P S Simeon1, M E Geffner, S R Levin, A M Lindsey.
Abstract
Continuous insulin infusion is sometimes used in very low birth weight infants with glucose intolerance. We studied the availability of insulin to the neonate by means of a laboratory-simulated clinical infusion under both flushed and non-flushed conditions. Although loss of insulin was noted with both solutions (0.5 U/ml), under the nonflushed condition only 55.9% of the prescribed dose for the first 2 hours was delivered before the delivery rate of the flushed system, 71.4%, was approached. These findings have implications for neonatal insulin administration.Entities:
Mesh:
Substances:
Year: 1994 PMID: 8176576 DOI: 10.1016/s0022-3476(05)81382-4
Source DB: PubMed Journal: J Pediatr ISSN: 0022-3476 Impact factor: 4.406