Accuracy of noninvasive and invasive point-of-care total blood hemoglobin measurement in an outpatient setting.

Marina Raikhel1.   

Abstract

OBJECTIVE: The purpose of this study was to determine the accuracy of noninvasive hemoglobin measurement using pulse CO-oximetry (Pronto® SpHb®, Masimo Corp.) and a commonly used, invasive, point-of-care, automated spectrophotometer (HemoCue 201+®, HemoCue, Inc.), in comparison with hemoglobin measurements obtained from a laboratory hematology analyzer (measuring total blood hemoglobin) in the outpatient setting.
METHODS: Adult patients presenting to an outpatient research clinic were tested for total blood hemoglobin measurement by 3 methods: noninvasive pulse CO-oximetry (SpHb®), finger-stick blood sample on a point-of-care device, and venous sample on a laboratory hematology analyzer (reference device). Bias and standard deviation (SD) of SpHb® and HemoCue 201+® compared with the values obtained with the laboratory hematology analyzer were calculated and Bland-Altman graphs were generated.
RESULTS: Samples from 152 subjects were assessed (average age, 46 years; 69% female). The bias ± SD compared with the reference method was -0.5 ± 1.0 g/dL for SpHb® and 0.3 ± 1.0 g/dL for HemoCue 201+®. The Bland-Altman plots assessing agreement of the test methods to the reference method had limits of agreement of -2.5 to 1.5 g/dL for SpHb® and -1.7 to 2.3 g/dL for HemoCue 201+®. A noninvasive measurement could not be obtained in 4 subjects after 2 attempts (2.5% failure rate), whereas the HemoCue 201+® measurements were obtained for all subjects.
CONCLUSION: Noninvasive SpHb® testing had bias and SD similar to those of HemoCue 201+®. Because SpHb® measurement is noninvasive, it may offer additional patient and provider benefits.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22913913     DOI: 10.3810/pgm.2012.07.2584

Source DB:  PubMed          Journal:  Postgrad Med        ISSN: 0032-5481            Impact factor:   3.840


  11 in total

1.  A noninvasive hemoglobin monitor in the pediatric intensive care unit.

Authors:  Michael R Phillips; Amal L Khoury; Andrey V Bortsov; Amy Marzinsky; Kathy A Short; Bruce A Cairns; Anthony G Charles; Benny L Joyner; Sean E McLean
Journal:  J Surg Res       Date:  2015-01-09       Impact factor: 2.192

2.  Comparison of the accuracy of noninvasive hemoglobin monitoring for preoperative evaluation between adult and pediatric patients: a retrospective study.

Authors:  Yong-Hee Park; Seho Lim; Hyun Kang; Hwa-Yong Shin; Chong Wha Baek; Young Cheol Woo
Journal:  J Clin Monit Comput       Date:  2018-01-04       Impact factor: 2.502

3.  SmartHeLP: Smartphone-based Hemoglobin Level Prediction Using an Artificial Neural Network.

Authors:  Md Kamrul Hasan; Md Munirul Haque; Riddhiman Adib; Jannatul F Tumpa; Azima Begum; Richard R Love; Young L Kim; I Ahamed Sheikh
Journal:  AMIA Annu Symp Proc       Date:  2018-12-05

4.  Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants.

Authors:  C Nicholas; R George; S Sardesai; M Durand; R Ramanathan; R Cayabyab
Journal:  J Perinatol       Date:  2015-03-05       Impact factor: 2.521

5.  Patient blood management implementation strategies and their effect on physicians' risk perception, clinical knowledge and perioperative practice - the frankfurt experience.

Authors:  Dania P Fischer; Kai D Zacharowski; Markus M Müller; Christof Geisen; Erhard Seifried; Heiko Müller; Patrick Meybohm
Journal:  Transfus Med Hemother       Date:  2015-03-12       Impact factor: 3.747

6.  Reliability of transcutaneous hemoglobin measurements in an outpatient pediatric clinic.

Authors:  Shae Margulies; Molly Posa; Melissa Fitzgerald; Stephanie Filipp; Diane Howell; Puneet Tung; Maria Kelly
Journal:  Postgrad Med       Date:  2021-12-27       Impact factor: 3.840

Review 7.  How Noninvasive Haemoglobin Measurement with Pulse CO-Oximetry Can Change Your Practice: An Expert Review.

Authors:  Gregor Lindner; Aristomenis K Exadaktylos
Journal:  Emerg Med Int       Date:  2013-08-22       Impact factor: 1.112

8.  Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer.

Authors:  N Shah; E A Osea; G J Martinez
Journal:  Int J Lab Hematol       Date:  2013-06-27       Impact factor: 2.877

Review 9.  Continuous Noninvasive Hemoglobin Monitoring: A Measured Response to a Critical Review.

Authors:  Steven J Barker; Aryeh Shander; Michael A Ramsay
Journal:  Anesth Analg       Date:  2016-02       Impact factor: 5.108

10.  Assessment of noninvasive, percutaneous hemoglobin measurement in pregnant and early postpartum women.

Authors:  Atsushi Yoshida; Keiko Saito; Kenji Ishii; Isao Azuma; Hidenori Sasa; Kenichi Furuya
Journal:  Med Devices (Auckl)       Date:  2014-01-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.