Overcoming x-ray tube small focal spot output limitations for high-resolution region of interest imaging.

Sandesh K Gupta1, Amit Jain, Daniel R Bednarek, Stephen Rudin.   

Abstract

We investigate methods to increase x-ray tube output to enable improved quantum image quality with a higher generalized-NEQ (GNEQ) while maintaining a small focal-spot size for the new high-resolution Micro-angiographic Fluoroscope (MAF) Region of Interest (ROI) imaging system. Rather than using a larger focal spot to increase tube-loading capacity with degraded resolution, we evaluated separately or in combination three methods to increase tube output: 1) reducing the anode angle and lengthening the filament to maintain a constant effective small focal-spot size, 2) using the standard medium focal spot viewed from a direction on the anode side of the field and 3) increasing the frame rate (frames/second) in combination with temporal filter. The GNEQ was compared for the MAF for the small focal-spot at the central axis, and for the medium focal-spot with a higher output on the anode side as well as for the small focal spot with different temporal recursive filtering weights. A net output increase of about 4.0 times could be achieved with a 2-degree anode angle (without the added filtration) and a 4 times longer filament compared to that of the standard 8-degree target. The GNEQ was also increased for the medium focal-spot due to its higher output capacity and for the temporally filtered higher frame rate. Thus higher tube output, while maintaining a small effective focal-spot, should be achievable using one or more of the three methods described with only small modifications of standard x-ray tube geometry.

Entities:  

Year:  2012        PMID: 22905312      PMCID: PMC3419545          DOI: 10.1117/12.910791

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  18 in total

1.  New device for accurate measurement of the x-ray intensity distribution of x-ray tube focal spots.

Authors:  K Doi; B Fromes; K Rossmann
Journal:  Med Phys       Date:  1975 Sep-Oct       Impact factor: 4.071

2.  Optimization of variable temporal averaging in digital fluoroscopy.

Authors:  C J Kotre; E Guibelalde
Journal:  Br J Radiol       Date:  2004-08       Impact factor: 3.039

3.  Calculation of x-ray spectra emerging from an x-ray tube. Part II. X-ray production and filtration in x-ray targets.

Authors:  Gavin G Poludniowski
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

4.  Implementation of a high-sensitivity Micro-Angiographic Fluoroscope (HS-MAF) for in-vivo endovascular image guided interventions (EIGI) and region-of-interest computed tomography (ROI-CT).

Authors:  C N Ionita; C Keleshis; V Patel; G Yadava; K R Hoffmann; D R Bednarek; A Jain; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008

5.  SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes.

Authors:  G Poludniowski; G Landry; F DeBlois; P M Evans; F Verhaegen
Journal:  Phys Med Biol       Date:  2009-09-01       Impact factor: 3.609

6.  Study of the Generalized MTF and DQE for a New Microangiographic System.

Authors:  Iacovos S Kyprianou; Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2004-05-06

7.  A method for measuring the presampled MTF of digital radiographic systems using an edge test device.

Authors:  E Samei; M J Flynn; D A Reimann
Journal:  Med Phys       Date:  1998-01       Impact factor: 4.071

8.  Generalized Performance Evaluation of X-ray Image Intensifier compared with a Microangiographic System.

Authors:  Girijesh K Yadava; Iacovos S Kyprianou; Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2005

9.  Generalized two-dimensional (2D) linear system analysis metrics (GMTF, GDQE) for digital radiography systems including the effect of focal spot, magnification, scatter, and detector characteristics.

Authors:  Amit Jain; Andrew T Kuhls-Gilcrist; Sandesh K Gupta; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2010-03-01

10.  Endovascular image-guided interventions (EIGIs).

Authors:  Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

View more
  2 in total

1.  Novel method to determine recursive filtration and noise reduction in fluoroscopic imaging - a comparison of four different vendors.

Authors:  Bente Konst; Jacob Nøtthellen; Stine Nalum Naess; Magnus Båth
Journal:  J Appl Clin Med Phys       Date:  2020-12-14       Impact factor: 2.102

2.  Focal spot size reduction using asymmetric collimation to enable reduced anode angles with a conventional angiographic x-ray tube for use with high resolution detectors.

Authors:  M Russ; A Shankar; S V Setlur Nagesh; C N Ionita; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.