| Literature DB >> 22733899 |
Eun-Suk Cho1, Jeong-Sik Yu, Jhii-Hyun Ahn, Joo Hee Kim, Jae-Joon Chung, Hyeon-Kyeong Lee, Kyung Hee Lee.
Abstract
OBJECTIVE: The objective of our study was to investigate the feasibility of an 80-kVp protocol using a moderate-concentration contrast medium (CM) for CT angiography of the renal arteries by comparison with a conventional 120-kVp protocol using a high-concentration CM. SUBJECTS AND METHODS: Attenuation values and signal-to-noise ratios (SNRs) were determined in a phantom for the 120-kVp protocol with a high-concentration CM and the 80-kVp protocol with a moderate-concentration CM. In addition, 50 patients were prospectively enrolled in the study: 25 were scanned with 120 kVp and 200 effective mAs (mAs(eff)) after the administration of 110 mL of high-concentration CM (370 mg I/mL), and the other 25 were scanned with 80 kVp and 585 mAs(eff) after the administration of 110 mL of moderate-concentration CM (300 mg I/mL). Images of the two groups were compared in terms of arterial attenuation, SNR, contrast-to-noise ratio (CNR), and subjective degree of arterial enhancement and image quality.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22733899 DOI: 10.2214/AJR.11.7450