corresponding U/S determinations. This ratio increased to 40% when taking into account an upper limit based on the depth of organs at risk below the cavity. The study shows that there is a significant discrepancy between cavity depths determined by U/S and CT. It also supports the concept that post-lumpectomy radiotherapy boosts should be tailored according to the needs and comfort of individual practices and institutions.Entities:
Keywords: Breast boost; cavity depth; lumpectomy
Year: 2012 PMID: 22973080 PMCID: PMC3437170 DOI: 10.4103/0971-6203.99234
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Lumpectomy depth measurement methods employed in various institutions
Figure 1Scatter plot of depth cavity from CT results against U/S for 25 patients. Pearson correlation was r=0.46 with P<0.0002
Patient and tumor characteristics employed in this study
Dosimetric parameters calculated from this study
Figure 2Difference between U/S and CT cavity depth measurements with horizontal lines showing the mean of limits of agreement
Figure 3Scatter plot of depth cavity from CT measurement versus OAR for 25 patients. The concordance correlation was estimated at P = 0.46
Figure 4Scatter plot of depth cavity from U/S measurement versus OAR for 25 patients. The concordance correlation was estimated at P = 0.37
Figure 5Scatter plot of depth cavity from U/S measurement versus breast volumefor 25 patients. The concordance correlation was estimated at P = 0.7
Figure 6Scatter plot of depth cavity from CT measurement versus breast volume for 25 patients. The concordance correlation was estimated at P = 0.31