The accuracy and efficacy of real-time continuous glucose monitoring sensor in Chinese diabetes patients: a multicenter study.

Jian Zhou1, Xiaofeng Lv, Yiming Mu, Xianling Wang, Jing Li, Xingguang Zhang, Jinxiao Wu, Yuqian Bao, Weiping Jia.   

Abstract

BACKGROUND: The purpose of this multicenter study was to investigate the accuracy of a real-time continuous glucose monitoring sensor in Chinese diabetes patients. SUBJECTS AND METHODS: In total, 48 patients with type 1 or 2 diabetes from three centers in China were included in the study. The MiniMed Paradigm(®) 722 insulin pump (Medtronic, Northridge, CA) was used to monitor the real-time continuous changes of blood glucose levels for three successive days. Venous blood of the subjects was randomly collected every 15 min for seven consecutive hours on the day when the subjects were wearing the sensor. Reference values were provided by the YSI(®) 2300 STAT PLUS™ glucose and lactate analyzer (YSI Life Sciences, Yellow Springs, OH).
RESULTS: In total, 1,317 paired YSI-sensor values were collected from the 48 patients. Of the sensor readings, 88.3% (95% confidence interval, 0.84-0.92) were within±20% of the YSI values, and 95.7% were within±30% of the YSI values. Clarke and consensus error grid analyses showed that the ratios of the YSI-sensor values in Zone A to the values in Zone B were 99.1% and 99.9%, respectively. Continuous error grid analysis showed that the ratios of the YSI-sensor values in the region of accurate reading, benign errors, and erroneous reading were 96.4%, 1.8%, and 1.8%, respectively. The mean absolute relative difference (ARD) for all subjects was 10.4%, and the median ARD was 7.8%. Bland-Altman analysis detected a mean blood glucose level of 3.84 mg/dL. Trend analysis revealed that 86.1% of the difference of the rates of change between the YSI values and the sensor readings occurred within the range of 1 mg/dL/min.
CONCLUSIONS: The Paradigm insulin pump has high accuracy in both monitoring the real-time continuous changes and predicting the trend of changes in blood glucose level. However, actual clinical manifestations should be taken into account for diagnosis of hypoglycemia.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22853722      PMCID: PMC3409457          DOI: 10.1089/dia.2012.0014

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  22 in total

1.  Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring.

Authors:  Dorothee Deiss; Jan Bolinder; Jean-Pierre Riveline; Tadej Battelino; Emanuele Bosi; Nadia Tubiana-Rufi; David Kerr; Moshe Phillip
Journal:  Diabetes Care       Date:  2006-12       Impact factor: 19.112

2.  Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.

Authors:  Richard L Weinstein; Sherwyn L Schwartz; Ronald L Brazg; Jolyon R Bugler; Thomas A Peyser; Geoffrey V McGarraugh
Journal:  Diabetes Care       Date:  2007-03-02       Impact factor: 19.112

3.  Accuracy of a new real-time continuous glucose monitoring algorithm.

Authors:  D Barry Keenan; Raymond Cartaya; John J Mastrototaro
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

4.  Evaluation of a novel continuous glucose measurement device in patients with diabetes mellitus across the glycemic range.

Authors:  Linda Morrow; Marcus Hompesch; Ann M Tideman; Jennifer Matson; Nancy Dunne; Scott Pardo; Joan L Parkes; Holly C Schachner; David A Simmons
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

5.  Evaluating clinical accuracy of systems for self-monitoring of blood glucose.

Authors:  W L Clarke; D Cox; L A Gonder-Frederick; W Carter; S L Pohl
Journal:  Diabetes Care       Date:  1987 Sep-Oct       Impact factor: 19.112

6.  [Accuracy and safety of continuous glucose monitoring system in diabetic and non-diabetic subjects].

Authors:  Li-fang Lü; Chun Wang; Yan-zhi Yang; Li-ping He; Guan-jian Liu; Da-wei Chen; Li Zhong; Li-hong Chen; Hao-ming Tian; Jian Zhou; Wei-ping Jia; Xing-wu Ran
Journal:  Zhonghua Yi Xue Za Zhi       Date:  2010-11-16

7.  The Continuous Glucose Monitoring System during pregnancy of women with type 1 diabetes mellitus: accuracy assessment.

Authors:  Anneloes Kerssen; Harold W de Valk; Gerard H A Visser
Journal:  Diabetes Technol Ther       Date:  2004-10       Impact factor: 6.118

8.  Analysis, modeling, and simulation of the accuracy of continuous glucose sensors.

Authors:  Marc Breton; Boris Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2008-09

9.  Continuous glucose monitoring and intensive treatment of type 1 diabetes.

Authors:  William V Tamborlane; Roy W Beck; Bruce W Bode; Bruce Buckingham; H Peter Chase; Robert Clemons; Rosanna Fiallo-Scharer; Larry A Fox; Lisa K Gilliam; Irl B Hirsch; Elbert S Huang; Craig Kollman; Aaron J Kowalski; Lori Laffel; Jean M Lawrence; Joyce Lee; Nelly Mauras; Michael O'Grady; Katrina J Ruedy; Michael Tansey; Eva Tsalikian; Stuart Weinzimer; Darrell M Wilson; Howard Wolpert; Tim Wysocki; Dongyuan Xing
Journal:  N Engl J Med       Date:  2008-09-08       Impact factor: 91.245

10.  Incremental value of continuous glucose monitoring when starting pump therapy in patients with poorly controlled type 1 diabetes: the RealTrend study.

Authors:  Denis Raccah; Véronique Sulmont; Yves Reznik; Bruno Guerci; Eric Renard; Hélène Hanaire; Nathalie Jeandidier; Marc Nicolino
Journal:  Diabetes Care       Date:  2009-09-18       Impact factor: 19.112

View more
  3 in total

1.  Cutaneous Complications With Continuous or Flash Glucose Monitoring Use: Systematic Review of Trials and Observational Studies.

Authors:  Nurul A Mohd Asarani; Andrew N Reynolds; Sara E Boucher; Martin de Bock; Benjamin J Wheeler
Journal:  J Diabetes Sci Technol       Date:  2019-08-27

2.  A Multicenter Evaluation of the Performance and Usability of a Novel Glucose Monitoring System in Chinese Adults With Diabetes.

Authors:  Linong Ji; Xiaohui Guo; Lixin Guo; Qian Ren; Nan Yu; Jie Zhang
Journal:  J Diabetes Sci Technol       Date:  2016-09-25

3.  Performance of a new real-time continuous glucose monitoring system: A multicenter pilot study.

Authors:  Jian Zhou; Shuo Zhang; Liang Li; Yufei Wang; Wei Lu; Chunjun Sheng; Yiming Li; Yuqian Bao; Weiping Jia
Journal:  J Diabetes Investig       Date:  2017-07-07       Impact factor: 4.232

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.