Biomechanical analysis of a novel locking plate with smooth pegs versus a conventional locking plate with threaded screws for proximal humerus fractures.

Nobuyuki Yamamoto1, Michio Hongo, Lawrence J Berglund, John W Sperling, Robert H Cofield, Kai-Nan An, Scott P Steinmann.   

Abstract

BACKGROUND: Repair of proximal humeral fractures with locking plates has greatly improved outcomes. However, an alarming rate of complications including screw cutout and impingement under the acromion has recently been reported. A novel locking plate with smooth pegs was developed to reduce these risks. The purpose of this study was to compare these 2 fixation methods with a cadaveric biomechanical study.
MATERIALS AND METHODS: Fourteen pairs of cadaveric proximal humeri (mean age, 77 years) were harvested, and bone density was measured. Osteosynthesis was performed on each pair using either a locking plate with threaded screws (TS group) or smooth pegs (SP group) on the contralateral side. Seven pairs of humeri were tested for cyclic bending, and 7 pairs for cyclic torsional evaluation: TS bending, SP bending, TS torsion, and SP torsion. The bending protocol consisted of cyclically loading to a maximum of 7.5 Nm bending moment for 10,000 cycles. The torsion protocol consisted of ±2 Nm of axial torque for 10,000 cycles. Surgical neck fractures were simulated by excising a 10-mm wedge of bone.
RESULTS: No implant failure or screw cutout was observed in any of the groups tested. Under bending loads, mean displacement of the distal fragment was significantly less for the SP group than that for the TS group over 5,000 cycles. In torsion tests, no significant differences were observed between the 2 fixation methods. DISCUSSION: The SP group demonstrated superior biomechanical characteristics to the TS group in regards to cyclic bending.
Copyright © 2013 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22743070     DOI: 10.1016/j.jse.2012.04.005

Source DB:  PubMed          Journal:  J Shoulder Elbow Surg        ISSN: 1058-2746            Impact factor:   3.019


  5 in total

1.  Comparison of CFR-PEEK and conventional titanium locking plates for proximal humeral fractures: a retrospective controlled study of patient outcomes.

Authors:  A Padolino; G Porcellini; B Guollo; E Fabbri; G N Kiran Kumar; P Paladini; G Merolla
Journal:  Musculoskelet Surg       Date:  2018-10-20

Review 2.  Biomechanical analysis of plate systems for proximal humerus fractures: a systematic literature review.

Authors:  Ali Jabran; Chris Peach; Lei Ren
Journal:  Biomed Eng Online       Date:  2018-04-27       Impact factor: 2.819

3.  Evaluation of open reduction and internal fixation of proximal humerus fractures using a locking plate with smooth pegs - A retrospective study.

Authors:  Frida Hansson; Magdalena Riddar; Anders Ekelund
Journal:  Shoulder Elbow       Date:  2018-08-07

4.  Hybrid blade and locking plate fixation for proximal humerus fractures: a comparative biomechanical analysis.

Authors:  Ali Jabran; Chris Peach; Zhenmin Zou; Lei Ren
Journal:  Biomed Eng Online       Date:  2018-01-25       Impact factor: 2.819

5.  Pegs not superior to screws for fixation of fractures of the proximal humerus.

Authors:  Ingrid Bønes; Anna Cecilie Karlberg; Maria Liljeholm; Alexander Nilsskog Fraser; Jan Erik Madsen; Tore Fjalestad
Journal:  J Orthop Surg Res       Date:  2022-02-02       Impact factor: 2.359

  5 in total