Literature DB >> 22795975

Prevention of capsular bag opacification with a modified hydrophilic acrylic disk-shaped intraocular lens.

Lisa Leishman1, Liliana Werner, Zachary Bodnar, Andrew Ollerton, Jennifer Michelson, Mason Schmutz, Nick Mamalis.   

Abstract

PURPOSE: To evaluate the stability and capsular bag opacification with a modified disk-shaped 1-piece hydrophilic acrylic intraocular lens (IOL) suspended between 2 complete haptic rings connected by a pillar of the haptic material and with a commercially available 1-piece hydrophilic acrylic IOL.
SETTING: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.
DESIGN: Experimental study.
METHODS: Study and control IOLs were implanted into the left eyes and right eyes of 5 New Zealand rabbits. Eyes were examined at the slitlamp from 1 through 4 weeks. At 4 weeks, the globes were enucleated and evaluated under a very-high-frequency ultrasound. Photographs were taken and capsular bag opacification scored from the posterior aspect (Miyake-Apple view), and the eyes were processed for complete histopathology.
RESULTS: At 4 weeks, the posterior capsule opacification score was 0.0 in the study group and 1.75 ± 0.5 (SD) in the control group (P=.005, paired t test). Ultrasound examination showed that 2 of the study IOLs had no contact between the posterior optic surface and the posterior capsule. Minimal proliferative cortical material was confined to the peripheral space between anterior and posterior rings of the study IOL haptics in localized areas at the equatorial region of the capsular bag. Anterior capsule opacification was absent in all eyes.
CONCLUSIONS: The study IOL is a modification of a previous design, incorporating haptic perforations between the peripheral rings. By maintaining an open capsular bag and enhancing endocapsular inflow of aqueous, this modified design appears to prevent capsular bag opacification.
Copyright © 2012 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22795975     DOI: 10.1016/j.jcrs.2012.04.040

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  [Intraocular lens as a drug delivery device].

Authors:  K H Eibl-Lindner; R Liegl; C Wertheimer; A Kampik
Journal:  Ophthalmologe       Date:  2013-10       Impact factor: 1.059

2.  Posterior capsular opacification in highly myopic eyes with an endocapsular equator ring.

Authors:  Tsutomu Hara; Takeshi Hara; Takako Hashimoto; Yuta Motoyama; Masaya Narita
Journal:  Jpn J Ophthalmol       Date:  2016-06-15       Impact factor: 2.447

3.  The human capsular bag model of posterior capsule opacification.

Authors:  I Michael Wormstone
Journal:  Eye (Lond)       Date:  2019-11-19       Impact factor: 3.775

4.  Further Enhancement of Intraocular Open-Capsule Devices for Prevention of Posterior Capsule Opacification.

Authors:  Lee Slutzky; Guy Kleinmann
Journal:  Transl Vis Sci Technol       Date:  2018-02-28       Impact factor: 3.283

5.  Growth factor restriction impedes progression of wound healing following cataract surgery: identification of VEGF as a putative therapeutic target.

Authors:  Julie A Eldred; Matthew McDonald; Helen S Wilkes; David J Spalton; I Michael Wormstone
Journal:  Sci Rep       Date:  2016-04-14       Impact factor: 4.379

  5 in total