Performance of atrial tachyarrhythmia-sensing algorithms in dual-chamber pacing using a fixed long AV delay in patients with sinus node dysfunction.

Sami Pakarinen1, Lauri Toivonen.   

Abstract

AIMS: The aim of the study was to evaluate the performance of pacemakers (PM) atrial tachyarrhythmia (AT)-sensing algorithms in sinus node dysfunction (SND) patients with DDDR pacing programmed with a fixed long atrioventricular (AV) delay.
METHODS: In a prospective study, a total of 60 patients with SND were implanted with a dual-chamber PM with two different algorithms for detection of ATs. The study was done with a 3 month data collection period retrieved from the memory of PM and with a 7 day external Holter recording period.
RESULTS: In 13 of 16 (81 %) patients whose Holter recording revealed the presence of ATs, episodes of AT sensing were retrieved from the PM memory with electrograms verifications, confirming that the devices had detected the ATs. Very short ATs seen in Holter recordings were missed by the PM with three patients. However, with all these patients after 3 months of follow up period, there were recognized periods of ATs by the PM. With ten (17 %) patients, there were intermittent periods of undersensing by the PM although continuous atrial fibrillation (AF) was seen in the Holter recording. Retrograde conduction caused false AT detection due to repetitive non-reentrant ventriculoatrial synchronous rhythm (RNRVAS) in six (25 %) of the 24 patients with retrograde conduction.
CONCLUSIONS: Even with long AV delay, ATs can be accurately identified. However, transient undersensing of continuous AF and non-detection of very short AT episodes can still occur. Programming a long AV delay predisposes to RNRVAS which can cause false AT detection and symptoms in SND patients who have retrograde conduction.

Entities:  

Mesh:

Year:  2012        PMID: 22688950     DOI: 10.1007/s10840-012-9691-4

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  16 in total

Review 1.  Pacemaker repetitive nonreentrant ventriculoatrial synchronous rhythm. A review.

Authors:  S S Barold; P A Levine
Journal:  J Interv Card Electrophysiol       Date:  2001-03       Impact factor: 1.900

2.  Morphological analysis of sinus and retrograde atrial waves detected through a permanent pacemaker atrial lead.

Authors:  Sami Pakarinen; Anne-Mari Vitikainen; Giorgio Corbucci; Lauri Toivonen
Journal:  J Interv Card Electrophysiol       Date:  2010-10-13       Impact factor: 1.900

3.  Prevalence and clinical implications of atrial fibrillation episodes detected by pacemaker in patients with sick sinus syndrome.

Authors:  H F Tse; C P Lau
Journal:  Heart       Date:  2005-03       Impact factor: 5.994

4.  Management of far-field R wave sensing for the avoidance of inappropriate mode switch in dual chamber pacemakers: results of the FFS-test study.

Authors:  Christof Kolb; Bernd Wille; Dominik Maurer; Andreas Schuchert; Ralf Weber; Volker Schibgilla; Norbert Klein; Alexander Hümmer; Claus Schmitt; Bernhard Zrenner
Journal:  J Cardiovasc Electrophysiol       Date:  2006-09

5.  Guidelines for cardiac pacing and cardiac resynchronization therapy. The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in collaboration with the European Heart Rhythm Association.

Authors:  Panos E Vardas; Angelo Auricchio; Jean-Jacques Blanc; Jean-Claude Daubert; Helmut Drexler; Hugo Ector; Maurizio Gasparini; Cecilia Linde; Francisco Bello Morgado; Ali Oto; Richard Sutton; Maria Trusz-Gluza
Journal:  Europace       Date:  2007-08-28       Impact factor: 5.214

6.  Palpitations after a pacemaker generator exchange: a new algorithm-induced cause of endless loop tachycardia.

Authors:  Juan C Velasquez-Castano; Michael S Lloyd
Journal:  Heart Rhythm       Date:  2009-04-01       Impact factor: 6.343

7.  Verification of pacemaker automatic mode switching for the detection of atrial fibrillation and atrial tachycardia with Holter recording.

Authors:  Willem G de Voogt; Norbert M van Hemel; Arjan A van de Bos; Juhani Koïstinen; Jules H Fast
Journal:  Europace       Date:  2006-11       Impact factor: 5.214

8.  Validation of dual-chamber pacemaker diagnostic data using dual-channel stored electrograms.

Authors:  Bernd Nowak; James McMeekin; Marie Knops; Bernd Wille; Eberhard Schröder; Concepcion Moro; Matthias Oelher; Eduardo Castellanos; Benoit Coutu; Ben Petersen; Wolfgang Pfeil; Joachim Kreuzer
Journal:  Pacing Clin Electrophysiol       Date:  2005-07       Impact factor: 1.976

9.  Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction.

Authors:  Michael O Sweeney; Anne S Hellkamp; Kenneth A Ellenbogen; Arnold J Greenspon; Roger A Freedman; Kerry L Lee; Gervasio A Lamas
Journal:  Circulation       Date:  2003-06-02       Impact factor: 29.690

10.  Eliminating right ventricular pacing may not be best for patients requiring implantable cardioverter-defibrillators.

Authors:  Brian Olshansky; John D Day; Darin R Lerew; Scott Brown; Kira Q Stolen
Journal:  Heart Rhythm       Date:  2007-04-06       Impact factor: 6.343

View more
  2 in total

Review 1.  A review of the atrial upper rate algorithms of St. Jude Medical (Abbott) cardiac implantable electronic devices : Incidence of repetitive nonreentrant ventriculoatrial synchrony (RNRVAS).

Authors:  S S Barold
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2017-09

2.  Enhanced detection of atrial tachyarrhythmias with pacing devices by using more accurate atrial sensing.

Authors:  Sami Pakarinen; Mika Lehto; Jaap Ruiter; Willem G de Voogt
Journal:  J Interv Card Electrophysiol       Date:  2021-10-01       Impact factor: 1.759

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.