Literature DB >> 22812488

A new transcutaneous bidirectional communication for monitoring implanted artificial heart using the human body as a conductive medium.

Eiji Okamoto1, Yoshikuni Kato, Kazuyuki Seino, Hidekazu Miura, Yasuyuki Shiraishi, Tomoyuki Yambe, Yoshinori Mitamura.   

Abstract

A transcutaneous communication system (TCS) is a key technology for monitoring and controlling artificial hearts and other artificial organs in the body. In this study, we developed a new TCS that uses the human body as a conductive medium. Direct data exchange provides a higher level of communication security compared to that of wireless methods without physical constraints such as an external wire. The external and internal units of the new TCS each consist mainly of a data transmitter and a data receiver. The data transmitter has an amplitude shift keying (ASK) modulator (carrier frequencies: 4 and 10 MHz) and an electrode. The ASK-modulated data current is led into the body through the electrode, and it flows back to the energy source through the body, the data receiver, and the earth ground that includes all conductors and dielectrics in the environment that are in close proximity to the patient. Performance of the TCS was evaluated by a communication test on the surface of the human body and in an animal experiment using a goat. The TCS was able to transmit data concurrently for 4 weeks between everywhere on the surface of the body and everywhere inside the body under full-duplex communication at a transmission rate of 115 kbps. The power consumption of each TCS unit was 125 mW with an ASK-modulated current of 7 mA (root-mean-square). While further study is required to secure its safety, the newly developed TCS has promise to be a next-generation transcutaneous communication device.
© 2012, Copyright the Authors. Artificial Organs © 2012, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22812488     DOI: 10.1111/j.1525-1594.2012.01480.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  2 in total

1.  Histological investigation of the titanium fiber mesh with one side sealed with non-porous material for its application to the artificial heart system.

Authors:  Eiji Okamoto; Kyoko Arimura; Yoshinori Mitamura
Journal:  J Artif Organs       Date:  2018-08-20       Impact factor: 1.731

2.  Electrical characteristic of the titanium mesh electrode for transcutaneous intrabody communication to monitor implantable artificial organs.

Authors:  Eiji Okamoto; Sakiko Kikuchi; Yoshinori Mitamura
Journal:  J Artif Organs       Date:  2016-02-26       Impact factor: 1.731

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.