Optimizing imaging quality in endourology with the Uro Dyna-CT: contrast agent dilution matters.

M Ritter1, C Weiss, M-C Rassweiler, M S Michel, A Häcker.   

Abstract

BACKGROUND AND
PURPOSE: The implementation of the Uro Dyna-CT (Siemens Healthcare Solutions, Erlangen, Germany) with interventional 3D- and cross-sectional imaging necessitates the development of new standards for endourologic procedures such as the customized use of diluted contrast agent. Our aim was to find the ideal contrast agent dilution (CAD) for the interventional use of the Uro Dyna-CT in a standardized setting and prove the experimental findings in a clinical case.
MATERIALS AND METHODS: Retrograde pyelographies were performed with 10 different CADs of Imeron 300 (Bracco, Konstanz, Germany) in 10 different pig's urinary tracts. Fluoroscopy, X-ray, 3D- and slice-image reconstruction was performed with the Uro Dyna-CT. Image quality was evaluated, blinded and randomized by 5 observers. Small plastic jars were filled with the CADs and 2 artificial stones (Plaster of paris). Images were evaluated by two observers. The ex vivo findings were transferred to a clinical setting in a complex percutaneous lithotomy procedure. Unweighted and weighted kappa coefficients were calculated to indicate the degree of observers' agreement.
RESULTS: Twenty percent diluted contrast agent provides the best image quality and stone detection when interventional cross-sectional imaging is considered without limitations in fluoroscopy or X-ray quality. This was proved in a percutaneous lithotomy of an obese patient.
CONCLUSIONS: Image quality of the Uro Dyna-CT can be optimized by the use of 20 % diluted contrast agent. This knowledge helps to provide high-quality 3D imaging in the endourological operation room.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22790391     DOI: 10.1007/s00345-012-0903-4

Source DB:  PubMed          Journal:  World J Urol        ISSN: 0724-4983            Impact factor:   4.226


  13 in total

1.  Laser-guided percutaneous kidney access with the Uro Dyna-CT: first experience of three-dimensional puncture planning with an ex vivo model.

Authors:  M Ritter; M-C Rassweiler; A Häcker; M S Michel
Journal:  World J Urol       Date:  2012-03-04       Impact factor: 4.226

2.  Use of Dyna-CT Angiography in Neuroendovascular Decision-Making. A Case Report.

Authors:  K Namba; Y Niimi; J K Song; A Berenstein
Journal:  Interv Neuroradiol       Date:  2009-04-15       Impact factor: 1.610

3.  Volume CT with a flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation in guidance of minimally invasive surgery.

Authors:  J H Siewerdsen; D J Moseley; S Burch; S K Bisland; A Bogaards; B C Wilson; D A Jaffray
Journal:  Med Phys       Date:  2005-01       Impact factor: 4.071

4.  Three-dimensional imaging and cone beam volume CT in C-arm angiography with flat panel detector.

Authors:  Sergin Akpek; Thomas Brunner; Goetz Benndorf; Charles Strother
Journal:  Diagn Interv Radiol       Date:  2005-03       Impact factor: 2.630

5.  [The urological Dyna-CT. Urological cross-sectional imaging on a newly developed urological intervention table].

Authors:  M S Michel; M Ritter; S Schönberg; A Häcker; G Weisser
Journal:  Urologe A       Date:  2012-06       Impact factor: 0.639

6.  A comparison of the use of contrast media with different iodine concentrations for multidetector CT of the kidney.

Authors:  Seung Chai Jung; Seung Hyup Kim; Jeong Yeon Cho
Journal:  Korean J Radiol       Date:  2011-09-27       Impact factor: 3.500

7.  Voiding cystourethrogram--are our protocols the same?

Authors:  Blake W Palmer; Faridali G Ramji; Charles T Snyder; Michael Hemphill; Bradley P Kropp; Dominic Frimberger
Journal:  J Urol       Date:  2011-08-19       Impact factor: 7.450

8.  Initial observations of endovascular aneurysm repair using Dyna-CT.

Authors:  Kari Ravn Eide; Asbjørn Ødegård; Hans Olav Myhre; Olav Haraldseth
Journal:  J Endovasc Ther       Date:  2007-02       Impact factor: 3.487

9.  Where contrast agent concentration really matters - a comparison of CT and MRI.

Authors:  Sabine Heiland; Gunter Erb; Simon Ziegler; Martin Krix
Journal:  Invest Radiol       Date:  2010-09       Impact factor: 6.016

10.  Three-dimensional CT pyelography for planning of percutaneous nephrostolithotomy: accuracy of stone measurement, stone depiction and pelvicalyceal reconstruction.

Authors:  Uday Patel; Richard Miles Walkden; Khurshid R Ghani; Ken Anson
Journal:  Eur Radiol       Date:  2009-01-14       Impact factor: 5.315

View more
  2 in total

1.  Artifacts caused by insufficient contrast medium filling during C-arm cone-beam CT scans: a phantom study.

Authors:  Mitsuaki Terabe; Hajime Ichikawa; Toyohiro Kato; Kichiro Koshida
Journal:  Radiol Phys Technol       Date:  2013-06-18

2.  Accuracy in detecting and measuring residual fragments with the Uro Dyna-CT.

Authors:  B Meister; M C Rassweiler; C Weiß; A Häcker; M Ritter
Journal:  World J Urol       Date:  2014-08-28       Impact factor: 4.226

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.