Efficacy and safety of the novel electrosurgical vessel sealing and cutting instrument BiCision®.

Ralf Rothmund1, Bernhard Kraemer, Felix Neis, Sara Brucker, Markus Wallwiener, Ali Reda, Andrea Hausch, Marcus Scharpf, Mara Natascha Szyrach.   

Abstract

BACKGROUND: The use of energy-based tissue-sealing and cutting instruments is becoming more and more popular in visceral, urological, and gynecological surgery. For their safe and efficacious use in clinical practice, such instruments have to reliably seal vessels with a minimal sealing failure rate, cause minimal thermal damage to adjacent tissue, and have good cutting qualities.
METHODS: The efficacy and safety of the novel energy-based instrument for dissection, hemostasis and cutting (BiCision(®), ERBE) was compared to a commercially available device (EnSeal(®), Ethicon Endo-Surgery). We investigated vessel-sealing reliability (success rate), sealing quality and sealing time, lateral thermal damage cutting quality, tissue sticking to the instrument, burst pressure and delayed complications in an acute and chronic pig model after splenectomy, small bowel resection, nephrectomy, salpingo-oophorectomy, and sealing of peripheral vessels.
RESULTS: For all parameters investigated, BiCision(®) was at least equivalent to EnSeal(®). BiCision(®) was even superior to EnSeal(®) with respect to the burst pressure in arteries (p = 0.044) and veins (p = 0.023) and the cut quality in all locations (arteries, p = 0.0009; veins, p = 0.043). The course of the 7-day chronic study was uneventful except for one animal that developed an intestinal obstruction. None of the animals showed any signs of postoperative bleeding. On second-look laparotomy at day 7, macroscopic inspection of the sealed tissue and vessels did not show any signs of complications or evidence that bleeding had occurred. Histologically, the integrity of vessel wall fusion, thermal alterations, and inflammatory reactions were comparable, confirming substantial equivalence.
CONCLUSION: We demonstrated that the efficacy and quality of vessel sealing with BiCision(®) is at least equivalent to those of EnSeal(®) for vessel diameters up to 7 mm. Since EnSeal(®) has already been shown to be safe in clinical practice, BiCision(®) should be as reliable as EnSeal(®) under clinical conditions.

Entities:  

Mesh:

Year:  2012        PMID: 22678171     DOI: 10.1007/s00464-012-2337-x

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  23 in total

Review 1.  Is laparoscopy back?

Authors:  Avner Hershlag; Jacob Markovitz
Journal:  Fertil Steril       Date:  2005-12       Impact factor: 7.329

2.  Comparison of four energy-based vascular sealing and cutting instruments: A porcine model.

Authors:  Benjamin Person; David A Vivas; Dan Ruiz; Michael Talcott; James E Coad; Steven D Wexner
Journal:  Surg Endosc       Date:  2007-12-20       Impact factor: 4.584

3.  Collagen-elastin ratio predicts burst pressure of arterial seals created using a bipolar vessel sealing device in a porcine model.

Authors:  David Sindram; Kimberly Martin; Jarrod P Meadows; Ajita S Prabhu; Jessica J Heath; Iain H McKillop; David A Iannitti
Journal:  Surg Endosc       Date:  2011-03-15       Impact factor: 4.584

4.  High-burst-strength, feedback-controlled bipolar vessel sealing.

Authors:  J S Kennedy; P L Stranahan; K D Taylor; J G Chandler
Journal:  Surg Endosc       Date:  1998-06       Impact factor: 4.584

Review 5.  Contemporary adhesion prevention.

Authors:  G S diZerega
Journal:  Fertil Steril       Date:  1994-02       Impact factor: 7.329

Review 6.  Postoperative adhesions: their treatment and relevance in clinical practice.

Authors:  D Menzies
Journal:  Ann R Coll Surg Engl       Date:  1993-05       Impact factor: 1.891

7.  Efficacy and safety of 5-mm-diameter bipolar and ultrasonic shears for cutting carotid arteries of the hybrid pig.

Authors:  René Mantke; W Halangk; A Habermann; B Peters; S Konrad; M Guenther; H Lippert
Journal:  Surg Endosc       Date:  2010-07-08       Impact factor: 4.584

8.  Prospective comparison of four laparoscopic vessel ligation devices.

Authors:  Gregory R Lamberton; Ryan S Hsi; Daniel H Jin; Tekisha U Lindler; Forrest C Jellison; D Duane Baldwin
Journal:  J Endourol       Date:  2008-10       Impact factor: 2.942

9.  Experience with a new energy source for tissue fusion in pediatric patients.

Authors:  Todd A Ponsky; Arjun Khosla; Steven S Rothenberg
Journal:  J Laparoendosc Adv Surg Tech A       Date:  2009-04       Impact factor: 1.878

10.  Thermal conduction, compression, and electrical current--an evaluation of major parameters of electrosurgical vessel sealing in a porcine in vitro model.

Authors:  Christian W Wallwiener; Taufiek K Rajab; Wolfgang Zubke; Keith B Isaacson; Markus Enderle; Daniel Schäller; Markus Wallwiener
Journal:  J Minim Invasive Gynecol       Date:  2008-07-21       Impact factor: 4.137

View more
  9 in total

1.  Prospective analysis of the sealing ability of the ENSEAL® G2 Articulating Tissue Sealer and transector on human mesenteric vessels in colorectal surgery.

Authors:  B Balachandran; G Melich; T Mustafa; S J Marecik; L M Prasad; M Gonzalez; S Sulo; F Dabbous; J J Park
Journal:  Tech Coloproctol       Date:  2017-01-31       Impact factor: 3.781

2.  Heat injury to the inferior vena cava by bipolar tissue sealer.

Authors:  Akira Chikamoto; Takayoshi Kaida; Kota Arima; Takaaki Higashi; Katsunobu Taki; Satoshi Ida; Hirohisa Okabe; Hidetoshi Nitta; Hiromitsu Hayashi; Daisuke Hashimoto; Masayuki Watanabe; Toru Beppu; Hideo Baba
Journal:  Surg Endosc       Date:  2015-07-03       Impact factor: 4.584

3.  A prospective, randomized clinical comparison between UltraCision and the novel sealing and cutting device BiCision in patients with laparoscopic supracervical hysterectomy.

Authors:  Ralf Rothmund; Mara Szyrach; Ali Reda; Markus D Enderle; Alexander Neugebauer; Florin-Andrei Taran; Sara Brucker; Andrea Hausch; Christian Wallwiener; Bernhard Kraemer
Journal:  Surg Endosc       Date:  2013-05-14       Impact factor: 4.584

4.  Surface-processing technology of a microgrooving and water-repellent coating improves the fusion potential of an ultrasonic energy device.

Authors:  Satoru Okada; Junichi Shimada; Kazuhiro Ito; Tatsuo Ishii; Koichiro Oshiumi
Journal:  Surg Endosc       Date:  2016-06-22       Impact factor: 4.584

5.  Comparison of the harmonic focus and the thunderbeat for open thyroidectomy.

Authors:  Sam Van Slycke; Jean-Pierre Gillardin; Klaas Van Den Heede; Joan Minguet; Hubert Vermeersch; Nele Brusselaers
Journal:  Langenbecks Arch Surg       Date:  2016-05-25       Impact factor: 3.445

6.  Evaluation of a novel electrosurgical sealing mode in an ex vivo and in vivo porcine model.

Authors:  Karolin Thiel; Walter Linzenbold; Markus D Enderle; B Nold; Alfred Königsrainer; Martin Schenk; Christian Thiel
Journal:  Surg Endosc       Date:  2017-09-18       Impact factor: 4.584

7.  Evaluation of the HystSim™-virtual reality trainer: an essential additional tool to train hysteroscopic skills outside the operation theater.

Authors:  Felix Neis; Sara Brucker; Melanie Henes; F Andrei Taran; Sascha Hoffmann; Markus Wallwiener; Birgitt Schönfisch; Nicole Ziegler; Angelika Larbig; Rudy Leon De Wilde
Journal:  Surg Endosc       Date:  2016-03-09       Impact factor: 4.584

8.  In silico evaluation of geometry variations with respect to the thermal spread during coagulation of egg white using bipolar vessel sealing instruments.

Authors:  Jay Wagenpfeil; Christina Schöllig; Volker Mayer; Ronny Feuer; Bernhard Nold; Alexander Neugebauer; Michael Ederer; Ralf Rothmund; Bernhard Krämer; Sara Brucker; Markus Enderle; Oliver Sawodny; Julia Rex
Journal:  Biomed Eng Online       Date:  2016-11-02       Impact factor: 2.819

9.  Safety and effectiveness of a novel generator algorithm for bipolar vessel sealing: a randomised controlled chronic animal study.

Authors:  Bernhard Kraemer; Christos Tsaousidis; Stephan Kruck; Martin Schenk; Marcus Scharpf; Stefan Kommoss; Sara Brucker; Daniela Nuessle; Markus D Enderle; Ulrich Biber
Journal:  BMC Surg       Date:  2019-11-05       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.