| Literature DB >> 22933938 |
Csilla Pesznyák1, István Polgár, Csaba Weisz, Réka Király, Pál Zaránd.
Abstract
BACKGROUND: The purpose of the study was to verify different values of quality parameters of portal images in radiotherapy.Entities:
Keywords: electronic portal imaging device; portal film; quality control
Year: 2010 PMID: 22933938 PMCID: PMC3423718 DOI: 10.2478/v10019-010-0052-6
Source DB: PubMed Journal: Radiol Oncol ISSN: 1318-2099 Impact factor: 2.991
FIGURE 1.Structure of the PTW EPID QC PHANTOM®, 1. Signal linearity and signal noise ratio, 2. Isotropy of signal linearity, 3. Geometric isotropy (distortion), 4. Low-contrast resolution, 5. High-contrast resolution (MTF).
FIGURE 2.Controlling the low contrast resolution with (A) Las Vegas test tool and (B) PTW EPID QC PHANTOM®.
FIGURE 3.The image analysis with the epidSoft®2.0 program.
Results of the portal image analysis with the epidSoft®2.0 program for the different equipments
| PortalVision™ | dicom 3.0 | 7+7 | 0.288 | 0.402 | 52.8 | 0.57 | 1.43 |
| PortalVision™ | dicom RI | 7+7 | 0.239 | 0.342 | 2532.7 | 0.67 | 1.63 |
| PortalVision™ | bmp | 7+7 | 0.251 | 0.355 | 107.3 | 0.60 | 1.50 |
| BeamView Plus® | dcm | 8 | 0.307 | 0.437 | 38.1 | 12.9 | 3.24 |
| BeamView Plus® 15MV | bmp | 8 | 0.225 | 0.378 | 40.7 | 11.0 | 2.16 |
| BeamView Plus® | bmp | 8 | 0.216 | 0.399 | 23.2 | 10.8 | 2.47 |
| BeamView Plus® | bmp inverse | 8 | 0.310 | 0.435 | 37.8 | 12.8 | 2.92 |
| BeamView Plus® | bmp | 10 | 0.242 | 0.402 | 54.2 | 11.6 | 2.75 |
| BeamView Plus® | bmp | 8+8 | 0.241 | 0.399 | 40.8 | 10.9 | 2.16 |
| BeamView Plus® 15MV | bmp | 8+8 | 0.23 | 0.388 | 56.6 | 12.3 | 2.46 |
| OptiVue500aSi® | dcm | 1 | 0.317 | 0.574 | 93.3 | 6.33 | 2.45 |
| OptiVue500aSi® | dcm | 2 | 0.315 | 0.573 | 105.3 | 6.15 | 2.48 |
| OptiVue500aSi® | dcm | 4 | 0.315 | 0.569 | 95.3 | 6.16 | 2.37 |
| OptiVue500aSi® | dcm | 6 | 0.315 | 0.563 | 86.4 | 6.10 | 2.32 |
| OptiVue500aSi® | dcm | 8 | 0.315 | 0.563 | 72.9 | 6.08 | 2.23 |
| Elekta iView® | bmp | 1 | 0.323 | 0.597 | 115.3 | 5.07 | 1.58 |
| Elekta iView® | bmp | 2 | 0.324 | 0.602 | 102.9 | 5.01 | 1.18 |
| Elekta iView® | bmp | 4 | 0.321 | 0.576 | 99.2 | 5.06 | 1.28 |
| Elekta iView® | bmp | 6 | 0.315 | 0.572 | 90.8 | 5.03 | 1.48 |
| Elekta iView® | bmp | 8 | 0.305 | 0.539 | 72.4 | 4.67 | 1.34 |
| X-OMAT® film | bmp | 20 | 0.322 | 0.609 | 248.1 | 4.62 | 1.52 |
| X-OMAT® film | tif | 20 | 0.333 | 0.548 | 167.8 | 3.61 | 2.59 |
| X-OMAT® film | dicom RI | 20 | 0.207 | 0.396 | 146.8 | 3.59 | 2.32 |
| X-OMAT® film | dicom RI | 40 | 0.275 | 0.692 | 105.6 | 2.34 | 2.68 |
| EC-L® film | dicom RI | 1 | 0.336 | 0.596 | 90.2 | 5.01 | 1.89 |
| EC-L® film | dicom RI | 2 | 0.316 | 0.569 | 100.7 | 4.71 | 2.36 |
| EC-L® film | dicom RI | 4 | 0.331 | 0.574 | 92.5 | 5.00 | 1.71 |
| EC-L® film | dicom RI | 6 | 0.306 | 0.563 | 88.6 | 6.84 | 3.96 |
| EC-L® film | tif | 1 | 0.324 | 0.572 | 113.1 | 4.97 | 1.89 |
| EC-L® film | tif | 2 | 0.312 | 0.563 | 119.9 | 4.66 | 2.48 |
| EC-L® film | tif | 4 | 0.324 | 0.584 | 110.2 | 4.96 | 1.71 |
| EC-L® film | tif | 10 | 0.291 | 0.600 | 88.2 | 10.02 | 3.63 |
LCSW, Linearity of Copper Step Wedge
LDL, Local Dependence of Linearity
Demonstration of the quantities to be used for the quality control of the EPIDs found in the references
| Clements at al. 2002, PIPSpro® QC-3V fantom [118] | ||||||
| Varian aS500™ | 512 × 384 | 0.78 | 5 | 260 | 0.392 | 0.600 |
| Elekta iViewGT® | 1024 × 1024 | 0.4 | 100 | 448 | 0.461 | 0.767 |
| Siemens FP-A | 1024 × 1024 | 0.4 | 100 | 611 | 0.454 | 0.696 |
| Hermann at al. 2001. [119] | ||||||
| BeamView Plus® | 512 × 512 | 0.204 | ||||
| PortalVision™ | 256 × 256 | 0.258 | ||||
| Wong, 1999. [120] | ||||||
| BeamView Plus® | 512 × 512 | 0,214 | ||||
| 0,192 (15 MV) | ||||||
| PortalVision™ | 256 × 256 | 0,258 | ||||
FIGURE 5.The graphic representation of the Varian PortalVision™ equipment’s stability test: A. linearity of copper step wedge, B. local dependence of linearity, C. MTF f(25) vertical and horizontal component, D. signal-to-noise ratio average value.
FIGURE 6.Comparison of results for low contrast resolution with PTW EPID QC PHANTOM® for Kodak X-OMAT® cassette with Kodak X-OMAT V® film and the Kodak EC-L Lightweight® cassette with Kodak Portal Localisation ReadyPack® film.