Literature DB >> 22959525

Results of total shoulder arthroplasty with a monoblock porous tantalum glenoid component: a prospective minimum 2-year follow-up study.

Matthew D Budge1, Elizabeth M Nolan, Meredith H Heisey, Kevin Baker, J Michael Wiater.   

Abstract

BACKGROUND: Aseptic loosening of all-polyethylene glenoid components remains a limiting factor in achieving long-term implant survival in total shoulder arthroplasty (TSA). This study prospectively evaluated the functional and radiographic outcomes of patients undergoing TSA with a novel, porous, tantalum-backed glenoid component, with a minimum 2 years of follow-up.
MATERIALS AND METHODS: Porous tantalum-backed glenoid components were used in 19 TSAs in 19 patients. All patients were available for radiographic follow-up at an average of 38 months (range, 24-64 months). Patients were evaluated prospectively using the American Shoulder and Elbow Surgeons (ASES) score and pain on a visual analog scale (VAS). Radiographs were evaluated for component loosening and failure of the porous tantalum backing at a minimum 2 years of follow-up.
RESULTS: The mean VAS decreased from 8.6 to 2.9 (P < .0001). The mean ASES score improved from 21 to 70 points (P < .05). Mean active forward elevation improved from 75° to 131° (P < .0001). At latest follow-up, all glenoid components except 1 had complete in-growth of the porous tantalum keel; however, 4 components (21%) failed by fracture at the keel-glenoid face junction.
CONCLUSIONS: There was an unacceptably high rate of glenoid component failure (21%) due to fracture at the keel-glenoid face junction in this series. The manufacturer has subsequently revised this early design to reduce the risk of failure. The results of this study illustrate that caution should be exercised in the use of novel implants with an unproven clinical track record.
Copyright © 2013 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22959525     DOI: 10.1016/j.jse.2012.06.001

Source DB:  PubMed          Journal:  J Shoulder Elbow Surg        ISSN: 1058-2746            Impact factor:   3.019


  16 in total

1.  Tantalum coating of porous carbon scaffold supplemented with autologous bone marrow stromal stem cells for bone regeneration in vitro and in vivo.

Authors:  Xiaowei Wei; Dewei Zhao; Benjie Wang; Wei Wang; Kai Kang; Hui Xie; Baoyi Liu; Xiuzhi Zhang; Jinsong Zhang; Zhenming Yang
Journal:  Exp Biol Med (Maywood)       Date:  2016-02-02

2.  Soft tissue balancing in total shoulder replacement.

Authors:  Maike Mueller; Gregory Hoy
Journal:  Curr Rev Musculoskelet Med       Date:  2014-03

3.  [Glenoid replacement for omarthritis : indications, technique, results and new developments].

Authors:  M Pfahler
Journal:  Orthopade       Date:  2013-07       Impact factor: 1.087

4.  Management of complications after total shoulder arthroplasty.

Authors:  Josef K Eichinger; Joseph W Galvin
Journal:  Curr Rev Musculoskelet Med       Date:  2015-03

5.  Radiographic and clinical outcomes of second generation Trabecular Metal™ glenoid components in total shoulder arthroplasty.

Authors:  Raymond E Chen; Alexander M Brown; Alexander S Greenstein; Sandeep Mannava; Ilya Voloshin
Journal:  Shoulder Elbow       Date:  2020-03-24

6.  Porous metals and alternate bearing surfaces in shoulder arthroplasty.

Authors:  Shannon R Carpenter; Ivan Urits; Anand M Murthi
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

Review 7.  Modern trabecular metal-backed glenoid components in total shoulder arthroplasty: What is the evidence? A systematic review.

Authors:  Michael-Alexander Malahias; Dimitrios Chytas; Lazaros Kostretzis; Angelos Trellopoulos; Emmanouil Brilakis; Emmanouil Antonogiannakis
Journal:  Shoulder Elbow       Date:  2020-09-25

8.  Management of failed metal-backed glenoid component in patients with bilateral total shoulder arthroplasty.

Authors:  Xinning Li; Josef K Eichinger; Laurence D Higgins
Journal:  Int J Shoulder Surg       Date:  2013-10

9.  Catastrophic failure of a low profile metal-backed glenoid component after total shoulder arthroplasty.

Authors:  Carley B Vuillermin; Mark E Trump; Shane A Barwood; Gregory A Hoy
Journal:  Int J Shoulder Surg       Date:  2015 Oct-Dec

Review 10.  The Clinical Application of Porous Tantalum and Its New Development for Bone Tissue Engineering.

Authors:  Gan Huang; Shu-Ting Pan; Jia-Xuan Qiu
Journal:  Materials (Basel)       Date:  2021-05-18       Impact factor: 3.623

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.