Marked discrepancy in pressure gradient between Doppler and catheter examinations on Medtronic Mosaic valve in aortic position.

Toshiaki Ito1, Atsuo Maekawa, Genyo Fujii, Sadanari Sawaki, Satoshi Hoshino, Yasunari Hayashi.   

Abstract

A 71-year-old woman underwent aortic valve replacement with 23 mm Medtronic Mosaic Ultra valve 4 years ago because of aortic stenosis. Although she had been asymptomatic since the operation, echocardiography showed 4 m/s of transprosthetic valve flow that implied early prosthetic valve failure. Catheter examination revealed that the mean transvalvular pressure gradient during systole was 15.1 mmHg on simultaneous pressure recording, and calculated valve area 1.82 cm(2). Her body surface area was 1.56 m(2). Prosthetic valve failure and prosthesis-patient mismatch were both denied. The discrepancy between Doppler study data and catheter data seemed to be due to fluid dynamical pressure recovery phenomenon. Net pressure difference between the left ventricle and the aorta may be significantly smaller than that estimated using Bernoulli's equation from transvalvular flow speed in some patients after aortic valve replacement.

Entities:  

Mesh:

Year:  2012        PMID: 22688582     DOI: 10.1007/s11748-012-0089-0

Source DB:  PubMed          Journal:  Gen Thorac Cardiovasc Surg        ISSN: 1863-6705


  6 in total

1.  "Overestimation" of catheter gradients by Doppler ultrasound in patients with aortic stenosis: a predictable manifestation of pressure recovery.

Authors:  H Baumgartner; T Stefenelli; J Niederberger; H Schima; G Maurer
Journal:  J Am Coll Cardiol       Date:  1999-05       Impact factor: 24.094

2.  Discrepancies between Doppler and catheter gradients in aortic prosthetic valves in vitro. A manifestation of localized gradients and pressure recovery.

Authors:  H Baumgartner; S Khan; M DeRobertis; L Czer; G Maurer
Journal:  Circulation       Date:  1990-10       Impact factor: 29.690

3.  Importance of pressure recovery for the assessment of aortic stenosis by Doppler ultrasound. Role of aortic size, aortic valve area, and direction of the stenotic jet in vitro.

Authors:  J Niederberger; H Schima; G Maurer; H Baumgartner
Journal:  Circulation       Date:  1996-10-15       Impact factor: 29.690

4.  Impact of prosthesis-patient size on functional recovery after aortic valve replacement.

Authors:  Colleen Gorman Koch; Farah Khandwala; Fawzy G Estafanous; Floyd D Loop; Eugene H Blackstone
Journal:  Circulation       Date:  2005-06-13       Impact factor: 29.690

Review 5.  The problem of valve prosthesis-patient mismatch.

Authors:  S H Rahimtoola
Journal:  Circulation       Date:  1978-07       Impact factor: 29.690

6.  Net pressure gradients in aortic prosthetic valves can be estimated by Doppler.

Authors:  Odd Bech-Hanssen; Peter Gjertsson; Erik Houltz; Bengt Wranne; Per Ask; Dan Loyd; Kenneth Caidahl
Journal:  J Am Soc Echocardiogr       Date:  2003-08       Impact factor: 5.251

  6 in total
  1 in total

1.  Comparable long-term results for porcine and pericardial prostheses after isolated aortic valve replacement.

Authors:  Martin Andreas; Stephanie Wallner; Kurt Ruetzler; Dominik Wiedemann; Marek Ehrlich; Georg Heinze; Thomas Binder; Anton Moritz; Michael J Hiesmayr; Alfred Kocher; Guenther Laufer
Journal:  Eur J Cardiothorac Surg       Date:  2014-12-18       Impact factor: 4.191

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.