Validation of dual-chamber pacemaker diagnostic data using dual-channel stored electrograms.

Bernd Nowak1, James McMeekin, Marie Knops, Bernd Wille, Eberhard Schröder, Concepcion Moro, Matthias Oelher, Eduardo Castellanos, Benoit Coutu, Ben Petersen, Wolfgang Pfeil, Joachim Kreuzer.   

Abstract

BACKGROUND: Pacemaker diagnostic counters are used to guide device programming and patient management. However, these data are susceptible to inappropriate classification of events. The aim of this multicenter study was to evaluate pacemaker diagnostic data using stored intracardiac electrograms (EGMs).
METHODS: The study included 351 patients (191 males, aged 71 +/- 10 years) with standard indications for dual-chamber pacemaker implantation. EGM triggers were atrial tachycardia (AT), ventricular tachycardia (VT), sudden bradycardia response (SBR), and pacemaker-mediated tachycardia (PMT). For this study, the devices could store up to 5 EGMs of 8s each (with marker annotation and onset recording). After 3 months, the EGMs were analyzed and classified as "confirmed" if the EGM validated the trigger and as "false positive" if the EGM showed an event different from the trigger.
RESULTS: Of the 1,003 EGMs available, the triggers were AT in 640 EGMs, VT in 76, SBR in 105, and PMT in 178 EGMs. Four EGMs were triggered by magnet application. The trigger was confirmed in 614 EGMs (62%): 62% of AT episodes, 18% of VT episodes, 100% of SBR episodes, and 54% of PMT episodes. In 385 cases (45%), the EGMs revealed false-positive events due to far-field sensing (39%), noise and myopotential sensing (26%), sinus tachycardias (21%), double counting (9%), exit block (4%), and undersensing (1%).
CONCLUSION: This large-scale study of stored EGMs revealed their value in validating diagnostic counter data. Therapeutic decisions should not be based on diagnostic counters alone; they should be validated by sophisticated tools like stored EGMs.

Entities:  

Mesh:

Year:  2005        PMID: 16008796     DOI: 10.1111/j.1540-8159.2005.00159.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

1.  Inappropriate mode switching clarified by using a chest radiograph.

Authors:  Brian Marino; Abhishek Jaiswal; Seth Goldbarg
Journal:  J Arrhythm       Date:  2015-02-14

2.  Atrioventricular nodal reentrant tachycardia diagnosed by pacemaker electrograms.

Authors:  Christophe d'Ivernois; Marc Hero; Isaac Sayumwe; Patrick Blanc
Journal:  J Interv Card Electrophysiol       Date:  2006-11-21       Impact factor: 1.900

3.  Atrial pacing for rhythm control of atrial fibrillation.

Authors:  I C van Gelder; A C P Wiesfeld
Journal:  Neth Heart J       Date:  2008-06       Impact factor: 2.380

4.  Performance of atrial tachyarrhythmia-sensing algorithms in dual-chamber pacing using a fixed long AV delay in patients with sinus node dysfunction.

Authors:  Sami Pakarinen; Lauri Toivonen
Journal:  J Interv Card Electrophysiol       Date:  2012-06-12       Impact factor: 1.900

5.  Diagnostic tools for atrial tachyarrhythmias in implantable pacemakers: a review of technical options and pitfalls.

Authors:  W G de Voogt; N M van Hemel
Journal:  Neth Heart J       Date:  2008-06       Impact factor: 2.380

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.