| Literature DB >> 22778566 |
Hyun Su Kim1, Young Cheol Yoon, Jong Won Kwon, Bong-Keun Choe.
Abstract
OBJECTIVE: To compare the image quality of volume isotropic turbo spin echo acquisition (VISTA) imaging method with that of the three-dimensional (3D) isotropic fast field echo (FFE) imaging method applied for ankle joint imaging.Entities:
Keywords: 3D isotropic imaging; Ankle; Cartilage; MRI; Tendon
Mesh:
Year: 2012 PMID: 22778566 PMCID: PMC3384826 DOI: 10.3348/kjr.2012.13.4.443
Source DB: PubMed Journal: Korean J Radiol ISSN: 1229-6929 Impact factor: 3.500
MR Sequence Parameters
Note.- MR = magnetic resonance, 3D = three-dimensional, VISTA = volume isotropic turbo spin echo acquisition, TSE = turbo spin echo, WATS-f = water selective, fluid, TR/TE = repetition time/echo time, FOV = field of view, SENSE = sensitivity encoding, SPAIR = spectral attenuated inversion recovery, ProSet = principle of selective excitation technique
Interobserver Agreement in Evaluation of Subjective Tissue Contrast
Note.- Data are mean κ values; numbers in parentheses are 95% confidence interval. *Non-squared data are expressed as percentage of agreement. F = fluid, C = cartilage, T = tendon, 3D = three-dimensional, VISTA = volume isotropic turbo spin echo acquisition, WATS-f = water selective, fluid
Fig. 1MR images of right ankle in 31-year-old woman.
For 3D VISTA (A), reader A graded excellent and reader B graded good for grade of contrast between Achilles tendon and joint fluid. For 3D WATS-f (B), both readers graded fair. CR of SI of tendon to SI of synovial fluid was 0.953 in 3D VISTA and 0.920 in 3D WATS-f. 3D = three-dimensional, VISTA = volume isotropic turbo spin echo acquisition, WATS-f = water selective, fluid, CR = contrast ratio; SI = signal intensity
Fig. 2MR images of right ankle in 30-year-old man.
For 3D VISTA (A), both readers graded fair for grade of contrast between cartilage and joint fluid. For 3D WATS-f (B), both readers graded excellent. CR of SI of cartilage to SI of synovial fluid was 0.274 in 3D VISTA and 0.297 in 3D WATS-f. 3D = three-dimensional, VISTA = volume isotropic turbo spin echo acquisition, WATS-f = water selective, fluid, CR = contrast ratio, SI = signal intensity
Image Contrast Ratios
Note.- Measured CRs are the means ± SDs. Significant difference at p < 0.05. FFE = fast field echo, SD = standard deviation, VISTA = volume isotropic turbo spin echo acquisition, WATS-f = water selective, fluid