Literature DB >> 22665568

What could be the cause of late syncope after pacemaker implantation?

Mohammad Hossein Mandegar1, Farideh Roshanali, Helia Farokhzad, Mahmood Eftekharzadeh.   

Abstract

A 21-year-old man who had suffered an episode of complete heart block 2 months earlier and thus undergone pacemaker implantation presented with syncope. Echocardiography showed cardiac tamponade and raised the suspicion of lead penetration. The effusion was drained. Multi-sliced CT scan confirmed that the right aspect of the pacemaker wire was extracardiac, causing the cardiac tamponade.

Entities:  

Mesh:

Year:  2012        PMID: 22665568      PMCID: PMC3291030          DOI: 10.1136/bcr.12.2011.5353

Source DB:  PubMed          Journal:  BMJ Case Rep        ISSN: 1757-790X


  5 in total

1.  Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT.

Authors:  David A Hirschl; Vineet R Jain; Hugo Spindola-Franco; Jay N Gross; Linda B Haramati
Journal:  Pacing Clin Electrophysiol       Date:  2007-01       Impact factor: 1.976

2.  Proper management of pericardial tamponade as a late complication of implantable cardiac device placement.

Authors:  Glenn M Polin; Erica Zado; Hemal Nayak; Joshua M Cooper; Andrea M Russo; Sanjay Dixit; David Lin; Francis E Marchlinski; Ralph J Verdino
Journal:  Am J Cardiol       Date:  2006-05-11       Impact factor: 2.778

3.  [Syncope of unknown origin in patients with permanent auriculoventricular block with an implanted pacemaker. Usefulness of the tilt table test].

Authors:  Manlio F Márquez; Carmen F Encarnación-Roa; Antonio G Hermosillo; William J Benítez-Pinto; Manuel Cárdenas
Journal:  Rev Esp Cardiol       Date:  2002-04       Impact factor: 4.753

4.  Delayed complications following pacemaker implantation.

Authors:  Kenneth A Ellenbogen; Mark A Wood; Richard K Shepard
Journal:  Pacing Clin Electrophysiol       Date:  2002-08       Impact factor: 1.976

5.  Ventricular pacing: to pace or not to pace.

Authors:  Paul A Levine
Journal:  Europace       Date:  2010-01       Impact factor: 5.214

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.