| Literature DB >> 2295897 |
G Greenblott1, S J Barker, K K Tremper, S Gerschultz, J L Gehrich.
Abstract
In a recent study, we compared a new intraarterial fiberoptic "optode" probe to continuously measure arterial oxygen and carbon dioxide tensions and pH with intermittently drawn blood samples in patients undergoing surgery. In one patient with a diagnosis of Arnold-Chiari type I malformation with outflow obstruction of the fourth ventricle, a major pulmonary air embolism occurred while the patient was undergoing suboccipital craniectomy and cervical laminectomy in the prone position. Three hours after the incision the optode-displayed oxygen tension decreased from a stable value of 225 +/- 8 mm Hg to 63 mm Hg over a 10-minute period. During the same interval, carbon dioxide tension increased and end-tidal carbon dioxide decreased; shortly thereafter, transcutaneous oxygen tension decreased also. Within 20 minutes after the inspired gas mixture was changed to 100% oxygen, the patient's respiratory variables returned to near baseline. No further complications ensued. This is the first time continuously monitored arterial oxygen tension values during a pulmonary embolism have been reported. With further refinement, intraarterial optode probes will add another valuable method of detecting pulmonary air embolism.Entities:
Mesh:
Substances:
Year: 1990 PMID: 2295897 DOI: 10.1007/bf02832181
Source DB: PubMed Journal: J Clin Monit ISSN: 0748-1977