| Literature DB >> 22750675 |
G A Godwin1, J B Simpson, K V Mugabe.
Abstract
The Apex® dynamic mini-multileaf collimator has recently been released by Elekta and attaches directly to the linear accelerator head. This paper details the work and results obtained in characterizing this mini-MLC for stereotactic usage within our department. A range of mechanical and dosimetric characteristics were investigated which included inter and intra leaf leakage, light/radiation field congruence, leaf position reproducibility, radiation penumbra, total scatter factors and mechanical rotational stability with the additional mini-MLC weight.Entities:
Mesh:
Year: 2012 PMID: 22750675 DOI: 10.1088/0031-9155/57/14/4643
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609