BACKGROUND: Arterial stiffness, as measured by the pulse wave velocity (PWV), is recommended for routine use in clinical practice as an important parameter for the evaluation of cardiovascular risk.1 New 24-hour monitors (eg, with Vasotens technology; Petr Telegin Company, Nizhny Novgorod, Russian Federation) provide single PWV measurements as well as several PWV measurements over a period of 24 hours.2 Such 24-hour pulse wave analysis led to the development of the novel Pulse Time Index of Norm (PTIN), which is defined as the percentage of a 24-hour period during which the PWV does not exceed the 10 m/second PWV threshold. The aim of this study is to test the new PTIN for correlation with the left ventricular mass index (LVMI). METHODS: Oscillometrically generated waveform files (n=137) used for clinical research studies were reanalyzed using the new 2013 version of the Vasotens technology program, which enables PTIN calculations. RESULTS: A good correlation (r=-0.72) between the PTIN and the LVMI was shown, which was significantly above the blood pressure load (r=0.41). CONCLUSION: The PTIN generated by the Vasotens technology can be recommended as an indicator of end organ damage via hypertension.
BACKGROUND: Arterial stiffness, as measured by the pulse wave velocity (PWV), is recommended for routine use in clinical practice as an important parameter for the evaluation of cardiovascular risk.1 New 24-hour monitors (eg, with Vasotens technology; Petr Telegin Company, Nizhny Novgorod, Russian Federation) provide single PWV measurements as well as several PWV measurements over a period of 24 hours.2 Such 24-hour pulse wave analysis led to the development of the novel Pulse Time Index of Norm (PTIN), which is defined as the percentage of a 24-hour period during which the PWV does not exceed the 10 m/second PWV threshold. The aim of this study is to test the new PTIN for correlation with the left ventricular mass index (LVMI). METHODS: Oscillometrically generated waveform files (n=137) used for clinical research studies were reanalyzed using the new 2013 version of the Vasotens technology program, which enables PTIN calculations. RESULTS: A good correlation (r=-0.72) between the PTIN and the LVMI was shown, which was significantly above the blood pressure load (r=0.41). CONCLUSION: The PTIN generated by the Vasotens technology can be recommended as an indicator of end organ damage via hypertension.
Arterial stiffness, as measured by the pulse wave velocity (PWV), is recommended for routine use in clinical practice as an important “vascular” index for the evaluation of cardiovascular risk.1 The PWV relates the capability of an artery to expand and contract under the influence of blood pressure (BP) changes. The PWV and other stiffness indices are independent predictors of cardiovascular morbidity and mortality in hypertensivepatients, in elderly populations, in patients with end-stage renal disease, and in patients with type 2 diabetes.1A novel cuff-based method, such as Vasotens® (Petr Telegin Company, Nizhny Novgorod, Russian Federation), determines the PWV from oscillometric measurements of the brachial artery waveform. This system for PWV measurements integrated into an ambulatory BP monitoring (ABPM) system provides single PWV measurements as well as several PWV measurements over a period of 24 hours. Such 24-hour pulse wave analysis has led to the development of the novel Pulse Time Index of Norm (PTIN), which is defined as the percentage of a 24-hour period during which the PWV does not exceed the threshold of the PWV of 10 m/second.2,3This cutoff value is well known as a threshold for hypertensive target organ damage.1,4,5 The PTIN calculation is similar to the principle of the BP load calculation. Several studies have shown that the mean 24-hour BP did not correlate with the left ventricular mass index (LVMI); however, an increased BP load was associated with a higher LVMI.6,7Does the PTIN provide additional information about cardiovascular risk? Recently registries have been established in European countries to assess the impact of ABPM on hypertension management and related outcomes. ABPM registries are more commonly focused on patients with documented or possible hypertension who would benefit from ABPM. The BPLab-Vasotens registry that was established this year is the project of dozens of small groups aimed at joint solutions for various scientific problems. One of the challenges is research into new ABPM “vascular” indices.To study the association of the PTIN with the LVMI, we used the ability to work with archives of the BPLab-Vasotens registry, which is provided by a unique feature of the Vasotens technology. This technology processes the oscillometrically generated waveform files that are received, digitized, and stored into the computer memory by the BPLab® (Petr Telegin Company) software for ABPM. The data processing can be performed during the analysis or at any time after the ABPM study.
Methods
Baseline characteristics
Overall, 137 oscillometrically generated waveform files that had previously been used in clinical research studies with ABPM were reanalyzed using the new 2013 version of the Vasotens technology program. The detailed baseline characteristics of the patients whose waveform files were used are shown in Table 1. These were patients with arterial hypertension; athletes and patients with a body mass index greater than 27.5 kg/m2 were excluded from the study.
Table 1
The baseline characteristics of the examined patients, n=137
Characteristic
Value
Male, n (%)
83 (61)
Age, mean (SD) years
46 (18)
Height, mean (SD) cm
168 (7)
Weight, mean (SD) kg
83 (17)
SBP, mean (SD) mmHg
144 (23)
DBP, mean (SD) mmHg
82 (13)
Heart rate, mean (SD) bpm
73 (14)
Hypertension, n (%)
137 (100)
Duration of hypertension history, mean (SD) years
6 (3)
Medication: thiazide diuretics, n (%)
80 (58)
Medication: angiotensin-receptor blockers, n (%)
91 (66)
Medication: beta-blockers, n (%)
60 (44)
Medication: other, n (%)
51 (37)
Drug combinations, n (%)
128 (93)
Previous MI or stroke, n (%)
26 (19)
LVH, n (%)
75 (55)
Renal disease, n (%)
4 (3)
Abbreviations: DBP, diastolic blood pressure; LVH, left ventricular hypertrophy; MI, myocardial infarction; SBP, systolic blood pressure; SD, standard deviation.
All of the studies in which the data were reanalyzed were authorized by the local ethics commissions. All of the enrolled patients provided written informed consent.
Vasotens
The Vasotens technology is an innovative method for pulse wave analysis based on oscillometric BP measurement by the BPLab device (Petr Telegin Company).8,9 The method of PTIN calculation consists of assessing the PWV. The PWV estimation method used for this parameter is based on identification of the reflected wave on the pulse curve at the arteria brachialis, taking into account the delay of the reflected wave relative to the direct wave. The transit time of both the direct wave and the reflected wave is equal to double the length of the aortic trunk. A more detailed description of the method was published in a recent review.2Recordings are made in the ambulatory (ie, 24-hour) measurement mode using a conventional BP cuff for adults. During the BP measurement, the pressure waveforms in the cuff are registered while performing a step-by-step deflation. The separation and timing of the forward and reflected pulse waves are performed by a Vasotens mathematical algorithm. The distance for the PWV equation used in this technology was measured in accordance with the manufacturer’s instructions. Quality control is based on visual assessment of the curves in the Vasotens clinical report screen. The PTIN calculation is illustrated in Figure 1.
Figure 1
The PTIN calculation principle.
Notes: T1, T2 and Tn are the periods in which the PWV does not exceed the cutoff value of 10 m/second. PTIN, % = (T1 + T2 +… + Tn)/Tm ×100, where Tm is the entire period of monitoring, Pulse Time index of norm.2,22
Abbreviations: PTIN, Pulse Time Index of Norm; PWV, pulse wave velocity.
BP load
We found it appropriate to use the identical BPLab device to evaluate the BP load because the hardware has an A/A class of accuracy for BP measurement.10 The BP load is defined as the percentage of a 24-hour period in which either the systolic BP exceeds the cutoff values of 135 mmHg during the day and 120 mmHg at night or the diastolic BP exceeds the cutoff values of 85 mmHg during the day and 70 mmHg during the night.
LVMI
The LVMI was determined for 137 hypertensivepatients using two-dimensional guided M-mode echocardiography performed with standard methods using an ultrasound system (Logiq 9 and Logiq 7 Ultrasound Machine; General Electric, Fairfield CT, USA) with a 2.25-MHz transducer. The left ventricular internal dimension (LVID), the interventricular septal thickness (IVST), and the posterior wall thickness (PWT) were measured during the end diastole period in accordance with the American Society of Echocardiography (ASE) recommendations.11 The left ventricular (LV) systolic and diastolic diameters and the IVST and the PWT were measured during the end diastole period which was defined by the beginning of the qRs-complex. The left ventricular mass (LVM) was calculated using the following corrected ASE formula by Devereux et al:12where LVID, left ventricular PWT, and IVST are measured in centimeters. The LVM was indexed by the body surface area (LVM/BSA, g/m2), which was calculated using the Dubois formula, and by height2.7 (LVM/height2.7, g/m2.7), as recommended by de Simone et al.13 We set the cutoff values for left ventricular hypertrophy (LVH) as LVM/BSA ≥116 g/m2 in males and ≥96 g/m2 in females; the cutoff value for the LVM/height2.7 was ≥49 g/m2.7 in males and ≥45 g/m2.7 in females.14
Statistics
We used BPStat (Petr Telegin Company) software, version 05.00.04 for the automatic tabulation of the indices of every measured waveform. Statistica (StatSoft Inc., Tulsa, OK, USA) software for Windows 8.0 was used for the analysis.
Results
The 137 hypertensivepatient ABPM waveform files were reanalyzed. First, it should be noted that the distribution of PTIN within this group could not be considered normal. The distribution is shown in Figure 2. Therefore, subsequent statistical analyses were performed with nonparametric tests that do not require the normality assumption that is usually used in parametric tests. For instance, to test for the correlation between the variables, we used Spearman’s r instead of the classical Pearson’s r coefficient.
Figure 2
Histogram of PTIN distribution – observed frequencies and expected normal curve.
Note: Pulse Time index of norm.2,22
Abbreviation: PTIN, Pulse Time Index of Norm.
Thus, correlation analysis with the LVM/height2.7 showed r=−0.72 (P<0.0001) for the PTIN (Figure 3A), r=0.41 (P<0.001) for the systolic BP load (Figure 3B), and r=0.47 (P<0.001) for the diastolic BP load (Figure 3C). The correlation between LVMI and PTIN was markedly higher.
Figure 3
Correlation between the LVMI (g/m2) and: (A) the PTIN (%), (B) the SBP load (%), and (C) the DBP load (%).
Notes: Pulse Time Index of norm.2,22
Abbreviations: DBP, diastolic blood pressure; SBP, systolic blood pressure; LVMI, left ventricular mass index; PTIN, Pulse Time Index of Norm.
Other correlations obtained in the study are presented in Table 2. It should be noted that the PTIN has more significant correlation coefficients with LVMI than the average PWV per day or its variability.
Table 2
The correlations between the echocardiographic data and the BPLab Vasotens® (Petr Telegin company, Nizhny Novgorod, Russian Federation) indices
LVM/height2.7
LVM/BSA
r
P-value
r
P-value
PTIN
−0.72
0.0001
−0.67
0.0001
Mean 24-hour PWV
0.32
0.01
0.32
0.01
Variability of PWV
0.12
>0.05
0.12
>0.05
SBP load
0.41
0.001
0.37
0.01
DBP load
0.47
0.001
0.37
0.01
Mean SBP
0.14
>0.05
0.15
>0.05
Mean DBP
0.29
>0.05
0.19
>0.05
Note: Pulse Time Index of Norm.2,22
Abbreviations: BSA, body surface area; DBP, diastolic blood pressure; LVM, left ventricular mass; PTIN, Pulse Time Index of Norm; PWV, pulse wave velocity; SBP, systolic blood pressure.
Important results were also obtained when PTIN was compared in patients with and without LVH. Figure 4 shows PTIN in 24-hour awake and asleep periods in these patients. As we can see, the boundaries of the PTIN percentile limits for the “asleep” period in these subgroups do not intersect. There are high significance levels (Yates corrected χ=34.2, P<0.001) if we construct a two-by-two table for awake and asleep periods in the subgroups listed above.
Figure 4
Medians, limits of the fifth and 95th percentiles and non-outlier range of PTIN according to 24-hour awake and asleep periods in patients with and without LVH.
Notes: Pulse Time Index of norm.2,22
Abbreviations: LVH, left ventricular hypertrophy; PTIN, Pulse Time Index of Norm; h, hour.
Discussion
LVH is the most commonly assessed target organ effect of hypertension, and it is associated with a significantly increased cardiac risk in adults.15 Several studies have shown that LVH, as with other markers of target organ damage, correlates with ambulatory BP more closely than with office B P.16,17 A number of additional indices that might be derived from the ABPM recordings have been investigated; however, their added predictive value is not clear, and they should be regarded as experimental.1 Some studies support the usefulness of circadian BP rhythm (a dipping pattern, morning BP) in predicting organ damage.18,19 Other studies show the usefulness of the BP load in predicting organ damage.6,7,20 In this present study, we obtained a similar correlation between the BP load and the LVMI. One study has also shown an association of the ambulatory arterial stiffness index with the LVMI.21Considering the above studies, our results from the PTIN study are very important. The correlation with the LVMI appears to be the highest among the correlations in similar studies of additional ABPM indexes. A difference can be seen in the correlations of the mean PWV with the LVMI (r=0.32) and of the PTIN with the LVMI (r=−0.72). In a comparison of recordings with similar average PWV levels, the differences in the PTIN can be largely explained by the differences in the PWV variability. Finally, the correlation of the PTIN of the “asleep” period with the presence of LVH is quite significant.Our study has some limitations, which might have affected the results. Because the distribution of the PTIN is non-normal, the adjustment of the correlation between PTIN and LVM using analysis of variance or analysis of covariance is not possible. Perhaps the correlation differs across age groups or in patients with varying clinical parameters.It is evident that only an event-based longitudinal study can provide an indisputable answer to the question of the clinical relevance of additional ABPM indices. Our previous work on the reliability of PTIN shows its good repeatability.22 Thus, given the above results, we believe that the PTIN especially during the “asleep” period should also be considered in forthcoming research.
Conclusion
The PTIN generated by the Vasotens technology can be recommended as an indicator of end organ damage resulting from hypertension.
Authors: Igor N Posokhov; Aleksandra O Konradi; Eugeny V Shlyakhto; Oleg V Mamontov; Artemy V Orlov; Anatoly N Rogoza Journal: Med Devices (Auckl) Date: 2014-02-27
Authors: Stefano Omboni; Igor N Posokhov; Yulia V Kotovskaya; Athanase D Protogerou; Jacques Blacher Journal: Curr Hypertens Rep Date: 2016-10 Impact factor: 5.369
Authors: Stefano Omboni; Igor N Posokhov; Gianfranco Parati; Alberto Avolio; Anatoly N Rogoza; Yulia V Kotovskaya; Giuseppe Mulè; Maria Lorenza Muiesan; Iana A Orlova; Elena A Grigoricheva; Ernesto Cardona Muñoz; Parounak H Zelveian; Telmo Pereira; João Manuel Peixoto Maldonado Journal: JMIR Res Protoc Date: 2016-06-29