High-dose MVCT image guidance for stereotactic body radiation therapy.

David C Westerly1, Tracey E Schefter, Brian D Kavanagh, Edward Chao, Dan Lucas, Ryan T Flynn, Moyed Miften.   

Abstract

PURPOSE: Stereotactic body radiation therapy (SBRT) is a potent treatment for early stage primary and limited metastatic disease. Accurate tumor localization is essential to administer SBRT safely and effectively. Tomotherapy combines helical IMRT with onboard megavoltage CT (MVCT) imaging and is well suited for SBRT; however, MVCT results in reduced soft tissue contrast and increased image noise compared with kilovoltage CT. The goal of this work was to investigate the use of increased imaging doses on a clinical tomotherapy machine to improve image quality for SBRT image guidance.
METHODS: Two nonstandard, high-dose imaging modes were created on a tomotherapy machine by increasing the linear accelerator (LINAC) pulse rate from the nominal setting of 80 Hz, to 160 Hz and 300 Hz, respectively. Weighted CT dose indexes (wCTDIs) were measured for the standard, medium, and high-dose modes in a 30 cm solid water phantom using a calibrated A1SL ion chamber. Image quality was assessed from scans of a customized image quality phantom. Metrics evaluated include: contrast-to-noise ratios (CNRs), high-contrast spatial resolution, image uniformity, and percent image noise. In addition, two patients receiving SBRT were localized using high-dose MVCT scans. Raw detector data collected after each scan were used to reconstruct standard-dose images for comparison.
RESULTS: MVCT scans acquired using a pitch of 1.0 resulted in wCTDI values of 2.2, 4.7, and 8.5 cGy for the standard, medium, and high-dose modes respectively. CNR values for both low and high-contrast materials were found to increase with the square root of dose. Axial high-contrast spatial resolution was comparable for all imaging modes at 0.5 lp∕mm. Image uniformity was improved and percent noise decreased as the imaging dose increased. Similar improvements in image quality were observed in patient images, with decreases in image noise being the most notable.
CONCLUSIONS: High-dose imaging modes are made possible on a clinical tomotherapy machine by increasing the LINAC pulse rate. Increasing the imaging dose results in increased CNRs; making it easier to distinguish the boundaries of low contrast objects. The imaging dose levels observed in this work are considered acceptable at our institution for SBRT treatments delivered in 3-5 fractions.

Entities:  

Mesh:

Year:  2012        PMID: 22894407     DOI: 10.1118/1.4736416

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Megavoltage CT imaging quality improvement on TomoTherapy via tensor framelet.

Authors:  Hao Gao; X Sharon Qi; Yu Gao; Daniel A Low
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

2.  Radiation-induced CT number changes in GTV and parotid glands during the course of radiation therapy for nasopharyngeal cancer.

Authors:  Shouping Xu; Zhaoxia Wu; Cungeng Yang; Lin Ma; Baolin Qu; Guangpei Chen; Weirong Yao; Shi Wang; Yaqiang Liu; X Allen Li
Journal:  Br J Radiol       Date:  2016-04-01       Impact factor: 3.039

3.  Improved contrast and noise of megavoltage computed tomography (MVCT) through cycle-consistent generative machine learning.

Authors:  Luciano Vinas; Jessica Scholey; Martina Descovich; Vasant Kearney; Atchar Sudhyadhom
Journal:  Med Phys       Date:  2020-12-27       Impact factor: 4.071

4.  Phase I dose escalation study of helical intensity-modulated radiotherapy-based stereotactic body radiotherapy for hepatocellular carcinoma.

Authors:  Jun Won Kim; Jinsil Seong; Ik Jae Lee; Joong Yeol Woo; Kwang-Hyub Han
Journal:  Oncotarget       Date:  2016-06-28
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.